Improving patient care and accuracy of given doses in radiation therapy using in vivo dosimetry verification

Objective This work aims to verify and improve the dose given for cancer patients in radiation therapy by using diodes to enhance patient in vivo dosimetry on a routine basis. Some characteristics of two available semi-conductor diode dosimetry systems were evaluated. Methods The diodes had been cal...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Oncology and translational medicine 2015-10, Vol.1 (5), p.212-217
Hauptverfasser: Shawata, Ahmed Shawky, El Nimr, Tarek, Elshahat, Khaled M.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 217
container_issue 5
container_start_page 212
container_title Oncology and translational medicine
container_volume 1
creator Shawata, Ahmed Shawky
El Nimr, Tarek
Elshahat, Khaled M.
description Objective This work aims to verify and improve the dose given for cancer patients in radiation therapy by using diodes to enhance patient in vivo dosimetry on a routine basis. Some characteristics of two available semi-conductor diode dosimetry systems were evaluated. Methods The diodes had been calibrated to read the dose at Dmax below the surface. Correction factors of clinical relevance were quantified to convert the diode readings into patient dose. The diode was irradiated at various gantry angles (increments of 45~), various Field Sizes and various Source to Surface Distances (SSDs). Results The maximal response variation in the angular response with respect to an arbitrary angle of 0~ was 1.9%, and the minimum variation was 0.5%. The response of the diode with respect to various field siz- es showed the minimum and the maximum variations in the measured dose from the diode; the calculated doses were -1.6% (for 5 cm x 5 cm field size) and 6.6% (for 40 cm x 40 cm field size). The diode exhibited a significant perturbation in the response, which decreased with increasing SSD. No discrepancies larger than 5% were detected between the expected dose and the measured dose. Conclusion The results indicate that the diodes exhibit excellent linearity, dose reproducibility and minimal anisotropy; that they can be used with confidence for patient dose verification. Furthermore, diodes render real time verification of the dose delivered to patients.
doi_str_mv 10.1007/s10330-014-0035-y
format Article
fullrecord <record><control><sourceid>wanfang_jour_cross</sourceid><recordid>TN_cdi_wanfang_journals_dgyx201505004</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><cqvip_id>666494794</cqvip_id><wanfj_id>dgyx201505004</wanfj_id><sourcerecordid>dgyx201505004</sourcerecordid><originalsourceid>FETCH-LOGICAL-c1424-880be1ce4e56329a86b213e1db1790627b06d225d155c825805d870db803acc93</originalsourceid><addsrcrecordid>eNo9kF1PwyAUhhujicvcD_COmHhZPVCg9NIsfixZ4o1eEwq0Y9loha3afy91i1echOc9H0-W3WJ4wADlY8RQFJADpjlAwfLxIpuRqmI5E0xcphpSXXGCr7NFjFsAIBUtOSOzbLfa96EbnG9Rrw7O-gPSKlikvEFK62NQekRdg1o3WI9MF21EzqOgjEt459FhY4PqR3SMU4_0Nbihm0C3t4cwosEG1zj9B99kV43aRbs4v_Ps8-X5Y_mWr99fV8unda4xJTQXAmqLtaWW8YJUSvCa4MJiU-OyAk7KGrghhBnMmBaECWBGlGBqAUXauSrm2f2p77fyjfKt3HbH4NNEadrxhwBmwABo4vCJ06GLMdhG9sHtVRglBjmZlSezMpmVk1k5pszdObPpfPuVjv4Pcc5p8lrR4hfmkXfr</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype></control><display><type>article</type><title>Improving patient care and accuracy of given doses in radiation therapy using in vivo dosimetry verification</title><source>DOAJ Directory of Open Access Journals</source><creator>Shawata, Ahmed Shawky ; El Nimr, Tarek ; Elshahat, Khaled M.</creator><creatorcontrib>Shawata, Ahmed Shawky ; El Nimr, Tarek ; Elshahat, Khaled M.</creatorcontrib><description>Objective This work aims to verify and improve the dose given for cancer patients in radiation therapy by using diodes to enhance patient in vivo dosimetry on a routine basis. Some characteristics of two available semi-conductor diode dosimetry systems were evaluated. Methods The diodes had been calibrated to read the dose at Dmax below the surface. Correction factors of clinical relevance were quantified to convert the diode readings into patient dose. The diode was irradiated at various gantry angles (increments of 45~), various Field Sizes and various Source to Surface Distances (SSDs). Results The maximal response variation in the angular response with respect to an arbitrary angle of 0~ was 1.9%, and the minimum variation was 0.5%. The response of the diode with respect to various field siz- es showed the minimum and the maximum variations in the measured dose from the diode; the calculated doses were -1.6% (for 5 cm x 5 cm field size) and 6.6% (for 40 cm x 40 cm field size). The diode exhibited a significant perturbation in the response, which decreased with increasing SSD. No discrepancies larger than 5% were detected between the expected dose and the measured dose. Conclusion The results indicate that the diodes exhibit excellent linearity, dose reproducibility and minimal anisotropy; that they can be used with confidence for patient dose verification. Furthermore, diodes render real time verification of the dose delivered to patients.