Calibration of the heavy ion beam probe parallel plate analyzer using the gas target and reference beam

The most important and difficult problem in heavy ion beam probe diagnostics is the absolute measurement of the plasma potential. The analyzer that measures the energy of the secondary ions must be perfectly calibrated. The accuracy necessary should be at least 10−4. We used two methods of calibrati...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Review of Scientific Instruments 1997-01, Vol.68 (1), p.308-311
Hauptverfasser: Melnikov, A. V., Kharchev, N. K., Eliseev, L. G., Bondarenko, I. S., Krupnik, L. I., Khrebtov, S. M., Nedzelskij, I. S., Trofimenko, Yu. V.
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 311
container_issue 1
container_start_page 308
container_title Review of Scientific Instruments
container_volume 68
creator Melnikov, A. V.
Kharchev, N. K.
Eliseev, L. G.
Bondarenko, I. S.
Krupnik, L. I.
Khrebtov, S. M.
Nedzelskij, I. S.
Trofimenko, Yu. V.
description The most important and difficult problem in heavy ion beam probe diagnostics is the absolute measurement of the plasma potential. The analyzer that measures the energy of the secondary ions must be perfectly calibrated. The accuracy necessary should be at least 10−4. We used two methods of calibration: in situ using the gas target, and in vitro using the reference beam. Calibration allows one to obtain the gain G and dynamic coefficient F of the analyzer. We developed a procedure of fast calibration that uses the dynamic curve Ub/Ua(δi), where Ub and Ua correspond to the accelerating voltage of the beam and the voltage on the analyzer, and δi is a normalized difference of currents on the plates. The analyzer used on the T-10 tokamak was calibrated by both methods. Results of both calibrations are in a reasonable agreement at the moderate values of the entrance angle. The nonideal effects caused by ionization of the gas target and asymmetric secondary electron exchange are considered.
doi_str_mv 10.1063/1.1147856
format Article
fullrecord <record><control><sourceid>scitation_cross</sourceid><recordid>TN_cdi_scitation_primary_10_1063_1_1147856</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>rsi</sourcerecordid><originalsourceid>FETCH-LOGICAL-c363t-e0ef4271c688be72b1bb9ced7e607fea73d5a927062f1c3f79a85e1c19821fde3</originalsourceid><addsrcrecordid>eNqdkE1LxDAQhoMouK4e_Ae5KnTNNG2SHmXxCxa86LlM00m3km1LUhfWX2_3A7w7l4GZZ16Gh7FbEAsQSj7AAiDTJldnbAbCFIlWqTxnMyFkliidmUt2FeOXmCoHmLFmib6tAo5t3_He8XFNfE243fH9oCLc8CH0FfEBA3pPng8eR-LYod_9UODfse2aw1mDkY8YGhqnbc0DOQrUWTqkXLMLhz7SzanP2efz08fyNVm9v7wtH1eJlUqOCQlyWarBKmMq0mkFVVVYqjUpoR2hlnWORaqFSh1Y6XSBJiewUJgUXE1yzu6OuTb0MU4_lENoNxh2JYhyb6iE8mRoYu-PbLTteDDwP3jbhz-wHGonfwEQPnX0</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype></control><display><type>article</type><title>Calibration of the heavy ion beam probe parallel plate analyzer using the gas target and reference beam</title><source>AIP Digital Archive</source><creator>Melnikov, A. V. ; Kharchev, N. K. ; Eliseev, L. G. ; Bondarenko, I. S. ; Krupnik, L. I. ; Khrebtov, S. M. ; Nedzelskij, I. S. ; Trofimenko, Yu. V.</creator><creatorcontrib>Melnikov, A. V. ; Kharchev, N. K. ; Eliseev, L. G. ; Bondarenko, I. S. ; Krupnik, L. I. ; Khrebtov, S. M. ; Nedzelskij, I. S. ; Trofimenko, Yu. V.