Precise radiochromic film dosimetry using a flat-bed document scanner

In this study, a measurement protocol is presented that improves the precision of dose measurements using a flat-bed document scanner in conjunction with two new GafChromic® film models, HS and Prototype A EBT exposed to 6 MV photon beams. We established two sources of uncertainties in dose measurem...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Medical physics (Lancaster) 2005-07, Vol.32 (7), p.2245-2253
Hauptverfasser: Devic, Slobodan, Seuntjens, Jan, Sham, Edwin, Podgorsak, Ervin B., Schmidtlein, C. Ross, Kirov, Assen S., Soares, Christopher G.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:In this study, a measurement protocol is presented that improves the precision of dose measurements using a flat-bed document scanner in conjunction with two new GafChromic® film models, HS and Prototype A EBT exposed to 6 MV photon beams. We established two sources of uncertainties in dose measurements, governed by measurement and calibration curve fit parameters contributions. We have quantitatively assessed the influence of different steps in the protocol on the overall dose measurement uncertainty. Applying the protocol described in this paper on the Agfa Arcus II flat-bed document scanner, the overall one-sigma dose measurement uncertainty for an uniform field amounts to 2% or less for doses above around 0.4 Gy in the case of the EBT (Prototype A), and for doses above 5 Gy in the case of the HS model GafChromic® film using a region of interest 2 × 2 mm 2 in size.
ISSN:0094-2405
2473-4209
DOI:10.1118/1.1929253