</description><identifier>ISSN: 2095-9621</identifier><identifier>EISSN: 2995-5858</identifier><identifier>DOI: 10.1007/s10330-014-0035-y</identifier><language>eng</language><publisher>Radiation Oncology Center, Tanta Military Hospital, Egypt%Department of Physics, Faculty of Science, Tanta University, Egypt%Department of Clinical Oncology, Faculty of Medicine, Al Azhar University, Egypt</publisher><subject>临床 ; 化疗 ; 癌症患者 ; 肿瘤学 ; 诊断</subject><ispartof>Oncology and translational medicine, 2015-10, Vol.1 (5), p.212-217</ispartof><rights>Copyright © Wanfang Data Co. Ltd. All Rights Reserved.</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><cites>FETCH-LOGICAL-c1424-880be1ce4e56329a86b213e1db1790627b06d225d155c825805d870db803acc93</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Uhttp://image.cqvip.com/vip1000/qk/90333B/90333B.jpg</thumbnail><link.rule.ids>314,780,784,864,27924,27925</link.rule.ids></links><search><creatorcontrib>Shawata, Ahmed Shawky</creatorcontrib><creatorcontrib>El Nimr, Tarek</creatorcontrib><creatorcontrib>Elshahat, Khaled M.</creatorcontrib><title>Improving patient care and accuracy of given doses in radiation therapy using in vivo dosimetry verification</title><title>Oncology and translational medicine</title><addtitle>Oncology and Translational Medicine</addtitle><description>Objective This work aims to verify and improve the dose given for cancer patients in radiation therapy by using diodes to enhance patient in vivo dosimetry on a routine basis. Some characteristics of two available semi-conductor diode dosimetry systems were evaluated. Methods The diodes had been calibrated to read the dose at Dmax below the surface. Correction factors of clinical relevance were quantified to convert the diode readings into patient dose. The diode was irradiated at various gantry angles (increments of 45~), various Field Sizes and various Source to Surface Distances (SSDs). Results The maximal response variation in the angular response with respect to an arbitrary angle of 0~ was 1.9%, and the minimum variation was 0.5%. The response of the diode with respect to various field siz- es showed the minimum and the maximum variations in the measured dose from the diode; the calculated doses were -1.6% (for 5 cm x 5 cm field size) and 6.6% (for 40 cm x 40 cm field size). The diode exhibited a significant perturbation in the response, which decreased with increasing SSD. No discrepancies larger than 5% were detected between the expected dose and the measured dose. Conclusion The results indicate that the diodes exhibit excellent linearity, dose reproducibility and minimal anisotropy; that they can be used with confidence for patient dose verification. Furthermore, diodes render real time verification of the dose delivered to patients.</description><subject>临床</subject><subject>化疗</subject><subject>癌症患者</subject><subject>肿瘤学</subject><subject>诊断</subject><issn>2095-9621</issn><issn>2995-5858</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2015</creationdate><recordtype>article</recordtype><recordid>eNo9kF1PwyAUhhujicvcD_COmHhZPVCg9NIsfixZ4o1eEwq0Y9loha3afy91i1echOc9H0-W3WJ4wADlY8RQFJADpjlAwfLxIpuRqmI5E0xcphpSXXGCr7NFjFsAIBUtOSOzbLfa96EbnG9Rrw7O-gPSKlikvEFK62NQekRdg1o3WI9MF21EzqOgjEt459FhY4PqR3SMU4_0Nbihm0C3t4cwosEG1zj9B99kV43aRbs4v_Ps8-X5Y_mWr99fV8unda4xJTQXAmqLtaWW8YJUSvCa4MJiU-OyAk7KGrghhBnMmBaECWBGlGBqAUXauSrm2f2p77fyjfKt3HbH4NNEadrxhwBmwABo4vCJ06GLMdhG9sHtVRglBjmZlSezMpmVk1k5pszdObPpfPuVjv4Pcc5p8lrR4hfmkXfr</recordid><startdate>201510</startdate><enddate>201510</enddate><creator>Shawata, Ahmed Shawky</creator><creator>El Nimr, Tarek</creator><creator>Elshahat, Khaled M.</creator><general>Radiation Oncology Center, Tanta Military Hospital, Egypt%Department of Physics, Faculty of Science, Tanta University, Egypt%Department of Clinical Oncology, Faculty of Medicine, Al Azhar University, Egypt</general><scope>2RA</scope><scope>92L</scope><scope>CQIGP</scope><scope>W91</scope><scope>~WA</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>2B.</scope><scope>4A8</scope><scope>92I</scope><scope>93N</scope><scope>PSX</scope><scope>TCJ</scope></search><sort><creationdate>201510</creationdate><title>Improving patient care and accuracy of given doses in radiation therapy using in vivo dosimetry verification</title><author>Shawata, Ahmed Shawky ; El Nimr, Tarek ; Elshahat, Khaled M.