Metal implants on abdominal CT: does split-filter dual-energy CT provide additional value over iterative metal artifact reduction?
Purpose To assess image quality and metal artifact reduction in split-filter dual-energy CT (sfDECT) of the abdomen with hip or spinal implants using virtual monoenergetic images (VMI) and iterative metal artifact reduction algorithm (iMAR). Methods 102 portal-venous abdominal sfDECTs of patients wi...
Gespeichert in:
Veröffentlicht in: | Abdominal imaging 2023-01, Vol.48 (1), p.424-435 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Purpose
To assess image quality and metal artifact reduction in split-filter dual-energy CT (sfDECT) of the abdomen with hip or spinal implants using virtual monoenergetic images (VMI) and iterative metal artifact reduction algorithm (iMAR).
Methods
102 portal-venous abdominal sfDECTs of patients with hip (
n
= 71) or spinal implants (
n
= 31) were included in this study. Images were reconstructed as 120kVp-equivalent images (Mixed) and VMI (40–190 keV), with and without iMAR. Quantitative artifact and image noise was measured using 12 different ROIs. Subjective image quality was rated by two readers using a five-point Likert-scale in six categories, including overall image quality and vascular contrast.
Results
Lowest quantitative artifact in both hip and spinal implants was measured in VMI
190keV-iMAR
. However, it was not significantly lower than in Mixed
iMAR
(for all ROIs,
p
= 1.00), which were rated best for overall image quality (hip: 1.00 [IQR: 1.00–2.00], spine: 3.00 [IQR:2.00–3.00]). VMI
50keV-iMAR
was rated best for vascular contrast (hip: 1.00 [IQR: 1.00–2.00], spine: 2.00 [IQR: 1.00–2.00]), which was significantly better than Mixed (both,
p
|
---|---|
ISSN: | 2366-0058 2366-004X 2366-0058 |
DOI: | 10.1007/s00261-022-03682-3 |