Automated differential scanning calorimetry at low temperatures
A Perkin--Elmer DSC-2 scanning calorimeter was operated by means of a PDP-11/34 computer with time-shared scanner and voltmeter. Special attention was paid to the problems of measurement below room temp. down to the low-temp. limit. It was found that neon rather than helium, should be used as a purg...
Gespeichert in:
Veröffentlicht in: | International journal of thermophysics 1985-03, Vol.6 (2), p.165-175 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A Perkin--Elmer DSC-2 scanning calorimeter was operated by means of a PDP-11/34 computer with time-shared scanner and voltmeter. Special attention was paid to the problems of measurement below room temp. down to the low-temp. limit. It was found that neon rather than helium, should be used as a purge gas, scans should always be started at a standardized liquid N level, and gas flow to the dry box should be stopped during measurements. Results on benzoic acid were then accurate to 0.6% from 120-300K. The specific heat of Sb was measured in the temp. interval 120-720K. 7 ref.--AA |
---|---|