Body Habitus and Dynamic Surgical Conditions Independently Impair Pulmonary Mechanics during Robotic-assisted Laparoscopic Surgery: A Cross-sectional Study
BACKGROUND:Body habitus, pneumoperitoneum, and Trendelenburg positioning may each independently impair lung mechanics during robotic laparoscopic surgery. This study hypothesized that increasing body mass index is associated with more mechanical strain and alveolar collapse, and these impairments ar...
Gespeichert in:
Veröffentlicht in: | Anesthesiology (Philadelphia) 2020-10, Vol.133 (4), p.750-763 |
---|---|
Hauptverfasser: | , , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | 763 |
---|---|
container_issue | 4 |
container_start_page | 750 |
container_title | Anesthesiology (Philadelphia) |
container_volume | 133 |
creator | Tharp, William G. Murphy, Serena Breidenstein, Max W. Love, Collin Booms, Alisha Rafferty, Melissa N. Friend, Alexander F. Perrapato, Scott Ahern, Thomas P. Dixon, Anne E. Bates, Jason H. T. Bender, S. Patrick |
description | BACKGROUND:Body habitus, pneumoperitoneum, and Trendelenburg positioning may each independently impair lung mechanics during robotic laparoscopic surgery. This study hypothesized that increasing body mass index is associated with more mechanical strain and alveolar collapse, and these impairments are exacerbated by pneumoperitoneum and Trendelenburg positioning.
METHODS:This cross-sectional study measured respiratory flow, airway pressures, and esophageal pressures in 91 subjects with body mass index ranging from 18.3 to 60.6 kg/m. Pulmonary mechanics were quantified at four stages(1) supine and level after intubation, (2) with pneumoperitoneum, (3) in Trendelenburg docked with the surgical robot, and (4) level without pneumoperitoneum. Subjects were stratified into five body mass index categories (less than 25, 25 to 29.9, 30 to 34.9, 35 to 39.9, and 40 or higher), and respiratory mechanics were compared over surgical stages using generalized estimating equations. The optimal positive end-expiratory pressure settings needed to achieve positive end-expiratory transpulmonary pressures were calculated.
RESULTS:At baseline, transpulmonary driving pressures increased in each body mass index category (1.9 ± 0.5 cm H2O; mean difference ± SD; P < 0.006), and subjects with a body mass index of 40 or higher had decreased mean end-expiratory transpulmonary pressures compared with those with body mass index of less than 25 (–7.5 ± 6.3 vs. –1.3 ± 3.4 cm H2O; P < 0.001). Pneumoperitoneum and Trendelenburg each further elevated transpulmonary driving pressures (2.8 ± 0.7 and 4.7 ± 1.0 cm H2O, respectively; P < 0.001) and depressed end-expiratory transpulmonary pressures (–3.4 ± 1.3 and –4.5 ± 1.5 cm H2O, respectively; P < 0.001) compared with baseline. Optimal positive end-expiratory pressure was greater than set positive end-expiratory pressure in 79% of subjects at baseline, 88% with pneumoperitoneum, 95% in Trendelenburg, and ranged from 0 to 36.6 cm H2O depending on body mass index and surgical stage.
