Design of an Innovative Wireless Left Ventricular Assist Device Driven by either Extracorporeal Magnets or an Intracorporeal Battery Pack

This study aimed to design a new wireless left ventricular assist device (LVAD) that solved the driveline problem of current LVADs and the heat problem of the transcutaneous energy transfer system (TETS). Our new wireless LVAD consisted of two blood pumps capable of driving using extracorporeal magn...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:ASAIO journal (1992) 2022-12
Hauptverfasser: Horie, Hideyuki, Isoyama, Takashi, Ishiyama, Kazushi
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page
container_issue
container_start_page
container_title ASAIO journal (1992)
container_volume
creator Horie, Hideyuki
Isoyama, Takashi
Ishiyama, Kazushi
description This study aimed to design a new wireless left ventricular assist device (LVAD) that solved the driveline problem of current LVADs and the heat problem of the transcutaneous energy transfer system (TETS). Our new wireless LVAD consisted of two blood pumps capable of driving using extracorporeal magnets and an intracorporeal battery pack. When one pump was driven, the other pump was stopped. The battery pack was wirelessly and slowly charged using TETS with low-power transmission, whereas the magnetic pump was driven wirelessly by extracorporeal magnets. We demonstrated the feasibility of our system in a bench-top durability test for 7 days. The distance between the extracorporeal magnets and the magnetic pump was 27.5 mm. Our LVAD system had steadily provided sufficient pressure and flow volume (approximately 108 mm Hg and 5.0 L/min, respectively) to the test loop for 7 days. Although loss of synchronism occurred once during the test, it recovered within a few minutes. The results demonstrate the feasibility of the proposed wireless LVAD system. Further technical improvements are required in our system, such as downsizing the electric devices inside the body, to conduct an in vivo test for the next step.
doi_str_mv 10.1097/MAT.0000000000001874
format Article
fullrecord <record><control><sourceid>wolterskluwer</sourceid><recordid>TN_cdi_wolterskluwer_health_00002480-990000000-00144</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>00002480-990000000-00144</sourcerecordid><originalsourceid>FETCH-wolterskluwer_health_00002480-990000000-001443</originalsourceid><addsrcrecordid>eNqdj8FOwzAQRH0AiUL5Aw77Ayl249LkWEgRSFTiUJVjZaJNY2rZaHeb0k_grwmiB7iyl5F2ZjR6Sl0ZPTK6nF4vZsuR_nWmmNoTNTB6UmTj0tycqXPmt_4_yXMzUJ8Vst9ESA24CI8xps6J7xBePGFAZnjCRmCFUcjXu-AIZsyeBSrsfI1QUZ-O8HoA9NIiwfxDyNWJ3hOhC7Bwm4jCkOhn4I9560SQDvDs6u1QnTYuMF4e9ULZ-_ny7iHbp9CHeBt2e6R12_ekXX-jjW2hs7I8gmY9k7X5P2tfF-BhBQ</addsrcrecordid><sourcetype>Publisher</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype></control><display><type>article</type><title>Design of an Innovative Wireless Left Ventricular Assist Device Driven by either Extracorporeal Magnets or an Intracorporeal Battery Pack</title><source>Journals@Ovid LWW Legacy Archive</source><source>Elektronische Zeitschriftenbibliothek - Frei zugängliche E-Journals</source><creator>Horie, Hideyuki ; Isoyama, Takashi ; Ishiyama, Kazushi</creator><creatorcontrib>Horie, Hideyuki ; Isoyama, Takashi ; Ishiyama, Kazushi</creatorcontrib><description>This study aimed to design a new wireless left ventricular assist device (LVAD) that solved the driveline problem of current LVADs and the heat problem of the transcutaneous energy transfer system (TETS). Our new wireless LVAD consisted of two blood pumps capable of driving using extracorporeal magnets and an intracorporeal battery pack. When one pump was driven, the other pump was stopped. The battery pack was wirelessly and slowly charged using TETS with low-power transmission, whereas the magnetic pump was driven wirelessly by extracorporeal magnets. We demonstrated the feasibility of our system in a bench-top durability test for 7 days. The distance between the extracorporeal magnets and the magnetic pump was 27.5 mm. Our LVAD system had steadily provided sufficient pressure and flow volume (approximately 108 mm Hg and 5.0 L/min, respectively) to the test loop for 7 days. Although loss of synchronism occurred once during the test, it recovered within a few minutes. The results demonstrate the feasibility of the proposed wireless LVAD system. Further technical improvements are required in our system, such as downsizing the electric devices inside the body, to conduct an in vivo test for the next step.</description><identifier>ISSN: 1058-2916</identifier><identifier>DOI: 10.1097/MAT.0000000000001874</identifier><language>eng</language><publisher>Lippincott Williams &amp; Wilkins</publisher><ispartof>ASAIO journal (1992), 2022-12</ispartof><rights>Lippincott Williams &amp; Wilkins</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktohtml>$$Uhttp://ovidsp.ovid.com/ovidweb.cgi?T=JS&amp;NEWS=n&amp;CSC=Y&amp;PAGE=fulltext&amp;D=ovft&amp;AN=00002480-990000000-00144$$EHTML$$P50$$Gwolterskluwer$$H</linktohtml><link.rule.ids>314,776,780,4595,27901,27902,65206</link.rule.ids></links><search><creatorcontrib>Horie, Hideyuki</creatorcontrib><creatorcontrib>Isoyama, Takashi</creatorcontrib><creatorcontrib>Ishiyama, Kazushi</creatorcontrib><title>Design of an Innovative Wireless Left Ventricular Assist Device Driven by either Extracorporeal Magnets or an Intracorporeal Battery Pack</title><title>ASAIO journal (1992)</title><description>This study aimed to design a new wireless left ventricular assist device (LVAD) that solved the driveline problem of current LVADs and the heat problem of the transcutaneous energy transfer system (TETS). Our new wireless LVAD consisted of two blood pumps capable of driving using extracorporeal magnets and an intracorporeal battery pack. When one pump was driven, the other pump was stopped. The battery pack was wirelessly and slowly charged using TETS with low-power transmission, whereas the magnetic pump was driven wirelessly by extracorporeal magnets. We demonstrated the feasibility of our system in a bench-top durability test for 7 days. The distance between the extracorporeal magnets and the magnetic pump was 27.5 mm. Our LVAD system had steadily provided sufficient pressure and flow volume (approximately 108 mm Hg and 5.0 L/min, respectively) to the test loop for 7 days. Although loss of synchronism occurred once during the test, it recovered within a few minutes. The results demonstrate the feasibility of the proposed wireless LVAD system. Further technical improvements are required in our system, such as downsizing the electric devices inside the body, to conduct an in vivo test for the next step.</description><issn>1058-2916</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2022</creationdate><recordtype>article</recordtype><sourceid/><recordid>eNqdj8FOwzAQRH0AiUL5Aw77Ayl249LkWEgRSFTiUJVjZaJNY2rZaHeb0k_grwmiB7iyl5F2ZjR6Sl0ZPTK6nF4vZsuR_nWmmNoTNTB6UmTj0tycqXPmt_4_yXMzUJ8Vst9ESA24CI8xps6J7xBePGFAZnjCRmCFUcjXu-AIZsyeBSrsfI1QUZ-O8HoA9NIiwfxDyNWJ3hOhC7Bwm4jCkOhn4I9560SQDvDs6u1QnTYuMF4e9ULZ-_ny7iHbp9CHeBt2e6R12_ekXX-jjW2hs7I8gmY9k7X5P2tfF-BhBQ</recordid><startdate>20221221</startdate><enddate>20221221</enddate><creator>Horie, Hideyuki</creator><creator>Isoyama, Takashi</creator><creator>Ishiyama, Kazushi</creator><general>Lippincott Williams &amp; Wilkins</general><scope/></search><sort><creationdate>20221221</creationdate><title>Design of an Innovative Wireless Left Ventricular Assist Device Driven by either Extracorporeal Magnets or an Intracorporeal Battery Pack</title><author>Horie, Hideyuki ; Isoyama, Takashi ; Ishiyama, Kazushi</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-wolterskluwer_health_00002480-990000000-001443</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2022</creationdate><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Horie, Hideyuki</creatorcontrib><creatorcontrib>Isoyama, Takashi</creatorcontrib><creatorcontrib>Ishiyama, Kazushi</creatorcontrib><jtitle>ASAIO journal (1992)</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Horie, Hideyuki</au><au>Isoyama, Takashi</au><au>Ishiyama, Kazushi</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Design of an Innovative Wireless Left Ventricular Assist Device Driven by either Extracorporeal Magnets or an Intracorporeal Battery Pack</atitle><jtitle>ASAIO journal (1992)</jtitle><date>2022-12-21</date><risdate>2022</risdate><issn>1058-2916</issn><abstract>This study aimed to design a new wireless left ventricular assist device (LVAD) that solved the driveline problem of current LVADs and the heat problem of the transcutaneous energy transfer system (TETS). Our new wireless LVAD consisted of two blood pumps capable of driving using extracorporeal magnets and an intracorporeal battery pack. When one pump was driven, the other pump was stopped. The battery pack was wirelessly and slowly charged using TETS with low-power transmission, whereas the magnetic pump was driven wirelessly by extracorporeal magnets. We demonstrated the feasibility of our system in a bench-top durability test for 7 days. The distance between the extracorporeal magnets and the magnetic pump was 27.5 mm. Our LVAD system had steadily provided sufficient pressure and flow volume (approximately 108 mm Hg and 5.0 L/min, respectively) to the test loop for 7 days. Although loss of synchronism occurred once during the test, it recovered within a few minutes. The results demonstrate the feasibility of the proposed wireless LVAD system. Further technical improvements are required in our system, such as downsizing the electric devices inside the body, to conduct an in vivo test for the next step.</abstract><pub>Lippincott Williams &amp; Wilkins</pub><doi>10.1097/MAT.0000000000001874</doi></addata></record>
fulltext fulltext
identifier ISSN: 1058-2916
ispartof ASAIO journal (1992), 2022-12
issn 1058-2916
language eng
recordid cdi_wolterskluwer_health_00002480-990000000-00144
source Journals@Ovid LWW Legacy Archive; Elektronische Zeitschriftenbibliothek - Frei zugängliche E-Journals
title Design of an Innovative Wireless Left Ventricular Assist Device Driven by either Extracorporeal Magnets or an Intracorporeal Battery Pack
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-02-03T16%3A44%3A43IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-wolterskluwer&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Design%20of%20an%20Innovative%20Wireless%20Left%20Ventricular%20Assist%20Device%20Driven%20by%20either%20Extracorporeal%20Magnets%20or%20an%20Intracorporeal%20Battery%20Pack&rft.jtitle=ASAIO%20journal%20(1992)&rft.au=Horie,%20Hideyuki&rft.date=2022-12-21&rft.issn=1058-2916&rft_id=info:doi/10.1097/MAT.0000000000001874&rft_dat=%3Cwolterskluwer%3E00002480-990000000-00144%3C/wolterskluwer%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_id=info:pmid/&rfr_iscdi=true