New polycrystalline solid state responsive electrodes for the determination of the selenite ion

New electrodes for the selective determination of selenite were constructed by compressing either HgSeO3 or a mixture of HgSeO3 and Hg2Cl2. The linear concentration range was 6.0 × 10−5−10−1 M with a slope of 29.5 mV/concentration decade. The usable pH range for both electrodes was wide (3–10) and t...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Electroanalysis (New York, N.Y.) N.Y.), 1995-06, Vol.7 (6), p.587-590
Hauptverfasser: Zareh, Mohsen M., Amin, Alaa S., Abdel-Aziz, Manar
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:New electrodes for the selective determination of selenite were constructed by compressing either HgSeO3 or a mixture of HgSeO3 and Hg2Cl2. The linear concentration range was 6.0 × 10−5−10−1 M with a slope of 29.5 mV/concentration decade. The usable pH range for both electrodes was wide (3–10) and their life span was up to 8 days. Selectivity coefficient values were determined for various anions and no significant interferences were observed for most of them. The solubility products of some sparingly soluble selenite salts were determined using the latter electrode.
ISSN:1040-0397
1521-4109
DOI:10.1002/elan.1140070616