CONCLUSIONS:Increasing body mass index induces significant alterations in lung mechanics during robotic laparoscopic surgery, but there is a wide range in the degree of impairment. Positive end-expiratory pressure settings may need individualization based on body mass index and surgical conditions. |
doi_str_mv | 10.1097/ALN.0000000000003442 |
format | Article |
fullrecord | <record><control><sourceid>wolterskluwer</sourceid><recordid>TN_cdi_wolterskluwer_health_10_1097_ALN_0000000000003442</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>00000542-202010000-00018</sourcerecordid><originalsourceid>FETCH-LOGICAL-s2492-a12f552e9c59e3a607be889713ad3132a69bcb433adbb8f691e69e04fc244a33</originalsourceid><addsrcrecordid>eNp9UMtOwzAQtBBIlMIfcPAPGPzKi1sJj1YKD9Heo43jtIbUruJEVb6FnyWBHhAHVtpdzWhnRxqELhm9YjSJrmfZ8xX9VUJKfoQmLOAxYSwKjtFkZImgnJ-iM-_fBxgFIp6gz1tX9ngOhWk7j8GW-K63sDUKL7tmbRTUOHW2NK1x1uOFLfVOD8O2dY8X2x2YBr929dZZaHr8pNUGrFEel11j7Bq_ucK1RhHw3vhWlziDHTTOK7c7OOimv8EznA6kJ16r0WfwXLZd2Z-jkwpqry8Oe4pWD_erdE6yl8dFOsuI5zLhBBivgoDrRAWJFhDSqNBxnERMQCmY4BAmhSqkGGBRxFWYMB0mmspKcSlBiCmKf97uXd3qxn_U3V43-UZD3W5yRvMx43zIOP-b8SCV_0i_7wLJCaecshGQoVksvgBQnYKn</addsrcrecordid><sourcetype>Publisher</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype></control><display><type>article</type><title>Body Habitus and Dynamic Surgical Conditions Independently Impair Pulmonary Mechanics during Robotic-assisted Laparoscopic Surgery: A Cross-sectional Study</title><source>Journals@Ovid Complete</source><source>EZB-FREE-00999 freely available EZB journals</source><creator>Tharp, William G. ; Murphy, Serena ; Breidenstein, Max W. ; Love, Collin ; Booms, Alisha ; Rafferty, Melissa N. ; Friend, Alexander F. ; Perrapato, Scott ; Ahern, Thomas P. ; Dixon, Anne E. ; Bates, Jason H. T. ; Bender, S. Patrick</creator><creatorcontrib>Tharp, William G. ; Murphy, Serena ; Breidenstein, Max W. ; Love, Collin ; Booms, Alisha ; Rafferty, Melissa N. ; Friend, Alexander F. ; Perrapato, Scott ; Ahern, Thomas P. ; Dixon, Anne E. ; Bates, Jason H. T. ; Bender, S. Patrick</creatorcontrib><description>BACKGROUND:Body habitus, pneumoperitoneum, and Trendelenburg positioning may each independently impair lung mechanics during robotic laparoscopic surgery. This study hypothesized that increasing body mass index is associated with more mechanical strain and alveolar collapse, and these impairments are exacerbated by pneumoperitoneum and Trendelenburg positioning.
METHODS:This cross-sectional study measured respiratory flow, airway pressures, and esophageal pressures in 91 subjects with body mass index ranging from 18.3 to 60.6 kg/m. Pulmonary mechanics were quantified at four stages(1) supine and level after intubation, (2) with pneumoperitoneum, (3) in Trendelenburg docked with the surgical robot, and (4) level without pneumoperitoneum. Subjects were stratified into five body mass index categories (less than 25, 25 to 29.9, 30 to 34.9, 35 to 39.9, and 40 or higher), and respiratory mechanics were compared over surgical stages using generalized estimating equations. The optimal positive end-expiratory pressure settings needed to achieve positive end-expiratory transpulmonary pressures were calculated.
RESULTS:At baseline, transpulmonary driving pressures increased in each body mass index category (1.9 ± 0.5 cm H2O; mean difference ± SD; P < 0.006), and subjects with a body mass index of 40 or higher had decreased mean end-expiratory transpulmonary pressures compared with those with body mass index of less than 25 (–7.5 ± 6.3 vs. –1.3 ± 3.4 cm H2O; P < 0.001). Pneumoperitoneum and Trendelenburg each further elevated transpulmonary driving pressures (2.8 ± 0.7 and 4.7 ± 1.0 cm H2O, respectively; P < 0.001) and depressed end-expiratory transpulmonary pressures (–3.4 ± 1.3 and –4.5 ± 1.5 cm H2O, respectively; P < 0.001) compared with baseline. Optimal positive end-expiratory pressure was greater than set positive end-expiratory pressure in 79% of subjects at baseline, 88% with pneumoperitoneum, 95% in Trendelenburg, and ranged from 0 to 36.6 cm H2O depending on body mass index and surgical stage.
