An image fusion tool for echo‐guided left ventricular lead placement in cardiac resynchronization therapy: Performance and workflow integration analysis
Background The response rate to cardiac resynchronization therapy (CRT) may be improved if echocardiographic‐derived parameters are used to guide the left ventricular (LV) lead deployment. Tools to visually integrate deformation imaging and fluoroscopy to take advantage of the combined information a...
Gespeichert in:
Veröffentlicht in: | Echocardiography (Mount Kisco, N.Y.) N.Y.), 2019-10, Vol.36 (10), p.1834-1845 |
---|---|
Hauptverfasser: | , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Background
The response rate to cardiac resynchronization therapy (CRT) may be improved if echocardiographic‐derived parameters are used to guide the left ventricular (LV) lead deployment. Tools to visually integrate deformation imaging and fluoroscopy to take advantage of the combined information are lacking.
Methods
An image fusion tool for echo‐guided LV lead placement in CRT was developed. A personalized average 3D cardiac model aided visualization of patient‐specific LV function in fluoroscopy. A set of coronary venography‐derived landmarks facilitated registration of the 3D model with fluoroscopy into a single multimodality image. The fusion was both performed and analyzed retrospectively in 30 cases. Baseline time‐to‐peak values from echocardiography speckle‐tracking radial strain traces were color‐coded onto the fused LV. LV segments with suspected scar tissue were excluded by cardiac magnetic resonance imaging. The postoperative augmented image was used to investigate: (a) registration accuracy and (b) agreement between LV pacing lead location, echo‐defined target segments, and CRT response.
Results
Registration time (264 ± 25 seconds) and accuracy (4.3 ± 2.3 mm) were found clinically acceptable. A good agreement between pacing location and echo‐suggested segments was found in 20 (out of 21) CRT responders. Perioperative integration of the proposed workflow was successfully tested in 2 patients. No additional radiation, compared with the existing workflow, was required.
Conclusions
The fusion tool facilitates understanding of the spatial relationship between the coronary veins and the LV function and may help targeted LV lead delivery. |
---|---|