A versatile technique for in situ investigation of the effect of thin film cracking on gas permeation of coated flexible polymers

In this work, we report the development of a measurement chamber linked with a quadrupole mass spectrometer (QMS) for in situ investigation of the effect of thin film cracking on the gas permeation of coated flexible polymeric substrates. The chamber enables the establishment of a bulged state of th...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Review of scientific instruments 2021-01, Vol.92 (1), p.015120-015120, Article 015120
Hauptverfasser: Kovács, Réka Lilla, Langer, Gábor, Gyöngyösi, Szilvia, Erdélyi, Zoltán
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:In this work, we report the development of a measurement chamber linked with a quadrupole mass spectrometer (QMS) for in situ investigation of the effect of thin film cracking on the gas permeation of coated flexible polymeric substrates. The chamber enables the establishment of a bulged state of the substrate/coating system, which causes the cracking of the coating layer. The increase in the gas permeation rate due to the presence of cracks can be monitored precisely using the QMS without movement or re-clamping of the samples between each measurement step. This method eliminates the probability of uncontrollable mechanical changes in the sample, and with the mass spectrometer, high sensitivity, reliability, and reproducibility of the experimental data become available.
ISSN:0034-6748
1089-7623
DOI:10.1063/5.0028783