Design and experimental study of a two-dimensional position sensitive X-ray detector

A prototype of a two-dimensional position sensitive X-ray detector was designed and constructed for small angle X-ray scattering experiments at BSFR (Beijing Synchrotron Radiation Facility). The detector is based on MWPC with cathode strip readout, and has a sensitive area of 200 mm×200 mm. The spat...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Chinese physics C 2008-11, Vol.32 (11), p.903-907
1. Verfasser: 王小胡 朱启明 陈元柏 陈昌 刘荣光 王岚 金艳 马骁妍 江山
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A prototype of a two-dimensional position sensitive X-ray detector was designed and constructed for small angle X-ray scattering experiments at BSFR (Beijing Synchrotron Radiation Facility). The detector is based on MWPC with cathode strip readout, and has a sensitive area of 200 mm×200 mm. The spatial resolution (FWHM) of about 210 μm along the anode wire direction was obtained from the ^55Fe X-ray test of the detector.
ISSN:1674-1137
0254-3052
2058-6132
DOI:10.1088/1674-1137/32/11/011