SU‐E‐T‐151: Effect of Surface Dose and Depth of Maximum Dose with Physical Wedge Filters for 6MV Flattened and 7MV Unflattened Photon Beams

Purpose: To evaluate impact on surface dose, Depth of dose maximum (dmax) and depth dose distribution due to physical wedge filters for different wedge angle were studied for 6MV flattened (6MV FB) and 7MV unflattened (7MV UFB) photon beams. Methods: The flattening filter and primary collimator are...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Medical Physics 2012-06, Vol.39 (6), p.3737-3738
Hauptverfasser: Ashokkumar, S, Sinha, SN, Yadav, Girigesh, Raman, Kothanda, Thiyagarajan, Rajesh, Nambiraj, N Arunai
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 3738
container_issue 6
container_start_page 3737
container_title Medical Physics
container_volume 39
creator Ashokkumar, S
Sinha, SN
Yadav, Girigesh
Raman, Kothanda
Thiyagarajan, Rajesh
Nambiraj, N Arunai
description Purpose: To evaluate impact on surface dose, Depth of dose maximum (dmax) and depth dose distribution due to physical wedge filters for different wedge angle were studied for 6MV flattened (6MV FB) and 7MV unflattened (7MV UFB) photon beams. Methods: The flattening filter and primary collimator are the major sources of producing the scattered radiation and these parameters affect the surface dose, dmax and dose distribution. In this study, open fields surface dose, dmax and depth dose distribution values were compared with physical wedge filter for 6MV FB and 7MV UFB .The measurement carried out in Siemens ‐ ARTISTE linear accelerator with diode detector along the central axis of the beam at 100 cm source to surface distance using IBA blue water phantom for 6MV FB and 7MV UFB x‐ray beams. Results: The surface dose increased as the field size increased for open and physical wedge fields for 6MV FB and 7MV UFB beams. For open fields, Surface doses relative to the dose at dmax ranged from 0.443 to 0.569 and 0.463 to 0.668 for field sizes of 5 × 5 to 20×20 cm2 for the 7MV UFB and 6MV FB beam respectively. The measured surface dose for 150, 300,450 and 600 wedge field values are 0.396 to 0.504, 0.366 to 0.484, 0.342 to 0.464 and 0.347 to 0.47 respectively for 7MV UFB and 0.424 to 0.566, 0.398 to 0.555, 0.3860 to 0.5430 and 0.389 to 0.55 respectively for 6MV FB. Conclusions: We found that dmax of wedged beams were higher than those open beams for field size up to 10 × 10 cm2, Surface doses of wedged beams were lower than those of open beams for 7MV UFB and 6MV FB. Surface dose of the 7MV UFB were lower than the 6MV FB for open and wedged beams.
doi_str_mv 10.1118/1.4735209
format Article
fullrecord <record><control><sourceid>proquest_wiley</sourceid><recordid>TN_cdi_wiley_primary_10_1118_1_4735209_MP5209</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>1900127955</sourcerecordid><originalsourceid>FETCH-LOGICAL-c2339-3dc04973796d20708da4708b82c8e1442e3556eca6311d0ca4ccda9a02dc80ac3</originalsourceid><addsrcrecordid>eNp9kc1u1DAQxy0EokvhwAsgHwEp7fgrH9xKuwuVumKlduEYufaENUriJXZU9sYj0Ffsk-CSpTc4jD36z08_Sx5CXjI4YoyVx-xIFkJxqB6RGU9tJlP_mMwAKplxCeqAPAvhGwDkQsFTcsBLxYoSyhm5vVzf_fw1T3WViin2js6bBk2kvqGX49Bog_TMB6S6t_QMt3FzP1nqH64bu2ly41K42uyCM7qlX9B-RbpwbcQh0MYPNF9-potWx4g92j-eIiXrvnnIVhsffU_fo-7Cc_Kk0W3AF_v7kKwX86vTj9nFpw_npycXmeFCVJmwBmRViKLKLYcCSqtlOq9LbkpkUnIUSuVodC4Ys2C0NMbqSgO3pgRtxCF5PXm3g_8-Yoh154LBttU9-jHUrAJgvKiUSuibCTWDD2HApt4OrtPDrmZQ32-gZvV-A4l9tdeO1x3aB_Lvlycgm4Ab1-Lu36Z6udoL3058MC7q6Hz_n9d_A5tbm_Q</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>1900127955</pqid></control><display><type>article</type><title>SU‐E‐T‐151: Effect of Surface Dose and Depth of Maximum Dose with Physical Wedge Filters for 6MV Flattened and 7MV Unflattened Photon Beams</title><source>Wiley Journals</source><source>Alma/SFX Local Collection</source><creator>Ashokkumar, S ; Sinha, SN ; Yadav, Girigesh ; Raman, Kothanda ; Thiyagarajan, Rajesh ; Nambiraj, N Arunai</creator><creatorcontrib>Ashokkumar, S ; Sinha, SN ; Yadav, Girigesh ; Raman, Kothanda ; Thiyagarajan, Rajesh ; Nambiraj, N Arunai</creatorcontrib><description>Purpose: To evaluate impact on surface dose, Depth of dose maximum (dmax) and depth dose distribution due to physical wedge filters for different wedge angle were studied for 6MV flattened (6MV FB) and 7MV unflattened (7MV UFB) photon beams. Methods: The flattening filter and primary collimator are the major sources of producing the scattered radiation and these parameters affect the surface dose, dmax and dose distribution. In this study, open fields surface dose, dmax and depth dose distribution values were compared with physical wedge filter for 6MV FB and 7MV UFB .The measurement carried out in Siemens ‐ ARTISTE linear accelerator with diode detector along the central axis of the beam at 100 cm source to surface distance using IBA blue water phantom for 6MV FB and 7MV UFB x‐ray beams. Results: The surface dose increased as the field size increased for open and physical wedge fields for 6MV FB and 7MV UFB beams. For open fields, Surface doses relative to the dose at dmax ranged from 0.443 to 0.569 and 0.463 to 0.668 for field sizes of 5 × 5 to 20×20 cm2 for the 7MV UFB and 6MV FB beam respectively. The measured surface dose for 150, 300,450 and 600 wedge field values are 0.396 to 0.504, 0.366 to 0.484, 0.342 to 0.464 and 0.347 to 0.47 respectively for 7MV UFB and 0.424 to 0.566, 0.398 to 0.555, 0.3860 to 0.5430 and 0.389 to 0.55 respectively for 6MV FB. Conclusions: We found that dmax of wedged beams were higher than those open beams for field size up to 10 × 10 cm2, Surface doses of wedged beams were lower than those of open beams for 7MV UFB and 6MV FB. Surface dose of the 7MV UFB were lower than the 6MV FB for open and wedged beams.</description><identifier>ISSN: 0094-2405</identifier><identifier>EISSN: 2473-4209</identifier><identifier>DOI: 10.1118/1.4735209</identifier><identifier>PMID: 28517808</identifier><identifier>CODEN: MPHYA6</identifier><language>eng</language><publisher>United States: American Association of Physicists in Medicine</publisher><subject>Collimators ; Field size ; Linear accelerators ; Particle beam detectors ; Photons ; Physical wedges ; Surface measurements ; Surface scattering</subject><ispartof>Medical Physics, 2012-06, Vol.39 (6), p.3737-3738</ispartof><rights>American Association of Physicists in Medicine</rights><rights>2012 American Association of Physicists in Medicine</rights><rights>2012 American Association of Physicists in Medicine.</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktohtml>$$Uhttps://onlinelibrary.wiley.com/doi/full/10.1118%2F1.4735209$$EHTML$$P50$$Gwiley$$H</linktohtml><link.rule.ids>309,310,314,780,784,789,790,1417,23930,23931,25140,27924,27925,45575</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/28517808$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Ashokkumar, S</creatorcontrib><creatorcontrib>Sinha, SN</creatorcontrib><creatorcontrib>Yadav, Girigesh</creatorcontrib><creatorcontrib>Raman, Kothanda</creatorcontrib><creatorcontrib>Thiyagarajan, Rajesh</creatorcontrib><creatorcontrib>Nambiraj, N Arunai</creatorcontrib><title>SU‐E‐T‐151: Effect of Surface Dose and Depth of Maximum Dose with Physical Wedge Filters for 6MV Flattened and 7MV Unflattened Photon Beams</title><title>Medical Physics</title><addtitle>Med Phys</addtitle><description>Purpose: To evaluate impact on surface dose, Depth of dose maximum (dmax) and depth dose distribution due to physical wedge filters for different wedge angle were studied for 6MV flattened (6MV FB) and 7MV unflattened (7MV UFB) photon beams. Methods: The flattening filter and primary