Vacuum Tube Circuit to Measure the Logarithm of a Direct Current

A circuit, using a single tube, and having an output voltage which is proportional to the logarithm of a direct current input, is described. The input current range extends from less than 10−7 ampere to 10−5 ampere if a closely linear logarithmic relationship is desired; and from 10−8 to 10−3 ampere...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Review of scientific instruments 1939-11, Vol.10 (11), p.336-339
Hauptverfasser: Meagher, Ralph E., Bentley, Edward P.
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A circuit, using a single tube, and having an output voltage which is proportional to the logarithm of a direct current input, is described. The input current range extends from less than 10−7 ampere to 10−5 ampere if a closely linear logarithmic relationship is desired; and from 10−8 to 10−3 ampere if only an approximately logarithmic characteristic is necessary. Because the circuit depends for its operation on a fundamental physical law (the Maxwellian distribution of velocities of the thermoelectrons emitted from a hot cathode), its characteristics are constant and predictable.
ISSN:0034-6748
1089-7623
DOI:10.1063/1.1751459