Kilovoltage cone-beam CT: Comparative dose and image quality evaluations in partial and full-angle scan protocols

Purpose: To assess imaging dose of partial and full-angle kilovoltage CBCT scan protocols and to evaluate image quality for each protocol. Methods: The authors obtained the CT dose index (CTDI) of the kilovoltage CBCT protocols in an on-board imager by ion chamber (IC) measurements and Monte Carlo (...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Medical physics (Lancaster) 2010-07, Vol.37 (7), p.3648-3659
Hauptverfasser: Kim, Sangroh, Yoo, Sua, Yin, Fang-Fang, Samei, Ehsan, Yoshizumi, Terry
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 3659
container_issue 7
container_start_page 3648
container_title Medical physics (Lancaster)
container_volume 37
creator Kim, Sangroh
Yoo, Sua
Yin, Fang-Fang
Samei, Ehsan
Yoshizumi, Terry
description Purpose: To assess imaging dose of partial and full-angle kilovoltage CBCT scan protocols and to evaluate image quality for each protocol. Methods: The authors obtained the CT dose index (CTDI) of the kilovoltage CBCT protocols in an on-board imager by ion chamber (IC) measurements and Monte Carlo (MC) simulations. A total of six new CBCT scan protocols were evaluated: Standard-dose head (100 kVp, 151 mA s, partial-angle), low-dose head (100 kVp, 75 mA s, partial-angle), high-quality head (100 kVp, 754 mA s, partial-angle), pelvis (125 kVp, 706 mA s, full-angle), pelvis spotlight (125 kVp, 752 mA s, partial-angle), and low-dose thorax (110 kVp, 271 mA s, full-angle). Using the point dose method, various CTDI values were calculated by (1) the conventional weighted CTDI ( CTDI w ) calculation and (2) Bakalyar’s method ( CTDI wb ) . The MC simulations were performed to obtain the CTDI w and CTDI wb , as well as from (3) central slice averaging ( CTDI 2 D ) and (4) volume averaging ( CTDI 3 D ) techniques. The CTDI values of the new protocols were compared to those of the old protocols (full-angle CBCT protocols). Image quality of the new protocols was evaluated following the CBCT image quality assurance (QA) protocol [S. Yoo et al. , “A quality assurance program for the on-board imager®,” Med. Phys. 33(11), 4431–4447 (2006)] testing Hounsfield unit (HU) linearity, spatial linearity/resolution, contrast resolution, and HU uniformity. Results: The CTDI w were found as 6.0, 3.2, 29.0, 25.4, 23.8, and 7.7 mGy for the new protocols, respectively. The CTDI w and CTDI wb differed within + 3 % between IC measurements and MC simulations. Method (2) results were within ±12% of method (1). In MC simulations, the CTDI w and CTDI wb were comparable to the CTDI 2 D and CTDI 3 D with the differences ranging from −4.3% to 20.6%. The CTDI 3 D were smallest among all the CTDI values. CTDI w of the new protocols were found as ∼ 14 times lower for standard head scan and 1.8 times lower for standard body scan than the old protocols, respectively. In the image quality QA tests, all the protocols except low-dose head and low-dose thorax protocols were within the tolerance in the HU verification test. The HU value for the two protocols was always higher than the nominal value. All the protocols passed the spatial linearity/resolution and HU uniformity tests. In the contrast resolution test, only high-quality head and pelvis scan protocols were within the tolerance. In addition, cres
doi_str_mv 10.1118/1.3438478
format Article
