Volumetric Breast Density Estimation on Mammograms Using Breast Tissue Equivalent Phantoms – An Update

Methods for improving the accuracy of a technique for estimating volumetric breast density are described. A breast tissue-equivalent phantom encompassing a range of thicknesses and compositions of tissue is used to evaluate the sources of error in the technique. The image acquisition parameters that...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Hauptverfasser: Augustine, Bindu J., Mawdsley, Gordon E., Boyd, Norman F., Yaffe, Martin J.
Format: Buchkapitel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 25
container_issue
container_start_page 18
container_title
container_volume
creator Augustine, Bindu J.
Mawdsley, Gordon E.
Boyd, Norman F.
Yaffe, Martin J.
description Methods for improving the accuracy of a technique for estimating volumetric breast density are described. A breast tissue-equivalent phantom encompassing a range of thicknesses and compositions of tissue is used to evaluate the sources of error in the technique. The image acquisition parameters that can affect the accuracy of calibration are considered, and sensitivity to these factors is evaluated. The robustness of the technique was tested by obtaining calibration images on 24 mammography machines, at 18 different sites, over a period of 3 years. The ability to use a single calibration on all machines of a given model type was assessed by comparing effective linear attenuation coefficients of fat and fibroglandular tissues, derived from the calibration phantom images obtained from various machines.
doi_str_mv 10.1007/11783237_3
format Book Chapter
fullrecord <record><control><sourceid>springer</sourceid><recordid>TN_cdi_springer_books_10_1007_11783237_3</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>springer_books_10_1007_11783237_3</sourcerecordid><originalsourceid>FETCH-LOGICAL-s188t-dec502c6348df447d33d42845d29ba089bc712023c6b0979b64976b8c7c6e9bc3</originalsourceid><addsrcrecordid>eNpFkMtOwzAURM1LIpRu-AIv2QRsX8ePZWnLQyqCRcM2chy3NTRJiV0kdvwDf8iXEEQRVyPdxRyNRoPQGSUXlBB5SalUwEAWsIdOIOMEMsEk30cJFZSmAFwfoKHuoZ2XyUOUECAs1ZLDMRqG8Ez6AyoZFQlaPbXrbe1i5y2-6pwJEU9cE3x8x9MQfW2ibxvc697UdbvsTB1wHnyz_KPnPoStw9PXrX8za9dE_LgyTWx77uvjE48anG8qE90pOlqYdXDD3R-g_Ho6H9-ms4ebu_FolgaqVEwrZzPCrACuqgXnsgKoOFM8q5guDVG6tJIywsCKkmipS8G1FKWy0grXmzBA57-5YdP1NV1XlG37EgpKip8Fi_8F4RuIZl9M</addsrcrecordid><sourcetype>Publisher</sourcetype><iscdi>true</iscdi><recordtype>book_chapter</recordtype></control><display><type>book_chapter</type><title>Volumetric Breast Density Estimation on Mammograms Using Breast Tissue Equivalent Phantoms – An Update</title><source>Springer Books</source><creator>Augustine, Bindu J. ; Mawdsley, Gordon E. ; Boyd, Norman F. ; Yaffe, Martin J.</creator><contributor>Astley, Susan M. ; Zwiggelaar, Reyer ; Brady, Michael ; Rose, Chris</contributor><creatorcontrib>Augustine, Bindu J. ; Mawdsley, Gordon E. ; Boyd, Norman F. ; Yaffe, Martin J. ; Astley, Susan M. ; Zwiggelaar, Reyer ; Brady, Michael ; Rose, Chris</creatorcontrib><description>Methods for improving the accuracy of a technique for estimating volumetric breast density are described. A breast tissue-equivalent phantom encompassing a range of thicknesses and compositions of tissue is used to evaluate the sources of error in the technique. The image acquisition parameters that can affect the accuracy of calibration are considered, and sensitivity to these factors is evaluated. The robustness of the technique was tested by obtaining calibration images on 24 mammography machines, at 18 different sites, over a period of 3 years. The ability to use a single calibration on all machines of a given model type was assessed by comparing effective linear attenuation coefficients of fat and fibroglandular tissues, derived from the calibration phantom images obtained from various machines.