Evaluation of the Cue Health point-of-care COVID-19 (SARS-CoV-2 nucleic acid amplification) test at a community drive through collection center
Point-of-care (POC) tests are in high demand in order to facilitate rapid care decisions for patients suspected of SARS-CoV-2. We conducted a clinical validation study of the Cue Health POC nucleic acid amplification test (NAAT) using the Cue lower nasal swab, compared to a reference NAAT using stan...
Gespeichert in:
Veröffentlicht in: | Diagnostic microbiology and infectious disease 2021-05, Vol.100 (1), p.115307-115307, Article 115307 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Point-of-care (POC) tests are in high demand in order to facilitate rapid care decisions for patients suspected of SARS-CoV-2. We conducted a clinical validation study of the Cue Health POC nucleic acid amplification test (NAAT) using the Cue lower nasal swab, compared to a reference NAAT using standard nasopharyngeal swab, in 292 symptomatic and asymptomatic outpatients for SARS-CoV-2 detection in a community drive through collection setting. Positive percent agreement between Cue COVID-19 and reference SARS-CoV-2 test was 91.7% (22 of 24); or 95.7% (22 of 23) when one patient with no tie-breaker method was excluded. Negative percent agreement was 98.4% (239 of 243), and there were 25 (8.6%) invalid or canceled results. The Cue COVID-19 test demonstrated very good positive and negative percent agreement with central laboratory tests and will be useful in settings where accurate POC testing is needed to facilitate management of patients suspected of COVID-19. |
---|---|
ISSN: | 0732-8893 1879-0070 |