fullrecord <record><control><sourceid>proquest_scita</sourceid><recordid>TN_cdi_scitation_primary_10_1118_1_3438478</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>754001925</sourcerecordid><originalsourceid>FETCH-LOGICAL-c5418-f6a651efb2a59882a45c46fdf1d845bed92c210440bc6985f9ea9f18d697ec5c3</originalsourceid><addsrcrecordid>eNqN0U9vFCEYBnBiNHZbPfgFDIkHU5OpwMAM9GBiNv6LNXqoZ8IwLxXDDLsDs2a_vWxnNb3UeOLAjycv74PQM0ouKKXyNb2oeS15Kx-gFeNtXXFG1EO0IkTxinEiTtBpSj8JIU0tyGN0woisKWnZCm0_-xB3MWRzA9jGEaoOzIDX15d4HYeNmUz2O8B9TIDN2GM_HOB2NsHnPYadCXMRcUzYj7jw7E24hW4OoTLjTQCcrCl3U8zRxpCeoEfOhARPj-cZ-v7-3fX6Y3X19cOn9durygpOZeUa0wgKrmNGKCmZ4cLyxvWO9pKLDnrFLKOEc9LZRknhFBjlqOwb1YIVtj5DL5bcmLLXyfoM9kf54Qg2a1YW1LSMFvVyUWW-7Qwp68EnCyGYEeKcdCs4IVQxUeT5Iu0UU5rA6c1UtjHtNSX6UIOm-lhDsc-PqXM3QP9X_tl7AdUCfvkA-_uT9Jdvx8A3iz_843bh97-506g-NKpLoyXg1X8H_Avv4nRnuk3v6t8EZMHL</addsrcrecordid><sourcetype>Open Access Repository</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>754001925</pqid></control><display><type>article</type><title>Kilovoltage cone-beam CT: Comparative dose and image quality evaluations in partial and full-angle scan protocols</title><source>MEDLINE</source><source>Wiley Online Library Journals Frontfile Complete</source><source>Alma/SFX Local Collection</source><creator>Kim, Sangroh ; Yoo, Sua ; Yin, Fang-Fang ; Samei, Ehsan ; Yoshizumi, Terry</creator><creatorcontrib>Kim, Sangroh ; Yoo, Sua ; Yin, Fang-Fang ; Samei, Ehsan ; Yoshizumi, Terry</creatorcontrib><description>Purpose: To assess imaging dose of partial and full-angle kilovoltage CBCT scan protocols and to evaluate image quality for each protocol. Methods: The authors obtained the CT dose index (CTDI) of the kilovoltage CBCT protocols in an on-board imager by ion chamber (IC) measurements and Monte Carlo (MC) simulations. A total of six new CBCT scan protocols were evaluated: Standard-dose head (100 kVp, 151 mA s, partial-angle), low-dose head (100 kVp, 75 mA s, partial-angle), high-quality head (100 kVp, 754 mA s, partial-angle), pelvis (125 kVp, 706 mA s, full-angle), pelvis spotlight (125 kVp, 752 mA s, partial-angle), and low-dose thorax (110 kVp, 271 mA s, full-angle). Using the point dose method, various CTDI values were calculated by (1) the conventional weighted CTDI ( CTDI w ) calculation and (2) Bakalyar’s method ( CTDI wb ) . The MC simulations were performed to obtain the CTDI w and CTDI wb , as well as from (3) central slice averaging ( CTDI 2 D ) and (4) volume averaging ( CTDI 3 D ) techniques. The CTDI values of the new protocols were compared to those of the old protocols (full-angle CBCT protocols). Image quality of the new protocols was evaluated following the CBCT image quality assurance (QA) protocol [S. Yoo et al. , “A quality assurance program for the on-board imager®,” Med. Phys. 33(11), 4431–4447 (2006)] testing Hounsfield unit (HU) linearity, spatial linearity/resolution, contrast resolution, and HU uniformity. Results: The CTDI w were found as 6.0, 3.2, 29.0, 25.4, 23.8, and 7.7 mGy for the new protocols, respectively. The CTDI w and CTDI wb differed within + 3 % between IC measurements and MC simulations. Method (2) results were within ±12% of method (1). In MC simulations, the CTDI w and CTDI wb were comparable to the CTDI 2 D and CTDI 3 D with the differences ranging from −4.3% to 20.6%. The CTDI 3 D were smallest among all the CTDI values. CTDI w of the new protocols were found as ∼ 14 times lower for standard head scan and 1.8 times lower for standard body scan than the old protocols, respectively. In the image quality QA tests, all the protocols except low-dose head and low-dose thorax protocols were within the