Nitinol Staples for Olecranon Osteotomy Fixation, Juxtacortical Versus Inset, Effect on Biomechanical Stability

Hardware prominence is a concern in the fixation of olecranon osteotomies. Staple fixation has provided low-profile secure fixation in other areas of orthopedics. Without insetting, staples still have subcutaneous prominence. This study examines whether nitinol staples, when inset into bone via cort...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of hand surgery global online 2021-07, Vol.3 (4), p.172-175
Hauptverfasser: Mahenthiran, Ajay, Kacena-Merrell, Ethan, Chen, Weinong W., Lim, Boon Him, Dineen, Hannah
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 175
container_issue 4
container_start_page 172
container_title Journal of hand surgery global online
container_volume 3
creator Mahenthiran, Ajay
Kacena-Merrell, Ethan
Chen, Weinong W.
Lim, Boon Him
Dineen, Hannah
description Hardware prominence is a concern in the fixation of olecranon osteotomies. Staple fixation has provided low-profile secure fixation in other areas of orthopedics. Without insetting, staples still have subcutaneous prominence. This study examines whether nitinol staples, when inset into bone via cortical notching, in an olecranon osteotomy can provide fixation strength sufficient for daily activities. Olecranon osteotomies were created in 8 cadaver arms and fixed with 2 nitinol staples. For inset and juxtacortical (noninset) staples, a micrometer measured the displacement between preplaced proximal and distal wires for 3 increasing loads: 0 N, 15 N, and 150 N. This measurement reflected the loss of osteotomy compression. We placed each arm in a pneumatic machine that flexed the elbow from 0° to 90° for 500 cycles at each load. We performed a 2-tailed t test (α value 0.05, β value 0.2) to evaluate for differences in the loss of compression between inset and noninset nitinol staples. We performed the displacement measurement procedure for both staple types at each of the 3 loads. At 0 N, the average displacement of inset was 0 mm and that of noninset was 0.02 mm. At 15 N, the average displacement of inset was 0.02 mm and that of noninset was 0.04 mm. At 150 N, the average displacement of inset was 0.05 mm and that of noninset was 0.09 mm. When comparing the displacement at the 3 force loads, there were no statistically significant differences between the staple types (P = .323). This study shows that inset staples do not considerably weaken osteotomy fixation with nitinol staples. Thus, nitinol staples may provide a low-profile, operatively-efficient fixation method compared with tension-band or screw-and-plate fixation methods for olecranon osteotomies. Future research can include comparing staples with plate constructs. Type of study/level of evidence: Therapeutic III.
doi_str_mv 10.1016/j.jhsg.2021.03.004
format Article
fullrecord <record><control><sourceid>proquest_pubme</sourceid><recordid>TN_cdi_pubmedcentral_primary_oai_pubmedcentral_nih_gov_8991424</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><els_id>S2589514121000220</els_id><sourcerecordid>2649995118</sourcerecordid><originalsourceid>FETCH-LOGICAL-c3704-c46093f057c87ab8c1ae70f46a461b5a95fdb6475cf01207f852afdf1877569b3</originalsourceid><addsrcrecordid>eNp9kUFvFCEYhidGY5vaP9CD4eihOwIDM0NiTGzTak3jHmx7JQzz0WXDwApM0_33sm5t6sUTJDzvy5fvqaoTgmuCSftxXa9X6b6mmJIaNzXG7FV1SHkvFpww8vrF_aA6TmmNMaadaDjt31YHDWeE8xYfVuGHzdYHh35mtXGQkAkRLR3oqHzwaJkyhBymLbq0jyrb4E_R9_kxKx1itlo5dAcxzQld-QT5FF0YAzqjkjyzYQK9Uv4PVdoH62zevqveGOUSHD-dR9Xt5cXN-bfF9fLr1fmX64VuOswWmrVYNAbzTvedGnpNFHTYsFaxlgxcCW7GoWUd1wYTijvTc6rMaEjfdbwVQ3NUfd73buZhglGDz1E5uYl2UnErg7Ly3xdvV_I-PMheCMIoKwUfngpi-DVDynKySYNzykOYk6QtE0JwQvqC0j2qY0gpgnn-hmC5kyXXcidL7mRJ3Mgiq4TevxzwOfJXTQE-7QEoa3qwEGXSFryG0cayYzkG-7_-39LJp9U</addsrcrecordid><sourcetype>Open Access Repository</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2649995118</pqid></control><display><type>article</type><title>Nitinol Staples for Olecranon Osteotomy Fixation, Juxtacortical Versus Inset, Effect on Biomechanical Stability</title><source>PubMed Central</source><source>Directory of Open Access