Field evaluation of a rapid antigen test (Panbio (TM) COVID-19 Ag Rapid Test Device) for COVID-19 diagnosis in primary healthcare centres
Objectives: To our knowledge no previous study has assessed the performance of a rapid antigen diagnostic immunoassay (RAD) conducted at the point of care (POC). We evaluated the Panbio (TM) COVID-19 Ag Rapid Test Device for diagnosis of coronavirus 2019 disease (COVID-19) in symptomatic patients (n...
Gespeichert in:
Veröffentlicht in: | Clinical microbiology and infection 2021-03, Vol.27 (3) |
---|---|
Hauptverfasser: | , , , , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Objectives: To our knowledge no previous study has assessed the performance of a rapid antigen diagnostic immunoassay (RAD) conducted at the point of care (POC). We evaluated the Panbio (TM) COVID-19 Ag Rapid Test Device for diagnosis of coronavirus 2019 disease (COVID-19) in symptomatic patients (n = 412) attending primary healthcare centres.
Methods: RAD was performed immediately after sampling following the manufacturer's instructions (reading at 15 min). RT-PCRs were carried out within 24 h of specimen collection. Samples displaying discordant results were processed for culture in Vero E6 cells. Presence of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) in cell cultures was confirmed by RT-PCR.
Results: Out of 412 patients, 43 (10.4%) tested positive by RT-PCR and RAD, and 358 (86.9%) tested negative by both methods; discordant results (RT-PCR+/RAD-) were obtained in 11 patients (2.7%). Overall specificity and sensitivity of rapid antigen detection (RAD) was 100% (95%CI 98.7-100%) and 79.6% (95%CI 67.0-88.8%), respectively, taking RT-PCR as the reference. Overall RAD negative predictive value for an estimated prevalence of 5% and 10% was 99% (95%CI 97.4-99.6%) and 97.9% (95%CI 95.9 -98.9), respectively. SARS-CoV-2 could not be cultured from specimens yielding RT-PCR+/RAD- results (n = 11).
Conclusion: The Panbio (TM) COVID-19 Ag Rapid Test Device performed well as a POC test for early diagnosis of COVID-19 in primary healthcare centres. More crucially, the data suggested that patients with RT-PCRproven COVID-19 testing negative by RAD are unlikely to be infectious. (C) 2020 European Society of Clinical Microbiology and Infectious Diseases. Published by Elsevier Ltd. All rights reserved. |
---|---|
ISSN: | 1198-743X 1469-0691 |
DOI: | 10.1016/j.cmi.2020.11.004 |