tolerance in the HU verification test. The HU value for the two protocols was always higher than the nominal value. All the protocols passed the spatial linearity/resolution and HU uniformity tests. In the contrast resolution test, only high-quality head and pelvis scan protocols were within the tolerance. In addition, crescent effect was found in the partial-angle scan protocols. Conclusions: The authors found that CTDI w of the new CBCT protocols has been significantly reduced compared to the old protocols with acceptable image quality. The CTDI w values in the point dose method were close to the volume averaging method within 9%–21% for all the CBCT scan protocols. The Bakalyar’s method produced more accurate dose estimation within 14%. The HU inaccuracy from low-dose head and low-dose thorax protocols can render incorrect dose results in the treatment planning system. When high soft-tissue contrast data are desired, high-quality head or pelvis scan protocol is recommended depending on the imaging area. The point dose method can be applicable to estimate CBCT dose with reasonable accuracy in the clinical environment.</description><identifier>ISSN: 0094-2405</identifier><identifier>EISSN: 2473-4209</identifier><identifier>DOI: 10.1118/1.3438478</identifier><identifier>PMID: 20831072</identifier><identifier>CODEN: MPHYA6</identifier><language>eng</language><publisher>United States: American Association of Physicists in Medicine</publisher><subject>60 APPLIED LIFE SCIENCES ; ACCURACY ; BEAMrnc ; CHEST ; Computed tomography ; computerised tomography ; COMPUTERIZED TOMOGRAPHY ; Cone beam computed tomography ; Cone-Beam Computed Tomography - methods ; cone-beam CT ; CTDI ; DOSIMETRY ; Dosimetry/exposure assessment ; EVALUATION ; HEAD ; image resolution ; IONIZATION CHAMBERS ; Linear Models ; Medical image contrast ; Medical image quality ; Medical imaging ; Medical X‐ray imaging ; Monte Carlo ; MONTE CARLO METHOD ; Monte Carlo methods ; Monte Carlo simulations ; partial-angle scan ; PELVIS ; Phantoms, Imaging ; PLANNING ; point dose method ; QUALITY ASSURANCE ; Quality Control ; Radiation Dosage ; RADIATION DOSES ; RADIATION PROTECTION AND DOSIMETRY ; Radiation treatment ; RESOLUTION ; SIMULATION ; Spatial resolution ; TESTING ; VERIFICATION</subject><ispartof>Medical physics (Lancaster), 2010-07, Vol.37 (7), p.3648-3659</ispartof><rights>American Association of Physicists in Medicine</rights><rights>2010 American Association of Physicists in Medicine</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c5418-f6a651efb2a59882a45c46fdf1d845bed92c210440bc6985f9ea9f18d697ec5c3</citedby><cites>FETCH-LOGICAL-c5418-f6a651efb2a59882a45c46fdf1d845bed92c210440bc6985f9ea9f18d697ec5c3</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktopdf>$$Uhttps://onlinelibrary.wiley.com/doi/pdf/10.1118%2F1.3438478$$EPDF$$P50$$Gwiley$$H</linktopdf><linktohtml>$$Uhttps://onlinelibrary.wiley.com/doi/full/10.1118%2F1.3438478$$EHTML$$P50$$Gwiley$$H</linktohtml><link.rule.ids>230,314,776,780,881,1411,27901,27902,45550,45551</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/20831072$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink><backlink>$$Uhttps://www.osti.gov/biblio/22096721$$D View this record in Osti.gov$$Hfree_for_read</backlink></links><search><creatorcontrib>Kim, Sangroh</creatorcontrib><creatorcontrib>Yoo, Sua</creatorcontrib><creatorcontrib>Yin, Fang-Fang</creatorcontrib><creatorcontrib>Samei, Ehsan</creatorcontrib><creatorcontrib>Yoshizumi, Terry</creatorcontrib><title>Kilovoltage