Journals</source><source>Alma/SFX Local Collection</source><source>EZB Electronic Journals Library</source><creator>Mahenthiran, Ajay ; Kacena-Merrell, Ethan ; Chen, Weinong W. ; Lim, Boon Him ; Dineen, Hannah</creator><creatorcontrib>Mahenthiran, Ajay ; Kacena-Merrell, Ethan ; Chen, Weinong W. ; Lim, Boon Him ; Dineen, Hannah</creatorcontrib><description>Hardware prominence is a concern in the fixation of olecranon osteotomies. Staple fixation has provided low-profile secure fixation in other areas of orthopedics. Without insetting, staples still have subcutaneous prominence. This study examines whether nitinol staples, when inset into bone via cortical notching, in an olecranon osteotomy can provide fixation strength sufficient for daily activities. Olecranon osteotomies were created in 8 cadaver arms and fixed with 2 nitinol staples. For inset and juxtacortical (noninset) staples, a micrometer measured the displacement between preplaced proximal and distal wires for 3 increasing loads: 0 N, 15 N, and 150 N. This measurement reflected the loss of osteotomy compression. We placed each arm in a pneumatic machine that flexed the elbow from 0° to 90° for 500 cycles at each load. We performed a 2-tailed t test (α value 0.05, β value 0.2) to evaluate for differences in the loss of compression between inset and noninset nitinol staples. We performed the displacement measurement procedure for both staple types at each of the 3 loads. At 0 N, the average displacement of inset was 0 mm and that of noninset was 0.02 mm. At 15 N, the average displacement of inset was 0.02 mm and that of noninset was 0.04 mm. At 150 N, the average displacement of inset was 0.05 mm and that of noninset was 0.09 mm. When comparing the displacement at the 3 force loads, there were no statistically significant differences between the staple types (P = .323). This study shows that inset staples do not considerably weaken osteotomy fixation with nitinol staples. Thus, nitinol staples may provide a low-profile, operatively-efficient fixation method compared with tension-band or screw-and-plate fixation methods for olecranon osteotomies. Future research can include comparing staples with plate constructs. Type of study/level of evidence: Therapeutic III.</description><identifier>ISSN: 2589-5141</identifier><identifier>EISSN: 2589-5141</identifier><identifier>DOI: 10.1016/j.jhsg.2021.03.004</identifier><identifier>PMID: 35415560</identifier><language>eng</language><publisher>United States: Elsevier Inc</publisher><subject>Olecranon ; Original Research ; Osteotomy ; Staples ; Troughing</subject><ispartof>Journal of hand surgery global online, 2021-07, Vol.3 (4), p.172-175</ispartof><rights>2021 The Authors</rights><rights>2021 The Authors.</rights><rights>2021 The Authors 2021</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c3704-c46093f057c87ab8c1ae70f46a461b5a95fdb6475cf01207f852afdf1877569b3</citedby><cites>FETCH-LOGICAL-c3704-c46093f057c87ab8c1ae70f46a461b5a95fdb6475cf01207f852afdf1877569b3</cites><orcidid>0000-0003-0157-8798</orcidid></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktopdf>$$Uhttps://www.ncbi.nlm.nih.gov/pmc/articles/PMC8991424/pdf/$$EPDF$$P50$$Gpubmedcentral$$Hfree_for_read</linktopdf><linktohtml>$$Uhttps://www.ncbi.nlm.nih.gov/pmc/articles/PMC8991424/$$EHTML$$P50$$Gpubmedcentral$$Hfree_for_read</linktohtml><link.rule.ids>230,314,723,776,780,860,881,27903,27904,53769,53771</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/35415560$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Mahenthiran, Ajay</creatorcontrib><creatorcontrib>Kacena-Merrell, Ethan</creatorcontrib><creatorcontrib>Chen, Weinong W.