Agreement between corneal diameter measurements obtained with an optical biometer and a Placido-based topographer
The purpose of this study was to compare the white-to-white distance measurements of two devices (IOL Master 500 and Atlas corneal topographer) commonly used in clinical practice to determine if they were interchangeable. Providing information on instrument interchangeability could eliminate several...
Gespeichert in:
Veröffentlicht in: | Arquivos brasileiros de oftalmologia 2022-07, Vol.86 (6) |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | |
---|---|
container_issue | 6 |
container_start_page | |
container_title | Arquivos brasileiros de oftalmologia |
container_volume | 86 |
creator | Teixeira, Eloisa Gomes Rosario M Gomes, Beatriz Fiuza Santana, João Dominice Santhiago, Marcony R Costa, Adroaldo Alencar Moraes, Jr, Haroldo V |
description | The purpose of this study was to compare the white-to-white distance measurements of two devices (IOL Master 500 and Atlas corneal topographer) commonly used in clinical practice to determine if they were interchangeable. Providing information on instrument interchangeability could eliminate several unnecessary tests and consequently reduce the economic burden for the patient and society.
In this prospective, comparative case series, the white-to-white distance was measured by independent examiners using the Atlas topographer (Carl Zeiss Meditec) and the IOL Master 500 (Carl Zeiss Meditec). One eye each of 184 patients was tested. Statistical analyses were performed using a paired t-test, Pearson correlation analysis, and Bland-Altman analysis to compare the measurement methods.
The mean white-to-white distance measurements with the Atlas topographer and the IOL Master 500 were 12.20 ± 0.44 mm and 12.12 ± 0.41 mm, respectively (p |
doi_str_mv | 10.5935/0004-2749.2021-0325 |
format | Article |
fullrecord | <record><control><sourceid>proquest_sciel</sourceid><recordid>TN_cdi_scielo_journals_S0004_27492022005007207</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><scielo_id>S0004_27492022005007207</scielo_id><sourcerecordid>2692755090</sourcerecordid><originalsourceid>FETCH-LOGICAL-c2545-4eda8258f496a88e3f36720e47e8275318f54ab2c51ae2f08c4a934e5e5ebd0d3</originalsourceid><addsrcrecordid>eNo9kU9r3TAQxEVpaF7TfoJC0bEXv67151k6htC0hUALSc5iLa8TBdtyJJvQb1-5L33sYS-_mYEZxj7VsNdW6q8AoCrRKLsXIOoKpNBv2K4-NKYSVui3bHciztn7nJ8AhLJWv2PnUhvdWKN37PnyIRGNNC28peWFaOI-polw4F3AkRZKfCTMa_oHZR7bBcNEHX8JyyPHicd5Cb7gbYhHHKeOI_89oA9drFrMBV7iHB8Szo-UPrCzHodMH1__Bbu__nZ39aO6-fX959XlTeWFVrpS1KER2vTKHtAYkr08NAJINWREo2Vteq2wFV7XSKIH4xVaqUiXazvo5AXbH32zDzRE9xTXNJVAd7vV4rZaSm8CQAMU46YIvhwFc4rPK-XFjSF7GgacKK7ZiYMtwRosFFQeUZ9izol6N6cwYvrjanDbOO6U4bZx3DZOUX1-DVjbkbqT5v8a8i_584jX</addsrcrecordid><sourcetype>Open Access Repository</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2692755090</pqid></control><display><type>article</type><title>Agreement between corneal diameter measurements obtained with an optical biometer and a Placido-based topographer</title><source>IngentaConnect Free/Open Access Journals</source><source>Elektronische Zeitschriftenbibliothek - Frei zugängliche E-Journals</source><creator>Teixeira, Eloisa Gomes Rosario M ; Gomes, Beatriz Fiuza ; Santana, João Dominice ; Santhiago, Marcony R ; Costa, Adroaldo Alencar ; Moraes, Jr, Haroldo V</creator><creatorcontrib>Teixeira, Eloisa Gomes Rosario M ; Gomes, Beatriz Fiuza ; Santana, João Dominice ; Santhiago, Marcony R ; Costa, Adroaldo Alencar ; Moraes, Jr, Haroldo V</creatorcontrib><description>The purpose of this study was to compare the white-to-white distance measurements of two devices (IOL Master 500 and Atlas corneal topographer) commonly used in clinical practice to determine if they were interchangeable. Providing information on instrument interchangeability could eliminate several unnecessary tests and consequently reduce the economic burden for the patient and society.
In this prospective, comparative case series, the white-to-white distance was measured by independent examiners using the Atlas topographer (Carl Zeiss Meditec) and the IOL Master 500 (Carl Zeiss Meditec). One eye each of 184 patients was tested. Statistical analyses were performed using a paired t-test, Pearson correlation analysis, and Bland-Altman analysis to compare the measurement methods.
