An analytical model for photomultiplier tube calibration

The purpose of the present article is to offer an analytical model that can be used for the calibration of photomultiplier tubes (PMTs). The derivation of the mathematical formulae is discussed extensively and we apply this machinery to a real problem; the gain determination of the R7081 Hamamatsu P...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:arXiv.org 2023-11
1. Verfasser: Kalousis, L N
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The purpose of the present article is to offer an analytical model that can be used for the calibration of photomultiplier tubes (PMTs). The derivation of the mathematical formulae is discussed extensively and we apply this machinery to a real problem; the gain determination of the R7081 Hamamatsu PMT. We hope that our work can provide a third, powerful alternative to the task of gain deconvolution. We also believe that the techniques explained in this article can be applied to other problems in PMT calibration and monitoring.
ISSN:2331-8422
DOI:10.48550/arxiv.2304.08735