cone-beam CT: Comparative dose and image quality evaluations in partial and full-angle scan protocols</title><title>Medical physics (Lancaster)</title><addtitle>Med Phys</addtitle><description>Purpose: To assess imaging dose of partial and full-angle kilovoltage CBCT scan protocols and to evaluate image quality for each protocol. Methods: The authors obtained the CT dose index (CTDI) of the kilovoltage CBCT protocols in an on-board imager by ion chamber (IC) measurements and Monte Carlo (MC) simulations. A total of six new CBCT scan protocols were evaluated: Standard-dose head (100 kVp, 151 mA s, partial-angle), low-dose head (100 kVp, 75 mA s, partial-angle), high-quality head (100 kVp, 754 mA s, partial-angle), pelvis (125 kVp, 706 mA s, full-angle), pelvis spotlight (125 kVp, 752 mA s, partial-angle), and low-dose thorax (110 kVp, 271 mA s, full-angle). Using the point dose method, various CTDI values were calculated by (1) the conventional weighted CTDI ( CTDI w ) calculation and (2) Bakalyar’s method ( CTDI wb ) . The MC simulations were performed to obtain the CTDI w and CTDI wb , as well as from (3) central slice averaging ( CTDI 2 D ) and (4) volume averaging ( CTDI 3 D ) techniques. The CTDI values of the new protocols were compared to those of the old protocols (full-angle CBCT protocols). Image quality of the new protocols was evaluated following the CBCT image quality assurance (QA) protocol [S. Yoo et al. , “A quality assurance program for the on-board imager®,” Med. Phys. 33(11), 4431–4447 (2006)] testing Hounsfield unit (HU) linearity, spatial linearity/resolution, contrast resolution, and HU uniformity. Results: The CTDI w were found as 6.0, 3.2, 29.0, 25.4, 23.8, and 7.7 mGy for the new protocols, respectively. The CTDI w and CTDI wb differed within + 3 % between IC measurements and MC simulations. Method (2) results were within ±12% of method (1). In MC simulations, the CTDI w and CTDI wb were comparable to the CTDI 2 D and CTDI 3 D with the differences ranging from −4.3% to 20.6%. The CTDI 3 D were smallest among all the CTDI values. CTDI w of the new protocols were found as ∼ 14 times lower for standard head scan and 1.8 times lower for standard body scan than the old protocols, respectively. In the image quality QA tests, all the protocols except low-dose head and low-dose thorax protocols were within the tolerance in the HU verification test. The HU value for the two protocols was always higher than the nominal value. All the protocols passed the spatial linearity/resolution and HU uniformity tests. In the contrast resolution test, only high-quality head and pelvis scan protocols were within the tolerance. In addition, crescent effect was found in the partial-angle scan protocols. Conclusions: The authors found that CTDI w of the new CBCT protocols has been significantly reduced compared to the old protocols with acceptable image quality. The CTDI w values in the point dose method were close to the volume averaging method within 9%–21% for all the CBCT scan protocols. The Bakalyar’s method produced more accurate dose estimation within 14%. The HU inaccuracy from low-dose head and low-dose thorax protocols can render incorrect dose results in the treatment planning system. When high soft-tissue contrast data are desired, high-quality head or pelvis scan protocol is recommended depending on the imaging area. The point dose method can be applicable to estimate CBCT dose with reasonable accuracy in the clinical environment.