</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c1424-880be1ce4e56329a86b213e1db1790627b06d225d155c825805d870db803acc93</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2015</creationdate><topic>临床</topic><topic>化疗</topic><topic>癌症患者</topic><topic>肿瘤学</topic><topic>诊断</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Shawata, Ahmed Shawky</creatorcontrib><creatorcontrib>El Nimr, Tarek</creatorcontrib><creatorcontrib>Elshahat, Khaled M.</creatorcontrib><collection>中文科技期刊数据库</collection><collection>中文科技期刊数据库-CALIS站点</collection><collection>中文科技期刊数据库-7.0平台</collection><collection>中文科技期刊数据库-医药卫生</collection><collection>中文科技期刊数据库- 镜像站点</collection><collection>CrossRef</collection><collection>Wanfang Data Journals - Hong Kong</collection><collection>WANFANG Data Centre</collection><collection>Wanfang Data Journals</collection><collection>万方数据期刊 - 香港版</collection><collection>China Online Journals (COJ)</collection><collection>China Online Journals (COJ)</collection><jtitle>Oncology and translational medicine</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Shawata, Ahmed Shawky</au><au>El Nimr, Tarek</au><au>Elshahat, Khaled M.</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Improving patient care and accuracy of given doses in radiation therapy using in vivo dosimetry verification</atitle><jtitle>Oncology and translational medicine</jtitle><addtitle>Oncology and Translational Medicine</addtitle><date>2015-10</date><risdate>2015</risdate><volume>1</volume><issue>5</issue><spage>212</spage><epage>217</epage><pages>212-217</pages><issn>2095-9621</issn><eissn>2995-5858</eissn><abstract>Objective This work aims to verify and improve the dose given for cancer patients in radiation therapy by using diodes to enhance patient in vivo dosimetry on a routine basis. Some characteristics of two available semi-conductor diode dosimetry systems were evaluated. Methods The diodes had been calibrated to read the dose at Dmax below the surface. Correction factors of clinical relevance were quantified to convert the diode readings into patient dose. The diode was irradiated at various gantry angles (increments of 45~), various Field Sizes and various Source to Surface Distances (SSDs). Results The maximal response variation in the angular response with respect to an arbitrary angle of 0~ was 1.9%, and the minimum variation was 0.5%. The response of the diode with respect to various field siz- es showed the minimum and the maximum variations in the measured dose from the diode; the calculated doses were -1.6% (for 5 cm x 5 cm field size) and 6.6% (for 40 cm x 40 cm field size). The diode exhibited a significant perturbation in the response, which decreased with increasing SSD. No discrepancies larger than 5% were detected between the expected dose and the measured dose. Conclusion The results indicate that the diodes exhibit excellent linearity, dose reproducibility and minimal anisotropy; that they can be used with confidence for patient dose verification. Furthermore, diodes render real time verification of the dose delivered to patients.</abstract><pub>Radiation Oncology Center, Tanta Military Hospital, Egypt%Department of Physics, Faculty of Science, Tanta University, Egypt%Department of Clinical Oncology, Faculty of Medicine, Al Azhar University, Egypt</pub><doi>10.1007/s10330-014-0035-y</doi><tpages>6</tpages><oa>free_for_read</oa></addata></record>
fulltext fulltext
identifier ISSN: 2095-9621
ispartof Oncology and translational medicine, 2015-10, Vol.1 (5), p.212-217
issn 2095-9621
2995-5858
language eng
recordid cdi_wanfang_journals_dgyx201505004
source DOAJ Directory of Open Access Journals
subjects 临床
化疗
癌症患者
肿瘤学
诊断
title Improving patient care and accuracy of given doses in radiation therapy using in vivo dosimetry verification
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2024-12-27T08%3A42%3A57IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-wanfang_jour_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Improving%20patient%20care%20and%20accuracy%20of%20given%20doses%20in%20radiation%20therapy%20using%20in%20vivo%20dosimetry%20verification&rft.jtitle=Oncology%20and%20translational%20medicine&rft.au=Shawata,%20Ahmed%20Shawky&rft.date=2015-10&rft.volume=1&rft.issue=5&rft.spage=212&rft.epage=217&rft.pages=212-217&rft.issn=2095-9621&rft.eissn=2995-5858&rft_id=info:doi/10.1007/s10330-014-0035-y&rft_dat=%3Cwanfang_jour_cross%3Edgyx201505004%3C/wanfang_jour_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_id=info:pmid/&rft_cqvip_id=666494794&rft_wanfj_id=dgyx201505004&rfr_iscdi=true