Fabry−Pérot Cavity Sensors for Multipoint On-Column Micro Gas Chromatography Detection

We developed and characterized a Fabry−Pérot (FP) sensor module based micro gas chromatography (μGC) detector for multipoint on-column detection. The FP sensor was fabricated by depositing a thin layer of metal and a layer of gas-sensitive polymer consecutively on the endface of an optical fiber, w...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Analytical chemistry (Washington) 2010-06, Vol.82 (11), p.4370-4375
Hauptverfasser: Liu, Jing, Sun, Yuze, Howard, Daniel J, Frye-Mason, Greg, Thompson, Aaron K, Ja, Shiou-jyh, Wang, Siao-Kwan, Bai, Mengjun, Taub, Haskell, Almasri, Mahmoud, Fan, Xudong
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We developed and characterized a Fabry−Pérot (FP) sensor module based micro gas chromatography (μGC) detector for multipoint on-column detection. The FP sensor was fabricated by depositing a thin layer of metal and a layer of gas-sensitive polymer consecutively on the endface of an optical fiber, which formed the FP cavity. Light partially reflected from the metal layer and the polymer−air interface generated an interference spectrum, which shifted as the polymer layer absorbed the gas analyte. The FP sensor module was then assembled by inserting the FP sensor into a hole drilled in the wall of a fused-silica capillary, which can be easily connected to the conventional gas chromatography (GC) column through a universal quick seal column connector, thus enabling on-column real-time detection. We characterized the FP sensor module based μGC detector. Sensitive detection of various gas analytes was achieved with subnanogram detection limits. The rapid separation capability of the FP sensor module assembled with both single- and tandem-column systems was demonstrated, in which gas analytes having a wide range of polarities and volatilities were well-resolved. The tandem-column system obtained increased sensitivity and selectivity by employing two FP sensor modules coated with different polymers, showing great system versatility.
ISSN:0003-2700
1520-6882