SU‐GG‐I‐152: Rapid Evaluation of X‐Ray/Light Field Alignment for Digital Radiographic Systems

Purpose: This study investigated the development of a device to rapidly quantify x‐ray and light field alignment. The device is a useful alternative to traditional measurement techniques requiring film and is particularly useful for all‐digital facilities that no longer have film processors. The sys...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Medical Physics 2010-06, Vol.37 (6), p.3136-3136
Hauptverfasser: Hoerner, M, Hintenlang, D
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 3136
container_issue 6
container_start_page 3136
container_title Medical Physics
container_volume 37
creator Hoerner, M
Hintenlang, D
description Purpose: This study investigated the development of a device to rapidly quantify x‐ray and light field alignment. The device is a useful alternative to traditional measurement techniques requiring film and is particularly useful for all‐digital facilities that no longer have film processors. The system that is investigated provides rapid and accurate evaluation of the difference between x‐ray and light fields. Method and Materials: A fiber‐optic coupled (FOC) dosimetry system was adapted to evaluate the misalignment between the light and x‐ray fields of mammography imaging systems. An FOC detector element providing a variable output as a function of exposed dosimeter length is used to measure the deviation of each side of the X‐ray/light field. Each detector element incorporates two FOC dosimeters which were constructed by coupling plastic scintillator fibers, 500 microns in diameter, one 2mm and the other 5cm in length, to sensitive photomultiplier tubes (PMTs). The detector element is positioned such that the 5 cm dosimeter intersects the edge of the light and X‐ray fields. The second dosimeter (2mm) is positioned entirely in the X‐ray field to provide a relative measure of beam intensity. Characterization involved varying the length of the 5 cm dosimeter exposed within the x‐ray field. Results: The system consistently demonstrated linearity with tube output and exposed length. Trials have shown a linear response of detected counts to length of scintillation material in the light field over the range of interest. The system's spatial resolution is evaluated and characterized for a variety of exposure parameters. Conclusion: Multiple PMTs allow all edges of the light field to be evaluated simultaneously, permitting x‐ray/light field alignment for a typical digital mammography system to be completed in approximately ten minutes. The time savings and improved accuracy over current alternatives makes the device an attractive tool for digital mammography quality control.
doi_str_mv 10.1118/1.3468188
format Article
fullrecord <record><control><sourceid>wiley_scita</sourceid><recordid>TN_cdi_scitation_primary_10_1118_1_3468188</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>MP8188</sourcerecordid><originalsourceid>FETCH-LOGICAL-c1778-e52d30981886aa4d6ad65f6454c499bd3a1ae3e0b11740dc63c1a55e30b754693</originalsourceid><addsrcrecordid>eNp9kM9OwkAQxjdGExE9-AZ71aSw0_3T1htBQBKMBiTx1gy727KmUNKtmt58BJ_RJ7EErnqYmcP85ss3HyHXwHoAEPehx4WKIY5PSCcUEQ9EyJJT0mEsEUEomDwnF96_McYUl6xD7GL58_U9mbRt2hbI8I7OcecMHX1g8Y61K7e0zOhru5xj05-5fF3TsbOFoYPC5duN3dY0Kyt673JXY9FeG1fmFe7WTtNF42u78ZfkLMPC26vj7JLlePQyfAhmT5PpcDALNERRHFgZGs6SvX2FKIxCo2SmhBRaJMnKcAS03LIVQCSY0YprQCktZ6tICpXwLrk56Oqq9L6yWbqr3AarJgWW7vNJIT3m07LBgf10hW3-BtPH5yN_e-C9bv_c5_KP-C-1_XTN</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype></control><display><type>article</type><title>SU‐GG‐I‐152: Rapid Evaluation of X‐Ray/Light Field Alignment for Digital Radiographic Systems</title><source>Access via Wiley Online Library</source><source>Alma/SFX Local Collection</source><creator>Hoerner, M ; Hintenlang, D</creator><creatorcontrib>Hoerner, M ; Hintenlang, D</creatorcontrib><description>Purpose: This study investigated the development of a device to rapidly quantify x‐ray and light field alignment. The device is a useful alternative to traditional measurement techniques requiring film and is particularly useful for all‐digital facilities that no longer have film processors. The system that is investigated provides rapid and accurate evaluation of the difference between x‐ray and light fields. Method and Materials: A fiber‐optic coupled (FOC) dosimetry system was adapted to evaluate the misalignment between the light and x‐ray fields of mammography imaging systems. An FOC detector element providing a variable output as a function of exposed dosimeter length is used to measure the deviation of each side of the X‐ray/light field. Each detector element incorporates two FOC dosimeters which were constructed by coupling plastic scintillator fibers, 500 microns in diameter, one 2mm and the other 5cm in length, to sensitive photomultiplier tubes (PMTs). The detector element is positioned such that the 5 cm dosimeter intersects the edge of the