Towards the reduction of matrix effects in inductively coupled plasma optical emission spectrometry: an argon-nitrogen-hydrogen mixed-gas plasma for the analysis of geological and environmental samplesElectronic supplementary information (ESI) available. See DOI: 10.1039/c7ja00112f
An argon-nitrogen-hydrogen mixed-gas plasma was optimized towards the minimization of matrix effects in inductively coupled plasma optical emission spectrometry. Analyte sensitivity was enhanced by a factor of 1 to 4 ( n = 26 elements) upon adding 3.7% v/v N 2 to the plasma gas and 1.1% v/v H 2 as a...
Gespeichert in:
Hauptverfasser: | , , , |
---|---|
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | An argon-nitrogen-hydrogen mixed-gas plasma was optimized towards the minimization of matrix effects in inductively coupled plasma optical emission spectrometry. Analyte sensitivity was enhanced by a factor of 1 to 4 (
n
= 26 elements) upon adding 3.7% v/v N
2
to the plasma gas and 1.1% v/v H
2
as a sheath around the Ar nebulizer gas flow in the central channel. However, detection limits were on average similar when using the mixed-gas plasma
versus
the Ar plasma, although at least 2-fold improvements resulted for Al and Ca in presence of 5% m/v Na. The 8.9 ± 1.7% (
n
= 26 elements) analyte signal suppression with an Ar plasma in a 0.23% m/v (or 0.1 M) Na matrix was reduced to 5.6 ± 2.3% with the mixed-gas plasma. The mixed-gas plasma was significantly more robust, with a Mg II/Mg I ratio of 13.3 ± 0.1 (
n
= 5 replicates), compared to 9.87 ± 0.16 for an Ar plasma optimized for robustness. The plasma excitation temperature also increased by 717 K for the mixed-gas plasma, confirming the improved excitation conditions. The superior analytical performance of the robust mixed-gas plasma for complex sample matrices was demonstrated through the direct quantitative multi-elemental analysis of a variety of certified reference materials (
i.e.
, lake sediment, till, stream sediment, and natural ore digests), without internal standardization or matrix-matched calibration.
A robust Ar-N
2
-H
2
plasma enables accurate multielemental analysis of geological and environmental samples without matrix-matching or internal standardization. |
---|---|
ISSN: | 0267-9477 1364-5544 |