CONCLUSIONS:Increasing body mass index induces significant alterations in lung mechanics during robotic laparoscopic surgery, but there is a wide range in the degree of impairment. Positive end-expiratory pressure settings may need individualization based on body mass index and surgical conditions.</description><identifier>ISSN: 0003-3022</identifier><identifier>EISSN: 1528-1175</identifier><identifier>DOI: 10.1097/ALN.0000000000003442</identifier><language>eng</language><publisher>the American Society of Anesthesiologists, Inc. Wolters Kluwer Health, Inc</publisher><ispartof>Anesthesiology (Philadelphia), 2020-10, Vol.133 (4), p.750-763</ispartof><rights>the American Society of Anesthesiologists, Inc. Wolters Kluwer Health, Inc. All Rights Reserved.</rights><rights>Copyright © by 2020, the American Society of Anesthesiologists, Inc. Wolters Kluwer Health, Inc. All Rights Reserved.</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,780,784,27924,27925</link.rule.ids></links><search><creatorcontrib>Tharp, William G.</creatorcontrib><creatorcontrib>Murphy, Serena</creatorcontrib><creatorcontrib>Breidenstein, Max W.</creatorcontrib><creatorcontrib>Love, Collin</creatorcontrib><creatorcontrib>Booms, Alisha</creatorcontrib><creatorcontrib>Rafferty, Melissa N.</creatorcontrib><creatorcontrib>Friend, Alexander F.</creatorcontrib><creatorcontrib>Perrapato, Scott</creatorcontrib><creatorcontrib>Ahern, Thomas P.</creatorcontrib><creatorcontrib>Dixon, Anne E.</creatorcontrib><creatorcontrib>Bates, Jason H. T.</creatorcontrib><creatorcontrib>Bender, S. Patrick</creatorcontrib><title>Body Habitus and Dynamic Surgical Conditions Independently Impair Pulmonary Mechanics during Robotic-assisted Laparoscopic Surgery: A Cross-sectional Study</title><title>Anesthesiology (Philadelphia)</title><description>BACKGROUND:Body habitus, pneumoperitoneum, and Trendelenburg positioning may each independently impair lung mechanics during robotic laparoscopic surgery. This study hypothesized that increasing body mass index is associated with more mechanical strain and alveolar collapse, and these impairments are exacerbated by pneumoperitoneum and Trendelenburg positioning.
METHODS:This cross-sectional study measured respiratory flow, airway pressures, and esophageal pressures in 91 subjects with body mass index ranging from 18.3 to 60.6 kg/m. Pulmonary mechanics were quantified at four stages(1) supine and level after intubation, (2) with pneumoperitoneum, (3) in Trendelenburg docked with the surgical robot, and (4) level without pneumoperitoneum. Subjects were stratified into five body mass index categories (less than 25, 25 to 29.9, 30 to 34.9, 35 to 39.9, and 40 or higher), and respiratory mechanics were compared over surgical stages using generalized estimating equations. The optimal positive end-expiratory pressure settings needed to achieve positive end-expiratory transpulmonary pressures were calculated.
RESULTS:At baseline, transpulmonary driving pressures increased in each body mass index category (1.9 ± 0.5 cm H2O; mean difference ± SD; P < 0.006), and subjects with a body mass index of 40 or higher had decreased mean end-expiratory transpulmonary pressures compared with those with body mass index of less than 25 (–7.5 ± 6.3 vs. –1.3 ± 3.4 cm H2O; P < 0.001). Pneumoperitoneum and Trendelenburg each further elevated transpulmonary driving pressures (2.8 ± 0.7 and 4.7 ± 1.0 cm H2O, respectively; P < 0.001) and depressed end-expiratory transpulmonary pressures (–3.4 ± 1.3 and –4.5 ± 1.5 cm H2O, respectively; P < 0.001) compared with baseline. Optimal positive end-expiratory pressure was greater than set positive end-expiratory pressure in 79% of subjects at baseline, 88% with pneumoperitoneum, 95% in Trendelenburg, and ranged from 0 to 36.6 cm H2O depending on body mass index and surgical stage.
