Heat on – Calibration of IR Spectroscopy at Process Conditions for Continuous Esterification

This work tries to close the gap between continuous synthesis, measurement, calibration, and analysis on R&D scale. By introducing new calibration methods, the full potential of inline ATR‐FTIR spectroscopy in continuous reactor setups is unraveled. A fast, fully automated calibration of the rea...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Chemie ingenieur technik 2020-10, Vol.92 (10), p.1467-1473
Hauptverfasser: Schnoor, Johann‐Kilian, Schulte, Lennard D., Liauw, Marcel A.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:This work tries to close the gap between continuous synthesis, measurement, calibration, and analysis on R&D scale. By introducing new calibration methods, the full potential of inline ATR‐FTIR spectroscopy in continuous reactor setups is unraveled. A fast, fully automated calibration of the reaction mixtures is carried out at process conditions. The use of different compositions underlines the robustness of this calibration method, which was used for the determination of kinetic parameters. The conversion‐based calibration significantly increased the precision of the analysis. In difference, the new method only requires the preparation of two calibration mixtures. By that, this method can also be applied to large‐scale processes and allows the use of a single measurement setup from R&D to industrial scale. Spectroscopic measurement techniques have found their way into R&D process development. A new calibration method was developed to allow precise measurements using ATR‐FTIR spectroscopy in continuous reactors in R&D and industrial scale. The calibration under process conditions increased precision and avoided temperature‐ and pressure‐related shifts.
ISSN:0009-286X
1522-2640
DOI:10.1002/cite.202000050