Assessment of ultrasound shear wave elastography within muscles using different region of interest sizes, manufacturers, probes and acquisition angles: an ex vivo study
The application of shear wave elastography (SWE) in assessment of the musculoskeletal system is affected by various factors. This study aimed to explore the influence of machines, probes, region of interest (ROI) sizes, and the acquisition angles on muscle shear wave speed (SWS). The SWS of isolated...
Gespeichert in:
Veröffentlicht in: | Quantitative imaging in medicine and surgery 2022-06, Vol.12 (6), p.3227-3237 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The application of shear wave elastography (SWE) in assessment of the musculoskeletal system is affected by various factors. This study aimed to explore the influence of machines, probes, region of interest (ROI) sizes, and the acquisition angles on muscle shear wave speed (SWS).
The SWS of
isolated muscles were acquired using 3 different machines (Aixplorer system, SuperSonic Imagine; Acuson S3000, Siemens Healthcare; Resona 7, Mindray) and 2 linear probes (Aixplorer system, SL 10-2 and SL 15-4). Also, 4 different ROI sizes (diameter 1-10 mm) and 9 different acquisition angles (0-40°) were tested. The SWS acquired under different conditions were compared, and the intra-class correlation coefficients (ICC) were used to evaluate reproducibility.
There was a significant difference in SWS acquired using the 3 different machines (P0.80), and nearly all ROI sizes demonstrated an almost perfect degree of intra- and inter-operator agreement (ICC, >0.80). The measurement reliability was higher when the acquisition angles were no more than 20°.
The 3 machines had different SWS values. Attention should be paid when comparing SWS results using different machines. For the Aixplorer system, the ROI size had no effect on the SWS values. Angles larger than 25° will lead to SWS measurements with greater variability compared to smaller angles (≤20°). |
---|---|