The Arclight vs. traditional ophthalmoscope: a cross-over trial

Background/objectives To compare skill acquisition of the new, cost-effective Arclight ophthalmoscope, with the traditional ophthalmoscope (TO), in medical students with no prior experience of ophthalmoscopy. Subjects/methods University of Dundee medical students took part in a cross-over trial. Stu...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Eye (London) 2021-03, Vol.35 (3), p.831-837
Hauptverfasser: Hytiris, Monica Lorraina, Fioratou, Evridiki, Gillan, Stewart N.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Background/objectives To compare skill acquisition of the new, cost-effective Arclight ophthalmoscope, with the traditional ophthalmoscope (TO), in medical students with no prior experience of ophthalmoscopy. Subjects/methods University of Dundee medical students took part in a cross-over trial. Students were divided into two groups and were alternately taught each device using a video tutorial. In period one, Group A was taught the TO first; Group B was taught the Arclight. They were then assessed using simulated objective, structured, clinical, examinations, examining four model heads with lettered fundal photographs of varying sizes of font. Groups crossed over following a 2-week washout period and were taught the second device and reassessed. A questionnaire was distributed to ascertain students’ opinions and preferences. Results Forty medical students participated. Overall, 92.5% of students performed better with the Arclight, irrespective of cross-over trial period. The mean difference in score in period one of the cross-over trial was 16.77 (95% CI: 11.63–21.93), with students performing better with the Arclight ( p  
ISSN:0950-222X
1476-5454
DOI:10.1038/s41433-020-0972-3