A unique test facility for calibration of domestic flow meters under dynamic flow conditions

In the early nineties a hot water test facility was planned and constructed for calibration and testing of volume and flow meters at the National Volume Measurement Laboratory at RISE (formerly SP Technical Research Institute of Sweden). The main feature of the test facility is the capability to mea...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Flow measurement and instrumentation 2021-06, Vol.79, p.101934, Article 101934
Hauptverfasser: Büker, Oliver, Stolt, Krister, Lindström, Kent, Wennergren, Per, Penttinen, Olle, Mattiasson, Kerstin
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:In the early nineties a hot water test facility was planned and constructed for calibration and testing of volume and flow meters at the National Volume Measurement Laboratory at RISE (formerly SP Technical Research Institute of Sweden). The main feature of the test facility is the capability to measure flow in a wide temperature and flow range with very high accuracy. The objective of the project, which was initiated in 1989, was to design equipment for calibration of flow meters with stable flow and temperature conditions. After many years of international debate whether static testing is adequate to represent the later more dynamic application of domestic water meters, the EMPIR project 17IND13 Metrology for real-world domestic water metering (“Metrowamet”) was launched in 2018. The project investigates the influence of dynamic flow testing on the measurement accuracy of different types of domestic flow meters. One of the main objectives of the project is the development of infrastructure to carry out dynamic flow measurements. The existing test facility at RISE was at the time of construction one of the best hot and cold-water test facilities in the world. Due to the Metrowamet project the test facility has been upgraded to meet the needs of an infrastructure for dynamic flow investigations. The first findings from dynamic consumption profile measurements are reported in this paper. •A unique water flow test facility has been upgraded and updated.•The test rig allows now traceable dynamic flow testing of domestic water meters.•The core elements are digital valves and a piston prover with an IMS as reference.•Measurements with a consumption profile give an impression of the test rig potential.
ISSN:0955-5986
1873-6998
1873-6998
DOI:10.1016/j.flowmeasinst.2021.101934