</description><subject>60 APPLIED LIFE SCIENCES</subject><subject>ACCURACY</subject><subject>BEAMrnc</subject><subject>CHEST</subject><subject>Computed tomography</subject><subject>computerised tomography</subject><subject>COMPUTERIZED TOMOGRAPHY</subject><subject>Cone beam computed tomography</subject><subject>Cone-Beam Computed Tomography - methods</subject><subject>cone-beam CT</subject><subject>CTDI</subject><subject>DOSIMETRY</subject><subject>Dosimetry/exposure assessment</subject><subject>EVALUATION</subject><subject>HEAD</subject><subject>image resolution</subject><subject>IONIZATION CHAMBERS</subject><subject>Linear Models</subject><subject>Medical image contrast</subject><subject>Medical image quality</subject><subject>Medical imaging</subject><subject>Medical X‐ray imaging</subject><subject>Monte Carlo</subject><subject>MONTE CARLO METHOD</subject><subject>Monte Carlo methods</subject><subject>Monte Carlo simulations</subject><subject>partial-angle scan</subject><subject>PELVIS</subject><subject>Phantoms, Imaging</subject><subject>PLANNING</subject><subject>point dose method</subject><subject>QUALITY ASSURANCE</subject><subject>Quality Control</subject><subject>Radiation Dosage</subject><subject>RADIATION DOSES</subject><subject>RADIATION PROTECTION AND DOSIMETRY</subject><subject>Radiation treatment</subject><subject>RESOLUTION</subject><subject>SIMULATION</subject><subject>Spatial resolution</subject><subject>TESTING</subject><subject>VERIFICATION</subject><issn>0094-2405</issn><issn>2473-4209</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2010</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNqN0U9vFCEYBnBiNHZbPfgFDIkHU5OpwMAM9GBiNv6LNXqoZ8IwLxXDDLsDs2a_vWxnNb3UeOLAjycv74PQM0ouKKXyNb2oeS15Kx-gFeNtXXFG1EO0IkTxinEiTtBpSj8JIU0tyGN0woisKWnZCm0_-xB3MWRzA9jGEaoOzIDX15d4HYeNmUz2O8B9TIDN2GM_HOB2NsHnPYadCXMRcUzYj7jw7E24hW4OoTLjTQCcrCl3U8zRxpCeoEfOhARPj-cZ-v7-3fX6Y3X19cOn9durygpOZeUa0wgKrmNGKCmZ4cLyxvWO9pKLDnrFLKOEc9LZRknhFBjlqOwb1YIVtj5DL5bcmLLXyfoM9kf54Qg2a1YW1LSMFvVyUWW-7Qwp68EnCyGYEeKcdCs4IVQxUeT5Iu0UU5rA6c1UtjHtNSX6UIOm-lhDsc-PqXM3QP9X_tl7AdUCfvkA-_uT9Jdvx8A3iz_843bh97-506g-NKpLoyXg1X8H_Avv4nRnuk3v6t8EZMHL</recordid><startdate>201007</startdate><enddate>201007</enddate><creator>Kim, Sangroh</creator><creator>Yoo, Sua</creator><creator>Yin, Fang-Fang</creator><creator>Samei, Ehsan</creator><creator>Yoshizumi, Terry</creator><general>American Association of Physicists in Medicine</general><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope><scope>OTOTI</scope></search><sort><creationdate>201007</creationdate><title>Kilovoltage cone-beam CT: Comparative dose and image quality evaluations in partial and full-angle scan protocols</title><author>Kim, Sangroh ; Yoo, Sua ; Yin, Fang-Fang ; Samei, Ehsan ; Yoshizumi, Terry</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c5418-f6a651efb2a59882a45c46fdf1d845bed92c210440bc6985f9ea9f18d697ec5c3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2010</creationdate><topic>60 APPLIED LIFE SCIENCES</topic><topic>ACCURACY</topic><topic>BEAMrnc</topic><topic>CHEST</topic><topic>Computed tomography</topic><topic>computerised tomography</topic><topic>COMPUTERIZED TOMOGRAPHY</topic><topic>Cone beam computed tomography</topic><topic>Cone-Beam Computed Tomography - methods</topic><topic>cone-beam CT</topic><topic>CTDI</topic><topic>DOSIMETRY</topic><topic>Dosimetry/exposure assessment</topic><topic>EVALUATION</topic><topic>HEAD</topic><topic>image resolution</topic><topic>IONIZATION CHAMBERS</topic><topic>Linear Models</topic><topic>Medical image