Accuracy of Dexcom G6 Continuous Glucose Monitoring in Non-Critically Ill Hospitalized Patients With Diabetes
OBJECTIVE Advances in continuous glucose monitoring (CGM) have transformed ambulatory diabetes management. Until recently, inpatient use of CGM has remained investigational, with limited data on its accuracy in the hospital setting. RESEARCH DESIGN AND METHODS To analyze the accuracy of Dexcom G6, w...
Gespeichert in:
Veröffentlicht in: | Diabetes care 2021-07, Vol.44 (7), p.1641-1646 |
---|---|
Hauptverfasser: | , , , , , , , , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | OBJECTIVE Advances in continuous glucose monitoring (CGM) have transformed ambulatory diabetes management. Until recently, inpatient use of CGM has remained investigational, with limited data on its accuracy in the hospital setting. RESEARCH DESIGN AND METHODS To analyze the accuracy of Dexcom G6, we compared retrospective matched-pair CGM and capillary point-of-care (POC) glucose data from three inpatient CGM studies (two interventional and one observational) in general medicine and surgery patients with diabetes treated with insulin. Analysis of accuracy metrics included mean absolute relative difference (MARD), median absolute relative difference (ARD), and proportion of CGM values within 15, 20, and 30% or 15, 20, and 30 mg/dL of POC reference values for blood glucose >100 mg/dL or |
---|---|