Application of single metal Au and Ag/Si3N4/Au plasmonic sensors for glucose refractive index measurements
We have employed single gold layer metal and multilayer Ag/Si3N4/Au plasmonic sensors to monitor changes in the refraction index of glucose/water solutions. Our results show that the sensor can resolve glucose refractive index down to 1% concentration. In addition, we have simulated the strength of...
Gespeichert in:
Hauptverfasser: | , , |
---|---|
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext bestellen |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We have employed single gold layer metal and multilayer Ag/Si3N4/Au plasmonic
sensors to monitor changes in the refraction index of glucose/water solutions.
Our results show that the sensor can resolve glucose refractive index down to
1% concentration. In addition, we have simulated the strength of the surface
plasmon resonance evanescent electric fields by using the finite difference
method of COMSOL. The electric field strength for the Ag/Si3N4/Au sensor is
much stronger than that for single gold layer SPR sensor. |
---|---|
DOI: | 10.48550/arxiv.2007.00064 |