collimator are the major sources of producing the scattered radiation and these parameters affect the surface dose, dmax and dose distribution. In this study, open fields surface dose, dmax and depth dose distribution values were compared with physical wedge filter for 6MV FB and 7MV UFB .The measurement carried out in Siemens ‐ ARTISTE linear accelerator with diode detector along the central axis of the beam at 100 cm source to surface distance using IBA blue water phantom for 6MV FB and 7MV UFB x‐ray beams. Results: The surface dose increased as the field size increased for open and physical wedge fields for 6MV FB and 7MV UFB beams. For open fields, Surface doses relative to the dose at dmax ranged from 0.443 to 0.569 and 0.463 to 0.668 for field sizes of 5 × 5 to 20×20 cm2 for the 7MV UFB and 6MV FB beam respectively. The measured surface dose for 150, 300,450 and 600 wedge field values are 0.396 to 0.504, 0.366 to 0.484, 0.342 to 0.464 and 0.347 to 0.47 respectively for 7MV UFB and 0.424 to 0.566, 0.398 to 0.555, 0.3860 to 0.5430 and 0.389 to 0.55 respectively for 6MV FB. Conclusions: We found that dmax of wedged beams were higher than those open beams for field size up to 10 × 10 cm2, Surface doses of wedged beams were lower than those of open beams for 7MV UFB and 6MV FB. Surface dose of the 7MV UFB were lower than the 6MV FB for open and wedged beams.</description><subject>Collimators</subject><subject>Field size</subject><subject>Linear accelerators</subject><subject>Particle beam detectors</subject><subject>Photons</subject><subject>Physical wedges</subject><subject>Surface measurements</subject><subject>Surface scattering</subject><issn>0094-2405</issn><issn>2473-4209</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2012</creationdate><recordtype>article</recordtype><recordid>eNp9kc1u1DAQxy0EokvhwAsgHwEp7fgrH9xKuwuVumKlduEYufaENUriJXZU9sYj0Ffsk-CSpTc4jD36z08_Sx5CXjI4YoyVx-xIFkJxqB6RGU9tJlP_mMwAKplxCeqAPAvhGwDkQsFTcsBLxYoSyhm5vVzf_fw1T3WViin2js6bBk2kvqGX49Bog_TMB6S6t_QMt3FzP1nqH64bu2ly41K42uyCM7qlX9B-RbpwbcQh0MYPNF9-potWx4g92j-eIiXrvnnIVhsffU_fo-7Cc_Kk0W3AF_v7kKwX86vTj9nFpw_npycXmeFCVJmwBmRViKLKLYcCSqtlOq9LbkpkUnIUSuVodC4Ys2C0NMbqSgO3pgRtxCF5PXm3g_8-Yoh154LBttU9-jHUrAJgvKiUSuibCTWDD2HApt4OrtPDrmZQ32-gZvV-A4l9tdeO1x3aB_Lvlycgm4Ab1-Lu36Z6udoL3058MC7q6Hz_n9d_A5tbm_Q</recordid><startdate>201206</startdate><enddate>201206</enddate><creator>Ashokkumar, S</creator><creator>Sinha, SN</creator><creator>Yadav, Girigesh</creator><creator>Raman, Kothanda</creator><creator>Thiyagarajan, Rajesh</creator><creator>Nambiraj, N Arunai</creator><general>American Association of Physicists in Medicine</general><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>201206</creationdate><title>SU‐E‐T‐151: Effect of Surface Dose and Depth of Maximum Dose with Physical Wedge Filters for 6MV Flattened and 7MV Unflattened Photon Beams</title><author>Ashokkumar, S ; Sinha, SN ; Yadav, Girigesh ; Raman, Kothanda ; Thiyagarajan, Rajesh ; Nambiraj, N Arunai</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c2339-3dc04973796d20708da4708b82c8e1442e3556eca6311d0ca4ccda9a02dc80ac3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2012</creationdate><topic>Collimators</topic><topic>Field size</topic><topic>Linear accelerators</topic><topic>Particle beam detectors</topic><topic>Photons</topic><topic>Physical wedges</topic><topic>Surface measurements</topic><topic>Surface scattering</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Ashokkumar, S</creatorcontrib><creatorcontrib>Sinha, SN</creatorcontrib><creatorcontrib>Yadav, Girigesh</creatorcontrib><creatorcontrib>Raman, Kothanda</creatorcontrib><creatorcontrib>Thiyagarajan, Rajesh</creatorcontrib><creatorcontrib>Nambiraj, N Arunai</creatorcontrib><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>Medical