Screen Printed Alizarin-Based Carbon Electrodes: Monitoring pH in Unbuffered Media
An alizarin based voltammetric pH sensor is introduced. This device is fabricated by screen printing the alizarin, which is a pH‐sensitive redox compound, mixed with carbon ink onto the carbon aluminum (aluminum foil coated with carbon paste). This electrode shows a Nernst response to both the buffe...
Gespeichert in:
Veröffentlicht in: | Electroanalysis (New York, N.Y.) N.Y.), 2015-04, Vol.27 (4), p.917-923 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | An alizarin based voltammetric pH sensor is introduced. This device is fabricated by screen printing the alizarin, which is a pH‐sensitive redox compound, mixed with carbon ink onto the carbon aluminum (aluminum foil coated with carbon paste). This electrode shows a Nernst response to both the buffered and unbuffered media, with a potential shift by 55.8 mV/pH unit and a working range up to pH 10. This approach suggests a disposable, reproducible, low‐cost method for pH sensing with the possibility of miniaturization. A further study of this type of sensor coated with nafion reveals a possible method to expand the working range and improve the stability. |
---|---|
ISSN: | 1040-0397 1521-4109 |
DOI: | 10.1002/elan.201400704 |