contrast</topic><topic>Medical image quality</topic><topic>Medical imaging</topic><topic>Medical X‐ray imaging</topic><topic>Monte Carlo</topic><topic>MONTE CARLO METHOD</topic><topic>Monte Carlo methods</topic><topic>Monte Carlo simulations</topic><topic>partial-angle scan</topic><topic>PELVIS</topic><topic>Phantoms, Imaging</topic><topic>PLANNING</topic><topic>point dose method</topic><topic>QUALITY ASSURANCE</topic><topic>Quality Control</topic><topic>Radiation Dosage</topic><topic>RADIATION DOSES</topic><topic>RADIATION PROTECTION AND DOSIMETRY</topic><topic>Radiation treatment</topic><topic>RESOLUTION</topic><topic>SIMULATION</topic><topic>Spatial resolution</topic><topic>TESTING</topic><topic>VERIFICATION</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Kim, Sangroh</creatorcontrib><creatorcontrib>Yoo, Sua</creatorcontrib><creatorcontrib>Yin, Fang-Fang</creatorcontrib><creatorcontrib>Samei, Ehsan</creatorcontrib><creatorcontrib>Yoshizumi, Terry</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><collection>OSTI.GOV</collection><jtitle>Medical physics (Lancaster)</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Kim, Sangroh</au><au>Yoo, Sua</au><au>Yin, Fang-Fang</au><au>Samei, Ehsan</au><au>Yoshizumi, Terry</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Kilovoltage cone-beam CT: Comparative dose and image quality evaluations in partial and full-angle scan protocols</atitle><jtitle>Medical physics (Lancaster)</jtitle><addtitle>Med Phys</addtitle><date>2010-07</date><risdate>2010</risdate><volume>37</volume><issue>7</issue><spage>3648</spage><epage>3659</epage><pages>3648-3659</pages><issn>0094-2405</issn><eissn>2473-4209</eissn><coden>MPHYA6</coden><abstract>Purpose: To assess imaging dose of partial and full-angle kilovoltage CBCT scan protocols and to evaluate image quality for each protocol. Methods: The authors obtained the CT dose index (CTDI) of the kilovoltage CBCT protocols in an on-board imager by ion chamber (IC) measurements and Monte Carlo (MC) simulations. A total of six new CBCT scan protocols were evaluated: Standard-dose head (100 kVp, 151 mA s, partial-angle), low-dose head (100 kVp, 75 mA s, partial-angle), high-quality head (100 kVp, 754 mA s, partial-angle), pelvis (125 kVp, 706 mA s, full-angle), pelvis spotlight (125 kVp, 752 mA s, partial-angle), and low-dose thorax (110 kVp, 271 mA s, full-angle). Using the point dose method, various CTDI values were calculated by (1) the conventional weighted CTDI ( CTDI w ) calculation and (2) Bakalyar’s method ( CTDI wb ) . The MC simulations were performed to obtain the CTDI w and CTDI wb , as well as from (3) central slice averaging ( CTDI 2 D ) and (4) volume averaging ( CTDI 3 D ) techniques. The CTDI values of the new protocols were compared to those of the old protocols (full-angle CBCT protocols). Image quality of the new protocols was evaluated following the CBCT image quality assurance (QA) protocol [S. Yoo et al. , “A quality assurance program for the on-board imager®,” Med. Phys. 33(11), 4431–4447 (2006)] testing Hounsfield unit (HU) linearity, spatial linearity/resolution, contrast resolution, and HU uniformity. Results: The CTDI w were found as 6.0, 3.2, 29.0, 25.4, 23.8, and 7.7 mGy for the new protocols, respectively. The CTDI w and CTDI wb differed within + 3 % between IC measurements and MC simulations. Method (2) results were within ±12% of method (1). In MC simulations, the CTDI w and CTDI wb were comparable to the CTDI 2 D and CTDI 3 D with the differences ranging from −4.3% to 