The i-TED Compton Camera Array for real-time boron imaging and determination during treatments in Boron Neutron Capture Therapy
This paper explores the adaptation and application of i-TED Compton imagers for real-time dosimetry in Boron Neutron Capture Therapy (BNCT). The i-TED array, previously utilized in nuclear astrophysics experiments at CERN, is being optimized for detecting and imaging 478 keV gamma-rays, critical for...
Gespeichert in:
Hauptverfasser: | , , , , , , , , , , , , , , |
---|---|
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext bestellen |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | This paper explores the adaptation and application of i-TED Compton imagers
for real-time dosimetry in Boron Neutron Capture Therapy (BNCT). The i-TED
array, previously utilized in nuclear astrophysics experiments at CERN, is
being optimized for detecting and imaging 478 keV gamma-rays, critical for
accurate BNCT dosimetry. Detailed Monte Carlo simulations were used to optimize
the i-TED detector configuration and enhance its performance in the challenging
radiation environment typical of BNCT. Additionally, advanced 3D image
reconstruction algorithms, including a combination of back-projection and
List-Mode Maximum Likelihood Expectation Maximization (LM-MLEM), are
implemented and validated through simulations. Preliminary experimental tests
at the Institut Laue-Langevin (ILL) demonstrate the potential of i-TED in a
clinical setting, with ongoing experiments focusing on improving imaging
capabilities in realistic BNCT conditions. |
---|---|
DOI: | 10.48550/arxiv.2409.10107 |