</creatorcontrib><creatorcontrib>Lim, Boon Him</creatorcontrib><creatorcontrib>Dineen, Hannah</creatorcontrib><title>Nitinol Staples for Olecranon Osteotomy Fixation, Juxtacortical Versus Inset, Effect on Biomechanical Stability</title><title>Journal of hand surgery global online</title><addtitle>J Hand Surg Glob Online</addtitle><description>Hardware prominence is a concern in the fixation of olecranon osteotomies. Staple fixation has provided low-profile secure fixation in other areas of orthopedics. Without insetting, staples still have subcutaneous prominence. This study examines whether nitinol staples, when inset into bone via cortical notching, in an olecranon osteotomy can provide fixation strength sufficient for daily activities. Olecranon osteotomies were created in 8 cadaver arms and fixed with 2 nitinol staples. For inset and juxtacortical (noninset) staples, a micrometer measured the displacement between preplaced proximal and distal wires for 3 increasing loads: 0 N, 15 N, and 150 N. This measurement reflected the loss of osteotomy compression. We placed each arm in a pneumatic machine that flexed the elbow from 0° to 90° for 500 cycles at each load. We performed a 2-tailed t test (α value 0.05, β value 0.2) to evaluate for differences in the loss of compression between inset and noninset nitinol staples. We performed the displacement measurement procedure for both staple types at each of the 3 loads. At 0 N, the average displacement of inset was 0 mm and that of noninset was 0.02 mm. At 15 N, the average displacement of inset was 0.02 mm and that of noninset was 0.04 mm. At 150 N, the average displacement of inset was 0.05 mm and that of noninset was 0.09 mm. When comparing the displacement at the 3 force loads, there were no statistically significant differences between the staple types (P = .323). This study shows that inset staples do not considerably weaken osteotomy fixation with nitinol staples. Thus, nitinol staples may provide a low-profile, operatively-efficient fixation method compared with tension-band or screw-and-plate fixation methods for olecranon osteotomies. Future research can include comparing staples with plate constructs. Type of study/level of evidence: Therapeutic III.</description><subject>Olecranon</subject><subject>Original Research</subject><subject>Osteotomy</subject><subject>Staples</subject><subject>Troughing</subject><issn>2589-5141</issn><issn>2589-5141</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2021</creationdate><recordtype>article</recordtype><recordid>eNp9kUFvFCEYhidGY5vaP9CD4eihOwIDM0NiTGzTak3jHmx7JQzz0WXDwApM0_33sm5t6sUTJDzvy5fvqaoTgmuCSftxXa9X6b6mmJIaNzXG7FV1SHkvFpww8vrF_aA6TmmNMaadaDjt31YHDWeE8xYfVuGHzdYHh35mtXGQkAkRLR3oqHzwaJkyhBymLbq0jyrb4E_R9_kxKx1itlo5dAcxzQld-QT5FF0YAzqjkjyzYQK9Uv4PVdoH62zevqveGOUSHD-dR9Xt5cXN-bfF9fLr1fmX64VuOswWmrVYNAbzTvedGnpNFHTYsFaxlgxcCW7GoWUd1wYTijvTc6rMaEjfdbwVQ3NUfd73buZhglGDz1E5uYl2UnErg7Ly3xdvV_I-PMheCMIoKwUfngpi-DVDynKySYNzykOYk6QtE0JwQvqC0j2qY0gpgnn-hmC5kyXXcidL7mRJ3Mgiq4TevxzwOfJXTQE-7QEoa3qwEGXSFryG0cayYzkG-7_-39LJp9U</recordid><startdate>20210701</startdate><enddate>20210701</enddate><creator>Mahenthiran, Ajay</creator><creator>Kacena-Merrell, Ethan</creator><creator>Chen, Weinong W.</creator><creator>Lim, Boon Him</creator><creator>Dineen, Hannah</creator><general>Elsevier Inc</general><general>Elsevier</general><scope>6I.</scope><scope>AAFTH</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope><scope>5PM</scope><orcidid>https://orcid.org/0000-0003-0157-8798</orcidid></search><sort><creationdate>20210701</creationdate><title>Nitinol Staples for Olecranon Osteotomy Fixation, Juxtacortical Versus Inset, Effect on Biomechanical Stability</title><author>Mahenthiran, Ajay ; Kacena-Merrell, Ethan ; Chen, Weinong W. ; Lim, Boon Him ; Dineen, Hannah</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c3704-c46093f057c87ab8c1ae70f46a461b5a95fdb6475cf01207f852afdf1877569b3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2021</creationdate><topic>Olecranon</topic><topic>Original Research</topic><topic>Osteotomy</topic><topic>Staples</topic><topic>Troughing</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Mahenthiran, Ajay</creatorcontrib><creatorcontrib>Kacena-Merrell, Ethan</creatorcontrib><creatorcontrib>Chen, Weinong W.</creatorcontrib><creatorcontrib>Lim, Boon Him</creatorcontrib><creatorcontrib>Dineen, Hannah</creatorcontrib><collection>ScienceDirect Open Access Titles</collection><collection>Elsevier:ScienceDirect:Open Access</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><collection>PubMed Central (Full Participant titles)</collection><jtitle>Journal of hand surgery global online</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Mahenthiran, Ajay</au><au>Kacena-Merrell, Ethan</au><au>Chen, Weinong W.