</description><identifier>ISSN: 0302-9743</identifier><identifier>ISBN: 9783540356257</identifier><identifier>ISBN: 3540356258</identifier><identifier>EISSN: 1611-3349</identifier><identifier>EISBN: 3540356274</identifier><identifier>EISBN: 9783540356271</identifier><identifier>DOI: 10.1007/11783237_3</identifier><language>eng</language><publisher>Berlin, Heidelberg: Springer Berlin Heidelberg</publisher><subject>Beam Hardening ; Calibration Image ; Fibroglandular Tissue ; Mammographic Density ; Mammography Unit</subject><ispartof>Digital Mammography, 2006, p.18-25</ispartof><rights>Springer-Verlag Berlin Heidelberg 2006</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><relation>Lecture Notes in Computer Science</relation></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktopdf>$$Uhttps://link.springer.com/content/pdf/10.1007/11783237_3$$EPDF$$P50$$Gspringer$$H</linktopdf><linktohtml>$$Uhttps://link.springer.com/10.1007/11783237_3$$EHTML$$P50$$Gspringer$$H</linktohtml><link.rule.ids>779,780,784,793,27924,38254,41441,42510</link.rule.ids></links><search><contributor>Astley, Susan M.</contributor><contributor>Zwiggelaar, Reyer</contributor><contributor>Brady, Michael</contributor><contributor>Rose, Chris</contributor><creatorcontrib>Augustine, Bindu J.</creatorcontrib><creatorcontrib>Mawdsley, Gordon E.</creatorcontrib><creatorcontrib>Boyd, Norman F.</creatorcontrib><creatorcontrib>Yaffe, Martin J.</creatorcontrib><title>Volumetric Breast Density Estimation on Mammograms Using Breast Tissue Equivalent Phantoms – An Update</title><title>Digital Mammography</title><description>Methods for improving the accuracy of a technique for estimating volumetric breast density are described. A breast tissue-equivalent phantom encompassing a range of thicknesses and compositions of tissue is used to evaluate the sources of error in the technique. The image acquisition parameters that can affect the accuracy of calibration are considered, and sensitivity to these factors is evaluated. The robustness of the technique was tested by obtaining calibration images on 24 mammography machines, at 18 different sites, over a period of 3 years. The ability to use a single calibration on all machines of a given model type was assessed by comparing effective linear attenuation coefficients of fat and fibroglandular tissues, derived from the calibration phantom images obtained from various machines.</description><subject>Beam Hardening</subject><subject>Calibration Image</subject><subject>Fibroglandular Tissue</subject><subject>Mammographic Density</subject><subject>Mammography Unit</subject><issn>0302-9743</issn><issn>1611-3349</issn><isbn>9783540356257</isbn><isbn>3540356258</isbn><isbn>3540356274</isbn><isbn>9783540356271</isbn><fulltext>true</fulltext><rsrctype>book_chapter</rsrctype><creationdate>2006</creationdate><recordtype>book_chapter</recordtype><sourceid/><recordid>eNpFkMtOwzAURM1LIpRu-AIv2QRsX8ePZWnLQyqCRcM2chy3NTRJiV0kdvwDf8iXEEQRVyPdxRyNRoPQGSUXlBB5SalUwEAWsIdOIOMEMsEk30cJFZSmAFwfoKHuoZ2XyUOUECAs1ZLDMRqG8Ez6AyoZFQlaPbXrbe1i5y2-6pwJEU9cE3x8x9MQfW2ibxvc697UdbvsTB1wHnyz_KPnPoStw9PXrX8za9dE_LgyTWx77uvjE48anG8qE90pOlqYdXDD3R-g_Ho6H9-ms4ebu_FolgaqVEwrZzPCrACuqgXnsgKoOFM8q5guDVG6tJIywsCKkmipS8G1FKWy0grXmzBA57-5YdP1NV1XlG37EgpKip8Fi_8F4RuIZl9M</recordid><startdate>2006</startdate><enddate>2006</enddate><creator>Augustine, Bindu J.</creator><creator>Mawdsley, Gordon E.</creator><creator>Boyd, Norman F.</creator><creator>Yaffe, Martin J.</creator><general>Springer Berlin Heidelberg</general><scope/></search><sort><creationdate>2006</creationdate><title>Volumetric Breast Density Estimation on Mammograms Using Breast Tissue Equivalent Phantoms – An Update</title><author>Augustine, Bindu J. ; Mawdsley, Gordon E. ; Boyd, Norman F. ; Yaffe, Martin J.