The mean white-to-white distance measurements with the Atlas topographer and the IOL Master 500 were 12.20 ± 0.44 mm and 12.12 ± 0.41 mm, respectively (p<0.001). The mean white-to-white difference between the two devices was 0.07 mm (95% confidence interval of mean difference: 0.04-0.11 mm). The Pearson correlation coefficient between the two devices was 0.85 (p<0.0001). The 95% limits of agreement between the two devices were -0.38 mm to 0.53 mm.
The Atlas topographer and IOL Master 500 can be used interchangeably with respect to white-to-white distance measurements, as the range of differences is unlikely to affect clinical practice and decision making.</description><identifier>ISSN: 0004-2749</identifier><identifier>ISSN: 1678-2925</identifier><identifier>EISSN: 1678-2925</identifier><identifier>DOI: 10.5935/0004-2749.2021-0325</identifier><identifier>PMID: 35857985</identifier><language>eng</language><publisher>Brazil: Conselho Brasileiro de Oftalmologia</publisher><subject>OPHTHALMOLOGY</subject><ispartof>Arquivos brasileiros de oftalmologia, 2022-07, Vol.86 (6)</ispartof><rights>This work is licensed under a Creative Commons Attribution 4.0 International License.</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><orcidid>0000-0003-3080-6278 ; 0000-0002-0687-7044</orcidid></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>230,314,776,780,881,27901,27902</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/35857985$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Teixeira, Eloisa Gomes Rosario M</creatorcontrib><creatorcontrib>Gomes, Beatriz Fiuza</creatorcontrib><creatorcontrib>Santana, João Dominice</creatorcontrib><creatorcontrib>Santhiago, Marcony R</creatorcontrib><creatorcontrib>Costa, Adroaldo Alencar</creatorcontrib><creatorcontrib>Moraes, Jr, Haroldo V</creatorcontrib><title>Agreement between corneal diameter measurements obtained with an optical biometer and a Placido-based topographer</title><title>Arquivos brasileiros de oftalmologia</title><addtitle>Arq Bras Oftalmol</addtitle><description>The purpose of this study was to compare the white-to-white distance measurements of two devices (IOL Master 500 and Atlas corneal topographer) commonly used in clinical practice to determine if they were interchangeable. Providing information on instrument interchangeability could eliminate several unnecessary tests and consequently reduce the economic burden for the patient and society.
In this prospective, comparative case series, the white-to-white distance was measured by independent examiners using the Atlas topographer (Carl Zeiss Meditec) and the IOL Master 500 (Carl Zeiss Meditec). One eye each of 184 patients was tested. Statistical analyses were performed using a paired t-test, Pearson correlation analysis, and Bland-Altman analysis to compare the measurement methods.
The mean white-to-white distance measurements with the Atlas topographer and the IOL Master 500 were 12.20 ± 0.44 mm and 12.12 ± 0.41 mm, respectively (p<0.001). The mean white-to-white difference between the two devices was 0.07 mm (95% confidence interval of mean difference: 0.04-0.11 mm). The Pearson correlation coefficient between the two devices was 0.85 (p<0.0001). The 95% limits of agreement between the two devices were -0.38 mm to 0.53 mm.
The Atlas topographer and IOL Master 500 can be used interchangeably with respect to white-to-white distance measurements, as the range of differences is unlikely to affect clinical practice and decision making.</description><subject>OPHTHALMOLOGY</subject><issn>0004-2749</issn><issn>1678-2925</issn><issn>1678-2925</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2022</creationdate><recordtype>article</recordtype><recordid>eNo9kU9r3TAQxEVpaF7TfoJC0bEXv67151k6htC0hUALSc5iLa8TBdtyJJvQb1-5L33sYS-_mYEZxj7VsNdW6q8AoCrRKLsXIOoKpNBv2K4-NKYSVui3bHciztn7nJ8AhLJWv2PnUhvdWKN37PnyIRGNNC28peWFaOI-polw4F3AkRZKfCTMa_oHZR7bBcNEHX8JyyPHicd5Cb7gbYhHHKeOI_89oA9drFrMBV7iHB8Szo-UPrCzHodMH1__Bbu__nZ39aO6-fX959XlTeWFVrpS1KER2vTKHtAYkr08NAJINWREo2Vteq2wFV7XSKIH4xVaqUiXazvo5AXbH32zDzRE9xTXNJVAd7vV4rZaSm8CQAMU46YIvhwFc4rPK-XFjSF7GgacKK7ZiYMtwRosFFQeUZ9izol6N6cwYvrjanDbOO6U4bZx3DZOUX1-DVjbkbqT5v8a8i_584jX</recordid><startdate>20220715</startdate><enddate>20220715</enddate><creator>Teixeira, Eloisa Gomes Rosario M</creator><creator>Gomes, Beatriz Fiuza</creator><creator>Santana, João Dominice</creator><creator>Santhiago, Marcony R</creator><creator>Costa, Adroaldo Alencar</creator><creator>Moraes, Jr, Haroldo V</creator><general>Conselho Brasileiro de Oftalmologia</general><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope><scope>GPN</scope><orcidid>https://orcid.org/0000-0003-3080-6278</orcidid><orcidid>https://orcid.org/0000-0002-0687-7044</orcidid></search><sort><creationdate>20220715</creationdate><title>Agreement