light and X‐ray fields. The second dosimeter (2mm) is positioned entirely in the X‐ray field to provide a relative measure of beam intensity. Characterization involved varying the length of the 5 cm dosimeter exposed within the x‐ray field. Results: The system consistently demonstrated linearity with tube output and exposed length. Trials have shown a linear response of detected counts to length of scintillation material in the light field over the range of interest. The system's spatial resolution is evaluated and characterized for a variety of exposure parameters. Conclusion: Multiple PMTs allow all edges of the light field to be evaluated simultaneously, permitting x‐ray/light field alignment for a typical digital mammography system to be completed in approximately ten minutes. The time savings and improved accuracy over current alternatives makes the device an attractive tool for digital mammography quality control.</description><identifier>ISSN: 0094-2405</identifier><identifier>EISSN: 2473-4209</identifier><identifier>DOI: 10.1118/1.3468188</identifier><identifier>CODEN: MPHYA6</identifier><language>eng</language><publisher>American Association of Physicists in Medicine</publisher><subject>Digital mammography ; Digital radiography ; Dosimetry ; Fiber materials ; Film mammography ; Mammography ; Photomultipliers ; Radiography ; Scintillation detectors ; Thin film devices</subject><ispartof>Medical Physics, 2010-06, Vol.37 (6), p.3136-3136</ispartof><rights>American Association of Physicists in Medicine</rights><rights>2010 American Association of Physicists in Medicine</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktohtml>$$Uhttps://onlinelibrary.wiley.com/doi/full/10.1118%2F1.3468188$$EHTML$$P50$$Gwiley$$H</linktohtml><link.rule.ids>309,310,314,780,784,789,790,1417,23930,23931,25140,27924,27925,45575</link.rule.ids></links><search><creatorcontrib>Hoerner, M</creatorcontrib><creatorcontrib>Hintenlang, D</creatorcontrib><title>SU‐GG‐I‐152: Rapid Evaluation of X‐Ray/Light Field Alignment for Digital Radiographic Systems</title><title>Medical Physics</title><description>Purpose: This study investigated the development of a device to rapidly quantify x‐ray and light field alignment. The device is a useful alternative to traditional measurement techniques requiring film and is particularly useful for all‐digital facilities that no longer have film processors. The system that is investigated provides rapid and accurate evaluation of the difference between x‐ray and light fields. Method and Materials: A fiber‐optic coupled (FOC) dosimetry system was adapted to evaluate the misalignment between the light and x‐ray fields of mammography imaging systems. An FOC detector element providing a variable output as a function of exposed dosimeter length is used to measure the deviation of each side of the X‐ray/light field. Each detector element incorporates two FOC dosimeters which were constructed by coupling plastic scintillator fibers, 500 microns in diameter, one 2mm and the other 5cm in length, to sensitive photomultiplier tubes (PMTs). The detector element is positioned such that the 5 cm dosimeter intersects the edge of the light and X‐ray fields. The second dosimeter (2mm) is positioned entirely in the X‐ray field to provide a relative measure of beam intensity. Characterization involved varying the length of the 5 cm dosimeter exposed within the x‐ray field. Results: The system consistently demonstrated linearity with tube output and exposed length. Trials have shown a linear response of detected counts to length of scintillation material in the light field over the range of interest. The system's spatial resolution is evaluated and characterized for a variety of exposure parameters. Conclusion: Multiple PMTs allow all edges of the light field to be evaluated simultaneously, permitting x‐ray/light field alignment for a typical digital mammography system to be completed in approximately ten minutes. The time savings and improved accuracy over current alternatives makes the device an attractive tool for digital mammography quality control.</description><subject>Digital mammography</subject><subject>Digital radiography</subject><subject>Dosimetry</subject><subject>Fiber materials</subject><subject>Film mammography</subject><subject>Mammography</subject><subject>Photomultipliers</subject><subject>Radiography</subject><subject>Scintillation detectors</subject><subject>Thin film devices</subject><issn>0094-2405</issn><issn>2473-4209</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2010</creationdate><recordtype>article</recordtype><recordid>eNp9kM9OwkAQxjdGExE9-AZ71aSw0_3T1htBQBKMBiTx1gy727KmUNKtmt58BJ_RJ7EErnqYmcP85ss3HyHXwHoAEPehx4WKIY5PSCcUEQ9EyJJT0mEsEUEomDwnF96_McYUl6xD7GL58_U9mbRt2hbI8I7OcecMHX1g8Y61K7e0zOhru5xj05-5fF3TsbOFoYPC5duN3dY0Kyt673JXY9FeG1fmFe7WTtNF42u78ZfkLMPC26vj7JLlePQyfAhmT5PpcDALNERRHFgZGs6SvX2FKIxCo2SmhBRaJMnKcAS03LIVQCSY0YprQCktZ6tICpXwLrk56Oqq9L6yWbqr3AarJgWW7vNJIT3m07LBgf10hW3-BtPH5yN_e-C9bv_c5_KP-C-1_XTN</recordid><startdate>201006</startdate><enddate>201006</enddate><creator>Hoerner, M</creator><creator>Hintenlang, D</creator><general>American