</creatorcontrib><description>The most important and difficult problem in heavy ion beam probe diagnostics is the absolute measurement of the plasma potential. The analyzer that measures the energy of the secondary ions must be perfectly calibrated. The accuracy necessary should be at least 10−4. We used two methods of calibration: in situ using the gas target, and in vitro using the reference beam. Calibration allows one to obtain the gain G and dynamic coefficient F of the analyzer. We developed a procedure of fast calibration that uses the dynamic curve Ub/Ua(δi), where Ub and Ua correspond to the accelerating voltage of the beam and the voltage on the analyzer, and δi is a normalized difference of currents on the plates. The analyzer used on the T-10 tokamak was calibrated by both methods. Results of both calibrations are in a reasonable agreement at the moderate values of the entrance angle. The nonideal effects caused by ionization of the gas target and asymmetric secondary electron exchange are considered.</description><identifier>ISSN: 0034-6748</identifier><identifier>EISSN: 1089-7623</identifier><identifier>DOI: 10.1063/1.1147856</identifier><identifier>CODEN: RSINAK</identifier><language>eng</language><ispartof>Review of Scientific Instruments, 1997-01, Vol.68 (1), p.308-311</ispartof><rights>American Institute of Physics</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c363t-e0ef4271c688be72b1bb9ced7e607fea73d5a927062f1c3f79a85e1c19821fde3</citedby></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktohtml>$$Uhttps://pubs.aip.org/rsi/article-lookup/doi/10.1063/1.1147856$$EHTML$$P50$$Gscitation$$H</linktohtml><link.rule.ids>309,310,314,777,781,786,787,1554,23911,23912,25121,27905,27906,76139</link.rule.ids></links><search><creatorcontrib>Melnikov, A. V.</creatorcontrib><creatorcontrib>Kharchev, N. K.</creatorcontrib><creatorcontrib>Eliseev, L. G.</creatorcontrib><creatorcontrib>Bondarenko, I. S.</creatorcontrib><creatorcontrib>Krupnik, L. I.</creatorcontrib><creatorcontrib>Khrebtov, S. M.</creatorcontrib><creatorcontrib>Nedzelskij, I. S.</creatorcontrib><creatorcontrib>Trofimenko, Yu. V.</creatorcontrib><title>Calibration of the heavy ion beam probe parallel plate analyzer using the gas target and reference beam</title><title>Review of Scientific Instruments</title><description>The most important and difficult problem in heavy ion beam probe diagnostics is the absolute measurement of the plasma potential. The analyzer that measures the energy of the secondary ions must be perfectly calibrated. The accuracy necessary should be at least 10−4. We used two methods of calibration: in situ using the gas target, and in vitro using the reference beam. Calibration allows one to obtain the gain G and dynamic coefficient F of the analyzer. We developed a procedure of fast calibration that uses the dynamic curve Ub/Ua(δi), where Ub and Ua correspond to the accelerating voltage of the beam and the voltage on the analyzer, and δi is a normalized difference of currents on the plates. The analyzer used on the T-10 tokamak was calibrated by both methods. Results of both calibrations are in a reasonable agreement at the moderate values of the entrance angle. The nonideal effects caused by ionization of the gas target and asymmetric secondary electron exchange are considered.</description><issn>0034-6748</issn><issn>1089-7623</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>1997</creationdate><recordtype>article</recordtype><recordid>eNqdkE1LxDAQhoMouK4e_Ae5KnTNNG2SHmXxCxa86LlM00m3km1LUhfWX2_3A7w7l4GZZ16Gh7FbEAsQSj7AAiDTJldnbAbCFIlWqTxnMyFkliidmUt2FeOXmCoHmLFmib6tAo5t3_He8XFNfE243fH9oCLc8CH0FfEBA3pPng8eR-LYod_9UODfse2aw1mDkY8YGhqnbc0DOQrUWTqkXLMLhz7SzanP2efz08fyNVm9v7wtH1eJlUqOCQlyWarBKmMq0mkFVVVYqjUpoR2hlnWORaqFSh1Y6XSBJiewUJgUXE1yzu6OuTb0MU4_lENoNxh2JYhyb6iE8mRoYu-PbLTteDDwP3jbhz-wHGonfwEQPnX0</recordid><startdate>199701</startdate><enddate>199701</enddate><creator>Melnikov, A. V.</creator><creator>Kharchev, N. K.</creator><creator>Eliseev, L. G.</creator><creator>Bondarenko, I. S.</creator><creator>Krupnik, L. I.</creator><creator>Khrebtov, S. M.</creator><creator>Nedzelskij, I. S.</creator><creator>Trofimenko, Yu. V.