Absorbed Dose in Ion Beams: Comparison of Ionisation and Fluence-based Measurements

A direct comparison measurement of fluorescent nuclear track detectors (FNTDs) and a thimble ionisation chamber is presented. Irradiations were performed using monoenergetic protons (142.66 MeV, 3*10^6 1/cm2) and carbon ions (270.55 MeV/u, 3*10^6 1/cm2). It was found that absorbed dose to water valu...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:arXiv.org 2014-02
Hauptverfasser: Julia-Maria Osinga, Brons, Stephan, Bartz, James A, Akselrod, Mark S, Jäkel, Oliver, Greilich, Steffen
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A direct comparison measurement of fluorescent nuclear track detectors (FNTDs) and a thimble ionisation chamber is presented. Irradiations were performed using monoenergetic protons (142.66 MeV, 3*10^6 1/cm2) and carbon ions (270.55 MeV/u, 3*10^6 1/cm2). It was found that absorbed dose to water values as determined by fluence measurements using FNTDs are, in case of protons, in good agreement (2.4 %) with ionization chamber measurements, if slower protons and Helium secondaries were accounted for by an effective stopping power. For carbon, however, a significant discrepancy of 4.5 % was seen, which could not be explained by fragmentation, uncertainties or experimental design. The results rather suggest a W-value of 32.10 eV +/- 2.6 %. Additionally, the abundance of secondary protons expected from Monte-Carlo transport simulation was not observed.
ISSN:2331-8422