</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-s188t-dec502c6348df447d33d42845d29ba089bc712023c6b0979b64976b8c7c6e9bc3</frbrgroupid><rsrctype>book_chapters</rsrctype><prefilter>book_chapters</prefilter><language>eng</language><creationdate>2006</creationdate><topic>Beam Hardening</topic><topic>Calibration Image</topic><topic>Fibroglandular Tissue</topic><topic>Mammographic Density</topic><topic>Mammography Unit</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Augustine, Bindu J.</creatorcontrib><creatorcontrib>Mawdsley, Gordon E.</creatorcontrib><creatorcontrib>Boyd, Norman F.</creatorcontrib><creatorcontrib>Yaffe, Martin J.</creatorcontrib></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Augustine, Bindu J.</au><au>Mawdsley, Gordon E.</au><au>Boyd, Norman F.</au><au>Yaffe, Martin J.</au><au>Astley, Susan M.</au><au>Zwiggelaar, Reyer</au><au>Brady, Michael</au><au>Rose, Chris</au><format>book</format><genre>bookitem</genre><ristype>CHAP</ristype><atitle>Volumetric Breast Density Estimation on Mammograms Using Breast Tissue Equivalent Phantoms – An Update</atitle><btitle>Digital Mammography</btitle><seriestitle>Lecture Notes in Computer Science</seriestitle><date>2006</date><risdate>2006</risdate><spage>18</spage><epage>25</epage><pages>18-25</pages><issn>0302-9743</issn><eissn>1611-3349</eissn><isbn>9783540356257</isbn><isbn>3540356258</isbn><eisbn>3540356274</eisbn><eisbn>9783540356271</eisbn><abstract>Methods for improving the accuracy of a technique for estimating volumetric breast density are described. A breast tissue-equivalent phantom encompassing a range of thicknesses and compositions of tissue is used to evaluate the sources of error in the technique. The image acquisition parameters that can affect the accuracy of calibration are considered, and sensitivity to these factors is evaluated. The robustness of the technique was tested by obtaining calibration images on 24 mammography machines, at 18 different sites, over a period of 3 years. The ability to use a single calibration on all machines of a given model type was assessed by comparing effective linear attenuation coefficients of fat and fibroglandular tissues, derived from the calibration phantom images obtained from various machines.</abstract><cop>Berlin, Heidelberg</cop><pub>Springer Berlin Heidelberg</pub><doi>10.1007/11783237_3</doi><tpages>8</tpages></addata></record>
fulltext fulltext
identifier ISSN: 0302-9743
ispartof Digital Mammography, 2006, p.18-25
issn 0302-9743
1611-3349
language eng
recordid cdi_springer_books_10_1007_11783237_3
source Springer Books
subjects Beam Hardening
Calibration Image
Fibroglandular Tissue
Mammographic Density
Mammography Unit
title Volumetric Breast Density Estimation on Mammograms Using Breast Tissue Equivalent Phantoms – An Update
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-10T22%3A58%3A41IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-springer&rft_val_fmt=info:ofi/fmt:kev:mtx:book&rft.genre=bookitem&rft.atitle=Volumetric%20Breast%20Density%20Estimation%20on%20Mammograms%20Using%20Breast%20Tissue%20Equivalent%20Phantoms%20%E2%80%93%20An%20Update&rft.btitle=Digital%20Mammography&rft.au=Augustine,%20Bindu%20J.&rft.date=2006&rft.spage=18&rft.epage=25&rft.pages=18-25&rft.issn=0302-9743&rft.eissn=1611-3349&rft.isbn=9783540356257&rft.isbn_list=3540356258&rft_id=info:doi/10.1007/11783237_3&rft_dat=%3Cspringer%3Espringer_books_10_1007_11783237_3%3C/springer%3E%3Curl%3E%3C/url%3E&rft.eisbn=3540356274&rft.eisbn_list=9783540356271&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_id=info:pmid/&rfr_iscdi=true