between corneal diameter measurements obtained with an optical biometer and a Placido-based topographer</title><author>Teixeira, Eloisa Gomes Rosario M ; Gomes, Beatriz Fiuza ; Santana, João Dominice ; Santhiago, Marcony R ; Costa, Adroaldo Alencar ; Moraes, Jr, Haroldo V</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c2545-4eda8258f496a88e3f36720e47e8275318f54ab2c51ae2f08c4a934e5e5ebd0d3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2022</creationdate><topic>OPHTHALMOLOGY</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Teixeira, Eloisa Gomes Rosario M</creatorcontrib><creatorcontrib>Gomes, Beatriz Fiuza</creatorcontrib><creatorcontrib>Santana, João Dominice</creatorcontrib><creatorcontrib>Santhiago, Marcony R</creatorcontrib><creatorcontrib>Costa, Adroaldo Alencar</creatorcontrib><creatorcontrib>Moraes, Jr, Haroldo V</creatorcontrib><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><collection>SciELO</collection><jtitle>Arquivos brasileiros de oftalmologia</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Teixeira, Eloisa Gomes Rosario M</au><au>Gomes, Beatriz Fiuza</au><au>Santana, João Dominice</au><au>Santhiago, Marcony R</au><au>Costa, Adroaldo Alencar</au><au>Moraes, Jr, Haroldo V</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Agreement between corneal diameter measurements obtained with an optical biometer and a Placido-based topographer</atitle><jtitle>Arquivos brasileiros de oftalmologia</jtitle><addtitle>Arq Bras Oftalmol</addtitle><date>2022-07-15</date><risdate>2022</risdate><volume>86</volume><issue>6</issue><issn>0004-2749</issn><issn>1678-2925</issn><eissn>1678-2925</eissn><abstract>The purpose of this study was to compare the white-to-white distance measurements of two devices (IOL Master 500 and Atlas corneal topographer) commonly used in clinical practice to determine if they were interchangeable. Providing information on instrument interchangeability could eliminate several unnecessary tests and consequently reduce the economic burden for the patient and society.
In this prospective, comparative case series, the white-to-white distance was measured by independent examiners using the Atlas topographer (Carl Zeiss Meditec) and the IOL Master 500 (Carl Zeiss Meditec). One eye each of 184 patients was tested. Statistical analyses were performed using a paired t-test, Pearson correlation analysis, and Bland-Altman analysis to compare the measurement methods.
The mean white-to-white distance measurements with the Atlas topographer and the IOL Master 500 were 12.20 ± 0.44 mm and 12.12 ± 0.41 mm, respectively (p<0.001). The mean white-to-white difference between the two devices was 0.07 mm (95% confidence interval of mean difference: 0.04-0.11 mm). The Pearson correlation coefficient between the two devices was 0.85 (p<0.0001). The 95% limits of agreement between the two devices were -0.38 mm to 0.53 mm.
The Atlas topographer and IOL Master 500 can be used interchangeably with respect to white-to-white distance measurements, as the range of differences is unlikely to affect clinical practice and decision making.</abstract><cop>Brazil</cop><pub>Conselho Brasileiro de Oftalmologia</pub><pmid>35857985</pmid><doi>10.5935/0004-2749.2021-0325</doi><orcidid>https://orcid.org/0000-0003-3080-6278</orcidid><orcidid>https://orcid.org/0000-0002-0687-7044</orcidid><oa>free_for_read</oa></addata></record> |
fulltext | fulltext |
identifier | ISSN: 0004-2749 |
ispartof | Arquivos brasileiros de oftalmologia, 2022-07, Vol.86 (6) |
issn | 0004-2749 1678-2925 1678-2925 |
language | eng |
recordid | cdi_scielo_journals_S0004_27492022005007207 |
source | IngentaConnect Free/Open Access Journals; Elektronische Zeitschriftenbibliothek - Frei zugängliche E-Journals |
subjects | OPHTHALMOLOGY |
title | Agreement between corneal diameter measurements obtained with an optical biometer and a Placido-based topographer |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-02-10T18%3A08%3A03IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_sciel&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Agreement%20between%20corneal%20diameter%20measurements%20obtained%20with%20an%20optical%20biometer%20and%20a%20Placido-based%20topographer&rft.jtitle=Arquivos%20brasileiros%20de%20oftalmologia&rft.au=Teixeira,%20Eloisa%20Gomes%20Rosario%20M&rft.date=2022-07-15&rft.volume=86&rft.issue=6&rft.issn=0004-2749&rft.eissn=1678-2925&rft_id=info:doi/10.5935/0004-2749.2021-0325&rft_dat=%3Cproquest_sciel%3E2692755090%3C/proquest_sciel%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=2692755090&rft_id=info:pmid/35857985&rft_scielo_id=S0004_27492022005007207&rfr_iscdi=true |