Physics</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Ashokkumar, S</au><au>Sinha, SN</au><au>Yadav, Girigesh</au><au>Raman, Kothanda</au><au>Thiyagarajan, Rajesh</au><au>Nambiraj, N Arunai</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>SU‐E‐T‐151: Effect of Surface Dose and Depth of Maximum Dose with Physical Wedge Filters for 6MV Flattened and 7MV Unflattened Photon Beams</atitle><jtitle>Medical Physics</jtitle><addtitle>Med Phys</addtitle><date>2012-06</date><risdate>2012</risdate><volume>39</volume><issue>6</issue><spage>3737</spage><epage>3738</epage><pages>3737-3738</pages><issn>0094-2405</issn><eissn>2473-4209</eissn><coden>MPHYA6</coden><abstract>Purpose: To evaluate impact on surface dose, Depth of dose maximum (dmax) and depth dose distribution due to physical wedge filters for different wedge angle were studied for 6MV flattened (6MV FB) and 7MV unflattened (7MV UFB) photon beams. Methods: The flattening filter and primary collimator are the major sources of producing the scattered radiation and these parameters affect the surface dose, dmax and dose distribution. In this study, open fields surface dose, dmax and depth dose distribution values were compared with physical wedge filter for 6MV FB and 7MV UFB .The measurement carried out in Siemens ‐ ARTISTE linear accelerator with diode detector along the central axis of the beam at 100 cm source to surface distance using IBA blue water phantom for 6MV FB and 7MV UFB x‐ray beams. Results: The surface dose increased as the field size increased for open and physical wedge fields for 6MV FB and 7MV UFB beams. For open fields, Surface doses relative to the dose at dmax ranged from 0.443 to 0.569 and 0.463 to 0.668 for field sizes of 5 × 5 to 20×20 cm2 for the 7MV UFB and 6MV FB beam respectively. The measured surface dose for 150, 300,450 and 600 wedge field values are 0.396 to 0.504, 0.366 to 0.484, 0.342 to 0.464 and 0.347 to 0.47 respectively for 7MV UFB and 0.424 to 0.566, 0.398 to 0.555, 0.3860 to 0.5430 and 0.389 to 0.55 respectively for 6MV FB. Conclusions: We found that dmax of wedged beams were higher than those open beams for field size up to 10 × 10 cm2, Surface doses of wedged beams were lower than those of open beams for 7MV UFB and 6MV FB. Surface dose of the 7MV UFB were lower than the 6MV FB for open and wedged beams.</abstract><cop>United States</cop><pub>American Association of Physicists in Medicine</pub><pmid>28517808</pmid><doi>10.1118/1.4735209</doi><tpages>2</tpages></addata></record>
fulltext fulltext
identifier ISSN: 0094-2405
ispartof Medical Physics, 2012-06, Vol.39 (6), p.3737-3738
issn 0094-2405
2473-4209
language eng
recordid cdi_wiley_primary_10_1118_1_4735209_MP5209
source Wiley Journals; Alma/SFX Local Collection
subjects Collimators
Field size
Linear accelerators
Particle beam detectors
Photons
Physical wedges
Surface measurements
Surface scattering
title SU‐E‐T‐151: Effect of Surface Dose and Depth of Maximum Dose with Physical Wedge Filters for 6MV Flattened and 7MV Unflattened Photon Beams
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-04T14%3A30%3A51IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_wiley&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=SU%E2%80%90E%E2%80%90T%E2%80%90151:%20Effect%20of%20Surface%20Dose%20and%20Depth%20of%20Maximum%20Dose%20with%20Physical%20Wedge%20Filters%20for%206MV%20Flattened%20and%207MV%20Unflattened%20Photon%20Beams&rft.jtitle=Medical%20Physics&rft.au=Ashokkumar,%20S&rft.date=2012-06&rft.volume=39&rft.issue=6&rft.spage=3737&rft.epage=3738&rft.pages=3737-3738&rft.issn=0094-2405&rft.eissn=2473-4209&rft.coden=MPHYA6&rft_id=info:doi/10.1118/1.4735209&rft_dat=%3Cproquest_wiley%3E1900127955%3C/proquest_wiley%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=1900127955&rft_id=info:pmid/28517808&rfr_iscdi=true