</creator><scope>AAYXX</scope><scope>CITATION</scope></search><sort><creationdate>199701</creationdate><title>Calibration of the heavy ion beam probe parallel plate analyzer using the gas target and reference beam</title><author>Melnikov, A. V. ; Kharchev, N. K. ; Eliseev, L. G. ; Bondarenko, I. S. ; Krupnik, L. I. ; Khrebtov, S. M. ; Nedzelskij, I. S. ; Trofimenko, Yu. V.</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c363t-e0ef4271c688be72b1bb9ced7e607fea73d5a927062f1c3f79a85e1c19821fde3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>1997</creationdate><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Melnikov, A. V.</creatorcontrib><creatorcontrib>Kharchev, N. K.</creatorcontrib><creatorcontrib>Eliseev, L. G.</creatorcontrib><creatorcontrib>Bondarenko, I. S.</creatorcontrib><creatorcontrib>Krupnik, L. I.</creatorcontrib><creatorcontrib>Khrebtov, S. M.</creatorcontrib><creatorcontrib>Nedzelskij, I. S.</creatorcontrib><creatorcontrib>Trofimenko, Yu. V.</creatorcontrib><collection>CrossRef</collection><jtitle>Review of Scientific Instruments</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Melnikov, A. V.</au><au>Kharchev, N. K.</au><au>Eliseev, L. G.</au><au>Bondarenko, I. S.</au><au>Krupnik, L. I.</au><au>Khrebtov, S. M.</au><au>Nedzelskij, I. S.</au><au>Trofimenko, Yu. V.</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Calibration of the heavy ion beam probe parallel plate analyzer using the gas target and reference beam</atitle><jtitle>Review of Scientific Instruments</jtitle><date>1997-01</date><risdate>1997</risdate><volume>68</volume><issue>1</issue><spage>308</spage><epage>311</epage><pages>308-311</pages><issn>0034-6748</issn><eissn>1089-7623</eissn><coden>RSINAK</coden><abstract>The most important and difficult problem in heavy ion beam probe diagnostics is the absolute measurement of the plasma potential. The analyzer that measures the energy of the secondary ions must be perfectly calibrated. The accuracy necessary should be at least 10−4. We used two methods of calibration: in situ using the gas target, and in vitro using the reference beam. Calibration allows one to obtain the gain G and dynamic coefficient F of the analyzer. We developed a procedure of fast calibration that uses the dynamic curve Ub/Ua(δi), where Ub and Ua correspond to the accelerating voltage of the beam and the voltage on the analyzer, and δi is a normalized difference of currents on the plates. The analyzer used on the T-10 tokamak was calibrated by both methods. Results of both calibrations are in a reasonable agreement at the moderate values of the entrance angle. The nonideal effects caused by ionization of the gas target and asymmetric secondary electron exchange are considered.</abstract><doi>10.1063/1.1147856</doi><tpages>4</tpages></addata></record>
fulltext fulltext
identifier ISSN: 0034-6748
ispartof Review of Scientific Instruments, 1997-01, Vol.68 (1), p.308-311
issn 0034-6748
1089-7623
language eng
recordid cdi_scitation_primary_10_1063_1_1147856
source AIP Digital Archive
title Calibration of the heavy ion beam probe parallel plate analyzer using the gas target and reference beam
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-19T00%3A23%3A06IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-scitation_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Calibration%20of%20the%20heavy%20ion%20beam%20probe%20parallel%20plate%20analyzer%20using%20the%20gas%20target%20and%20reference%20beam&rft.jtitle=Review%20of%20Scientific%20Instruments&rft.au=Melnikov,%20A.%20V.&rft.date=1997-01&rft.volume=68&rft.issue=1&rft.spage=308&rft.epage=311&rft.pages=308-311&rft.issn=0034-6748&rft.eissn=1089-7623&rft.coden=RSINAK&rft_id=info:doi/10.1063/1.1147856&rft_dat=%3Cscitation_cross%3Ersi%3C/scitation_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_id=info:pmid/&rfr_iscdi=true