20.6%. The CTDI 3 D were smallest among all the CTDI values. CTDI w of the new protocols were found as ∼ 14 times lower for standard head scan and 1.8 times lower for standard body scan than the old protocols, respectively. In the image quality QA tests, all the protocols except low-dose head and low-dose thorax protocols were within the tolerance in the HU verification test. The HU value for the two protocols was always higher than the nominal value. All the protocols passed the spatial linearity/resolution and HU uniformity tests. In the contrast resolution test, only high-quality head and pelvis scan protocols were within the tolerance. In addition, crescent effect was found in the partial-angle scan protocols. Conclusions: The authors found that CTDI w of the new CBCT protocols has been significantly reduced compared to the old protocols with acceptable image quality. The CTDI w values in the point dose method were close to the volume averaging method within 9%–21% for all the CBCT scan protocols. The Bakalyar’s method produced more accurate dose estimation within 14%. The HU inaccuracy from low-dose head and low-dose thorax protocols can render incorrect dose results in the treatment planning system. When high soft-tissue contrast data are desired, high-quality head or pelvis scan protocol is recommended depending on the imaging area. The point dose method can be applicable to estimate CBCT dose with reasonable accuracy in the clinical environment.</abstract><cop>United States</cop><pub>American Association of Physicists in Medicine</pub><pmid>20831072</pmid><doi>10.1118/1.3438478</doi><tpages>12</tpages></addata></record>
fulltext fulltext
identifier ISSN: 0094-2405
ispartof Medical physics (Lancaster), 2010-07, Vol.37 (7), p.3648-3659
issn 0094-2405
2473-4209
language eng
recordid cdi_scitation_primary_10_1118_1_3438478
source MEDLINE; Wiley Online Library Journals Frontfile Complete; Alma/SFX Local Collection
subjects 60 APPLIED LIFE SCIENCES
ACCURACY
BEAMrnc
CHEST
Computed tomography
computerised tomography
COMPUTERIZED TOMOGRAPHY
Cone beam computed tomography
Cone-Beam Computed Tomography - methods
cone-beam CT
CTDI
DOSIMETRY
Dosimetry/exposure assessment
EVALUATION
HEAD
image resolution
IONIZATION CHAMBERS
Linear Models
Medical image contrast
Medical image quality
Medical imaging
Medical X‐ray imaging
Monte Carlo
MONTE CARLO METHOD
Monte Carlo methods
Monte Carlo simulations
partial-angle scan
PELVIS
Phantoms, Imaging
PLANNING
point dose method
QUALITY ASSURANCE
Quality Control
Radiation Dosage
RADIATION DOSES
RADIATION PROTECTION AND DOSIMETRY
Radiation treatment
RESOLUTION
SIMULATION
Spatial resolution
TESTING
VERIFICATION
title Kilovoltage cone-beam CT: Comparative dose and image quality evaluations in partial and full-angle scan protocols
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-02-08T13%3A04%3A11IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_scita&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Kilovoltage%20cone-beam%20CT:%20Comparative%20dose%20and%20image%20quality%20evaluations%20in%20partial%20and%20full-angle%20scan%20protocols&rft.jtitle=Medical%20physics%20(Lancaster)&rft.au=Kim,%20Sangroh&rft.date=2010-07&rft.volume=37&rft.issue=7&rft.spage=3648&rft.epage=3659&rft.pages=3648-3659&rft.issn=0094-2405&rft.eissn=2473-4209&rft.coden=MPHYA6&rft_id=info:doi/10.1118/1.3438478&rft_dat=%3Cproquest_scita%3E754001925%3C/proquest_scita%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=754001925&rft_id=info:pmid/20831072&rfr_iscdi=true