Association of Physicists in Medicine</general><scope>AAYXX</scope><scope>CITATION</scope></search><sort><creationdate>201006</creationdate><title>SU‐GG‐I‐152: Rapid Evaluation of X‐Ray/Light Field Alignment for Digital Radiographic Systems</title><author>Hoerner, M ; Hintenlang, D</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c1778-e52d30981886aa4d6ad65f6454c499bd3a1ae3e0b11740dc63c1a55e30b754693</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2010</creationdate><topic>Digital mammography</topic><topic>Digital radiography</topic><topic>Dosimetry</topic><topic>Fiber materials</topic><topic>Film mammography</topic><topic>Mammography</topic><topic>Photomultipliers</topic><topic>Radiography</topic><topic>Scintillation detectors</topic><topic>Thin film devices</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Hoerner, M</creatorcontrib><creatorcontrib>Hintenlang, D</creatorcontrib><collection>CrossRef</collection><jtitle>Medical Physics</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Hoerner, M</au><au>Hintenlang, D</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>SU‐GG‐I‐152: Rapid Evaluation of X‐Ray/Light Field Alignment for Digital Radiographic Systems</atitle><jtitle>Medical Physics</jtitle><date>2010-06</date><risdate>2010</risdate><volume>37</volume><issue>6</issue><spage>3136</spage><epage>3136</epage><pages>3136-3136</pages><issn>0094-2405</issn><eissn>2473-4209</eissn><coden>MPHYA6</coden><abstract>Purpose: This study investigated the development of a device to rapidly quantify x‐ray and light field alignment. The device is a useful alternative to traditional measurement techniques requiring film and is particularly useful for all‐digital facilities that no longer have film processors. The system that is investigated provides rapid and accurate evaluation of the difference between x‐ray and light fields. Method and Materials: A fiber‐optic coupled (FOC) dosimetry system was adapted to evaluate the misalignment between the light and x‐ray fields of mammography imaging systems. An FOC detector element providing a variable output as a function of exposed dosimeter length is used to measure the deviation of each side of the X‐ray/light field. Each detector element incorporates two FOC dosimeters which were constructed by coupling plastic scintillator fibers, 500 microns in diameter, one 2mm and the other 5cm in length, to sensitive photomultiplier tubes (PMTs). The detector element is positioned such that the 5 cm dosimeter intersects the edge of the light and X‐ray fields. The second dosimeter (2mm) is positioned entirely in the X‐ray field to provide a relative measure of beam intensity. Characterization involved varying the length of the 5 cm dosimeter exposed within the x‐ray field. Results: The system consistently demonstrated linearity with tube output and exposed length. Trials have shown a linear response of detected counts to length of scintillation material in the light field over the range of interest. The system's spatial resolution is evaluated and characterized for a variety of exposure parameters. Conclusion: Multiple PMTs allow all edges of the light field to be evaluated simultaneously, permitting x‐ray/light field alignment for a typical digital mammography system to be completed in approximately ten minutes. The time savings and improved accuracy over current alternatives makes the device an attractive tool for digital mammography quality control.</abstract><pub>American Association of Physicists in Medicine</pub><doi>10.1118/1.3468188</doi><tpages>1</tpages></addata></record>
fulltext fulltext
identifier ISSN: 0094-2405
ispartof Medical Physics, 2010-06, Vol.37 (6), p.3136-3136
issn 0094-2405
2473-4209
language eng
recordid cdi_scitation_primary_10_1118_1_3468188
source Access via Wiley Online Library; Alma/SFX Local Collection
subjects Digital mammography
Digital radiography
Dosimetry
Fiber materials
Film mammography
Mammography
Photomultipliers
Radiography
Scintillation detectors
Thin film devices
title SU‐GG‐I‐152: Rapid Evaluation of X‐Ray/Light Field Alignment for Digital Radiographic Systems
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-01T23%3A33%3A25IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-wiley_scita&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=SU%E2%80%90GG%E2%80%90I%E2%80%90152:%20Rapid%20Evaluation%20of%20X%E2%80%90Ray/Light%20Field%20Alignment%20for%20Digital%20Radiographic%20Systems&rft.jtitle=Medical%20Physics&rft.au=Hoerner,%20M&rft.date=2010-06&rft.volume=37&rft.issue=6&rft.spage=3136&rft.epage=3136&rft.pages=3136-3136&rft.issn=0094-2405&rft.eissn=2473-4209&rft.coden=MPHYA6&rft_id=info:doi/10.1118/1.3468188&rft_dat=%3Cwiley_scita%3EMP8188%3C/wiley_scita%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_id=info:pmid/&rfr_iscdi=true