CONCLUSIONS:Increasing body mass index induces significant alterations in lung mechanics during robotic laparoscopic surgery, but there is a wide range in the degree of impairment. Positive end-expiratory pressure settings may need individualization based on body mass index and surgical conditions.</description><issn>0003-3022</issn><issn>1528-1175</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2020</creationdate><recordtype>article</recordtype><sourceid/><recordid>eNp9UMtOwzAQtBBIlMIfcPAPGPzKi1sJj1YKD9Heo43jtIbUruJEVb6FnyWBHhAHVtpdzWhnRxqELhm9YjSJrmfZ8xX9VUJKfoQmLOAxYSwKjtFkZImgnJ-iM-_fBxgFIp6gz1tX9ngOhWk7j8GW-K63sDUKL7tmbRTUOHW2NK1x1uOFLfVOD8O2dY8X2x2YBr929dZZaHr8pNUGrFEel11j7Bq_ucK1RhHw3vhWlziDHTTOK7c7OOimv8EznA6kJ16r0WfwXLZd2Z-jkwpqry8Oe4pWD_erdE6yl8dFOsuI5zLhBBivgoDrRAWJFhDSqNBxnERMQCmY4BAmhSqkGGBRxFWYMB0mmspKcSlBiCmKf97uXd3qxn_U3V43-UZD3W5yRvMx43zIOP-b8SCV_0i_7wLJCaecshGQoVksvgBQnYKn</recordid><startdate>202010</startdate><enddate>202010</enddate><creator>Tharp, William G.</creator><creator>Murphy, Serena</creator><creator>Breidenstein, Max W.</creator><creator>Love, Collin</creator><creator>Booms, Alisha</creator><creator>Rafferty, Melissa N.</creator><creator>Friend, Alexander F.</creator><creator>Perrapato, Scott</creator><creator>Ahern, Thomas P.</creator><creator>Dixon, Anne E.</creator><creator>Bates, Jason H. T.</creator><creator>Bender, S. Patrick</creator><general>the American Society of Anesthesiologists, Inc. Wolters Kluwer Health, Inc</general><general>Copyright by , the American Society of Anesthesiologists, Inc. Wolters Kluwer Health, Inc</general><scope/></search><sort><creationdate>202010</creationdate><title>Body Habitus and Dynamic Surgical Conditions Independently Impair Pulmonary Mechanics during Robotic-assisted Laparoscopic Surgery: A Cross-sectional Study</title><author>Tharp, William G. ; Murphy, Serena ; Breidenstein, Max W. ; Love, Collin ; Booms, Alisha ; Rafferty, Melissa N. ; Friend, Alexander F. ; Perrapato, Scott ; Ahern, Thomas P. ; Dixon, Anne E. ; Bates, Jason H. T. ; Bender, S. Patrick</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-s2492-a12f552e9c59e3a607be889713ad3132a69bcb433adbb8f691e69e04fc244a33</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2020</creationdate><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Tharp, William G.</creatorcontrib><creatorcontrib>Murphy, Serena</creatorcontrib><creatorcontrib>Breidenstein, Max W.</creatorcontrib><creatorcontrib>Love, Collin</creatorcontrib><creatorcontrib>Booms, Alisha</creatorcontrib><creatorcontrib>Rafferty, Melissa N.</creatorcontrib><creatorcontrib>Friend, Alexander F.</creatorcontrib><creatorcontrib>Perrapato, Scott</creatorcontrib><creatorcontrib>Ahern, Thomas P.</creatorcontrib><creatorcontrib>Dixon, Anne E.</creatorcontrib><creatorcontrib>Bates, Jason H. T.</creatorcontrib><creatorcontrib>Bender, S. Patrick</creatorcontrib><jtitle>Anesthesiology (Philadelphia)</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Tharp, William G.</au><au>Murphy, Serena</au><au>Breidenstein, Max W.</au><au>Love, Collin</au><au>Booms, Alisha</au><au>Rafferty, Melissa N.</au><au>Friend, Alexander F.</au><au>Perrapato, Scott</au><au>Ahern, Thomas P.</au><au>Dixon, Anne E.</au><au>Bates, Jason H. T.</au><au>Bender, S. Patrick</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Body Habitus and Dynamic Surgical Conditions Independently Impair Pulmonary Mechanics during Robotic-assisted Laparoscopic Surgery: A Cross-sectional Study</atitle><jtitle>Anesthesiology (Philadelphia)</jtitle><date>2020-10</date><risdate>2020</risdate><volume>133</volume><issue>4</issue><spage>750</spage><epage>763</epage><pages>750-763</pages><issn>0003-3022</issn><eissn>1528-1175</eissn><abstract>BACKGROUND:Body habitus, pneumoperitoneum, and Trendelenburg positioning may each independently impair lung mechanics during robotic laparoscopic surgery. This study hypothesized that increasing body mass index is associated with more mechanical strain and alveolar collapse, and these impairments are exacerbated by pneumoperitoneum and Trendelenburg positioning.