</au><au>Lim, Boon Him</au><au>Dineen, Hannah</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Nitinol Staples for Olecranon Osteotomy Fixation, Juxtacortical Versus Inset, Effect on Biomechanical Stability</atitle><jtitle>Journal of hand surgery global online</jtitle><addtitle>J Hand Surg Glob Online</addtitle><date>2021-07-01</date><risdate>2021</risdate><volume>3</volume><issue>4</issue><spage>172</spage><epage>175</epage><pages>172-175</pages><issn>2589-5141</issn><eissn>2589-5141</eissn><abstract>Hardware prominence is a concern in the fixation of olecranon osteotomies. Staple fixation has provided low-profile secure fixation in other areas of orthopedics. Without insetting, staples still have subcutaneous prominence. This study examines whether nitinol staples, when inset into bone via cortical notching, in an olecranon osteotomy can provide fixation strength sufficient for daily activities. Olecranon osteotomies were created in 8 cadaver arms and fixed with 2 nitinol staples. For inset and juxtacortical (noninset) staples, a micrometer measured the displacement between preplaced proximal and distal wires for 3 increasing loads: 0 N, 15 N, and 150 N. This measurement reflected the loss of osteotomy compression. We placed each arm in a pneumatic machine that flexed the elbow from 0° to 90° for 500 cycles at each load. We performed a 2-tailed t test (α value 0.05, β value 0.2) to evaluate for differences in the loss of compression between inset and noninset nitinol staples. We performed the displacement measurement procedure for both staple types at each of the 3 loads. At 0 N, the average displacement of inset was 0 mm and that of noninset was 0.02 mm. At 15 N, the average displacement of inset was 0.02 mm and that of noninset was 0.04 mm. At 150 N, the average displacement of inset was 0.05 mm and that of noninset was 0.09 mm. When comparing the displacement at the 3 force loads, there were no statistically significant differences between the staple types (P = .323). This study shows that inset staples do not considerably weaken osteotomy fixation with nitinol staples. Thus, nitinol staples may provide a low-profile, operatively-efficient fixation method compared with tension-band or screw-and-plate fixation methods for olecranon osteotomies. Future research can include comparing staples with plate constructs. Type of study/level of evidence: Therapeutic III.</abstract><cop>United States</cop><pub>Elsevier Inc</pub><pmid>35415560</pmid><doi>10.1016/j.jhsg.2021.03.004</doi><tpages>4</tpages><orcidid>https://orcid.org/0000-0003-0157-8798</orcidid><oa>free_for_read</oa></addata></record>
fulltext fulltext
identifier ISSN: 2589-5141
ispartof Journal of hand surgery global online, 2021-07, Vol.3 (4), p.172-175
issn 2589-5141
2589-5141
language eng
recordid cdi_pubmedcentral_primary_oai_pubmedcentral_nih_gov_8991424
source PubMed Central; Directory of Open Access Journals; Alma/SFX Local Collection; EZB Electronic Journals Library
subjects Olecranon
Original Research
Osteotomy
Staples
Troughing
title Nitinol Staples for Olecranon Osteotomy Fixation, Juxtacortical Versus Inset, Effect on Biomechanical Stability
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-24T22%3A29%3A45IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_pubme&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Nitinol%20Staples%20for%20Olecranon%20Osteotomy%20Fixation,%20Juxtacortical%20Versus%20Inset,%20Effect%20on%20Biomechanical%20Stability&rft.jtitle=Journal%20of%20hand%20surgery%20global%20online&rft.au=Mahenthiran,%20Ajay&rft.date=2021-07-01&rft.volume=3&rft.issue=4&rft.spage=172&rft.epage=175&rft.pages=172-175&rft.issn=2589-5141&rft.eissn=2589-5141&rft_id=info:doi/10.1016/j.jhsg.2021.03.004&rft_dat=%3Cproquest_pubme%3E2649995118%3C/proquest_pubme%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=2649995118&rft_id=info:pmid/35415560&rft_els_id=S2589514121000220&rfr_iscdi=true