Design and experimental study of a two-dimensional position sensitive X-ray detector
A prototype of a two-dimensional position sensitive X-ray detector was designed and constructed for small angle X-ray scattering experiments at BSFR (Beijing Synchrotron Radiation Facility). The detector is based on MWPC with cathode strip readout, and has a sensitive area of 200 mm×200 mm. The spat...
Gespeichert in:
Veröffentlicht in: | Chinese physics C 2008-11, Vol.32 (11), p.903-907 |
---|---|
1. Verfasser: | |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A prototype of a two-dimensional position sensitive X-ray detector was designed and constructed for small angle X-ray scattering experiments at BSFR (Beijing Synchrotron Radiation Facility). The detector is based on MWPC with cathode strip readout, and has a sensitive area of 200 mm×200 mm. The spatial resolution (FWHM) of about 210 μm along the anode wire direction was obtained from the ^55Fe X-ray test of the detector. |
---|---|
ISSN: | 1674-1137 0254-3052 2058-6132 |
DOI: | 10.1088/1674-1137/32/11/011 |