METHODS:This cross-sectional study measured respiratory flow, airway pressures, and esophageal pressures in 91 subjects with body mass index ranging from 18.3 to 60.6 kg/m. Pulmonary mechanics were quantified at four stages(1) supine and level after intubation, (2) with pneumoperitoneum, (3) in Trendelenburg docked with the surgical robot, and (4) level without pneumoperitoneum. Subjects were stratified into five body mass index categories (less than 25, 25 to 29.9, 30 to 34.9, 35 to 39.9, and 40 or higher), and respiratory mechanics were compared over surgical stages using generalized estimating equations. The optimal positive end-expiratory pressure settings needed to achieve positive end-expiratory transpulmonary pressures were calculated.
RESULTS:At baseline, transpulmonary driving pressures increased in each body mass index category (1.9 ± 0.5 cm H2O; mean difference ± SD; P < 0.006), and subjects with a body mass index of 40 or higher had decreased mean end-expiratory transpulmonary pressures compared with those with body mass index of less than 25 (–7.5 ± 6.3 vs. –1.3 ± 3.4 cm H2O; P < 0.001). Pneumoperitoneum and Trendelenburg each further elevated transpulmonary driving pressures (2.8 ± 0.7 and 4.7 ± 1.0 cm H2O, respectively; P < 0.001) and depressed end-expiratory transpulmonary pressures (–3.4 ± 1.3 and –4.5 ± 1.5 cm H2O, respectively; P < 0.001) compared with baseline. Optimal positive end-expiratory pressure was greater than set positive end-expiratory pressure in 79% of subjects at baseline, 88% with pneumoperitoneum, 95% in Trendelenburg, and ranged from 0 to 36.6 cm H2O depending on body mass index and surgical stage.
CONCLUSIONS:Increasing body mass index induces significant alterations in lung mechanics during robotic laparoscopic surgery, but there is a wide range in the degree of impairment. Positive end-expiratory pressure settings may need individualization based on body mass index and surgical conditions.</abstract><pub>the American Society of Anesthesiologists, Inc. Wolters Kluwer Health, Inc</pub><doi>10.1097/ALN.0000000000003442</doi><tpages>14</tpages></addata></record> |
fulltext | fulltext |
identifier | ISSN: 0003-3022 |
ispartof | Anesthesiology (Philadelphia), 2020-10, Vol.133 (4), p.750-763 |
issn | 0003-3022 1528-1175 |
language | eng |
recordid | cdi_wolterskluwer_health_10_1097_ALN_0000000000003442 |
source | Journals@Ovid Complete; EZB-FREE-00999 freely available EZB journals |
title | Body Habitus and Dynamic Surgical Conditions Independently Impair Pulmonary Mechanics during Robotic-assisted Laparoscopic Surgery: A Cross-sectional Study |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2024-12-27T15%3A40%3A03IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-wolterskluwer&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Body%20Habitus%20and%20Dynamic%20Surgical%20Conditions%20Independently%20Impair%20Pulmonary%20Mechanics%20during%20Robotic-assisted%20Laparoscopic%20Surgery:%20A%20Cross-sectional%20Study&rft.jtitle=Anesthesiology%20(Philadelphia)&rft.au=Tharp,%20William%20G.&rft.date=2020-10&rft.volume=133&rft.issue=4&rft.spage=750&rft.epage=763&rft.pages=750-763&rft.issn=0003-3022&rft.eissn=1528-1175&rft_id=info:doi/10.1097/ALN.0000000000003442&rft_dat=%3Cwolterskluwer%3E00000542-202010000-00018%3C/wolterskluwer%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_id=info:pmid/&rfr_iscdi=true |