Functional retinal imaging using adaptive optics swept-source OCT at 1.6 MHz

Objective optical assessment of photoreceptor function may permit earlier diagnosis of retinal disease than current methods such as perimetry, electrophysiology, and clinical imaging. In this work, we describe an adaptive optics (AO) optical coherence tomography (OCT) system designed to measure func...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Optica 2019-03, Vol.6 (3), p.300-303
Hauptverfasser: Azimipour, Mehdi, Migacz, Justin V, Zawadzki, Robert J, Werner, John S, Jonnal, Ravi S
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 303
container_issue 3
container_start_page 300
container_title Optica
container_volume 6
creator Azimipour, Mehdi
Migacz, Justin V
Zawadzki, Robert J
Werner, John S
Jonnal, Ravi S
description Objective optical assessment of photoreceptor function may permit earlier diagnosis of retinal disease than current methods such as perimetry, electrophysiology, and clinical imaging. In this work, we describe an adaptive optics (AO) optical coherence tomography (OCT) system designed to measure functional responses of single cones to visible stimuli. The OCT subsystem consisted of a raster-scanning Fourier-domain mode-locked laser that acquires A scans at 1.64 MHz with a center wavelength of 1063 nm and an AO system operating in closed-loop. Analysis of serial volumetric images revealed phase changes of cone photoreceptors consistent with outer segment elongation and proportional to stimulus intensity, as well as other morphological changes in the outer segment and retinal pigment epithelium.
doi_str_mv 10.1364/OPTICA.6.000300
format Article
fullrecord <record><control><sourceid>proquest_pubme</sourceid><recordid>TN_cdi_pubmedcentral_primary_oai_pubmedcentral_nih_gov_7839998</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>2483816252</sourcerecordid><originalsourceid>FETCH-LOGICAL-c393t-721d3d7fd91fd4078dbaebd5eaf04d4553ee6b14a66010443724577b6df651163</originalsourceid><addsrcrecordid>eNpVUU1Lw0AQXUSxUnv2Jnv0knQ3sx_JRZBirVCph3peNtlNjaRJ3E0q-utNSS31Mm9g3rx5zEPohpKQgmDT1ev6efYQipAQAoScoasIgAURB3F-0o_QxPuPnkOBEZ6QSzQC4JRGXF6h5byrsraoK11iZ9tij8VWb4pqgzu_r9ropi12Ftc9ZB77L9u0ga87l1m8mq2xbjENBX5Z_Fyji1yX3k4OOEZv88f1bBEsV0-902WQQQJtICNqwMjcJDQ3jMjYpNqmhludE2YY52CtSCnTQhBKGAMZMS5lKkwuet8Cxuh-0G26dGtNZqvW6VI1rnfuvlWtC_V_UhXvalPvlIwhSZK4F7g7CLj6s7O-VdvCZ7YsdWXrzquIxRBTEfGop04HauZq753Nj2coUfsY1BCDEmqIod-4PXV35P89HX4BIlOC7A</addsrcrecordid><sourcetype>Open Access Repository</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2483816252</pqid></control><display><type>article</type><title>Functional retinal imaging using adaptive optics swept-source OCT at 1.6 MHz</title><source>DOAJ Directory of Open Access Journals</source><source>EZB-FREE-00999 freely available EZB journals</source><creator>Azimipour, Mehdi ; Migacz, Justin V ; Zawadzki, Robert J ; Werner, John S ; Jonnal, Ravi S</creator><creatorcontrib>Azimipour, Mehdi ; Migacz, Justin V ; Zawadzki, Robert J ; Werner, John S ; Jonnal, Ravi S</creatorcontrib><description>Objective optical assessment of photoreceptor function may permit earlier diagnosis of retinal disease than current methods such as perimetry, electrophysiology, and clinical imaging. In this work, we describe an adaptive optics (AO) optical coherence tomography (OCT) system designed to measure functional responses of single cones to visible stimuli. The OCT subsystem consisted of a raster-scanning Fourier-domain mode-locked laser that acquires A scans at 1.64 MHz with a center wavelength of 1063 nm and an AO system operating in closed-loop. Analysis of serial volumetric images revealed phase changes of cone photoreceptors consistent with outer segment elongation and proportional to stimulus intensity, as well as other morphological changes in the outer segment and retinal pigment epithelium.</description><identifier>ISSN: 2334-2536</identifier><identifier>EISSN: 2334-2536</identifier><identifier>DOI: 10.1364/OPTICA.6.000300</identifier><identifier>PMID: 33511257</identifier><language>eng</language><publisher>United States</publisher><ispartof>Optica, 2019-03, Vol.6 (3), p.300-303</ispartof><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c393t-721d3d7fd91fd4078dbaebd5eaf04d4553ee6b14a66010443724577b6df651163</citedby><cites>FETCH-LOGICAL-c393t-721d3d7fd91fd4078dbaebd5eaf04d4553ee6b14a66010443724577b6df651163</cites><orcidid>0000-0002-9574-156X ; 0000-0002-2033-8207</orcidid></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>230,314,777,781,861,882,27905,27906</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/33511257$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Azimipour, Mehdi</creatorcontrib><creatorcontrib>Migacz, Justin V</creatorcontrib><creatorcontrib>Zawadzki, Robert J</creatorcontrib><creatorcontrib>Werner, John S</creatorcontrib><creatorcontrib>Jonnal, Ravi S</creatorcontrib><title>Functional retinal imaging using adaptive optics swept-source OCT at 1.6 MHz</title><title>Optica</title><addtitle>Optica</addtitle><description>Objective optical assessment of photoreceptor function may permit earlier diagnosis of retinal disease than current methods such as perimetry, electrophysiology, and clinical imaging. In this work, we describe an adaptive optics (AO) optical coherence tomography (OCT) system designed to measure functional responses of single cones to visible stimuli. The OCT subsystem consisted of a raster-scanning Fourier-domain mode-locked laser that acquires A scans at 1.64 MHz with a center wavelength of 1063 nm and an AO system operating in closed-loop. Analysis of serial volumetric images revealed phase changes of cone photoreceptors consistent with outer segment elongation and proportional to stimulus intensity, as well as other morphological changes in the outer segment and retinal pigment epithelium.</description><issn>2334-2536</issn><issn>2334-2536</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2019</creationdate><recordtype>article</recordtype><recordid>eNpVUU1Lw0AQXUSxUnv2Jnv0knQ3sx_JRZBirVCph3peNtlNjaRJ3E0q-utNSS31Mm9g3rx5zEPohpKQgmDT1ev6efYQipAQAoScoasIgAURB3F-0o_QxPuPnkOBEZ6QSzQC4JRGXF6h5byrsraoK11iZ9tij8VWb4pqgzu_r9ropi12Ftc9ZB77L9u0ga87l1m8mq2xbjENBX5Z_Fyji1yX3k4OOEZv88f1bBEsV0-902WQQQJtICNqwMjcJDQ3jMjYpNqmhludE2YY52CtSCnTQhBKGAMZMS5lKkwuet8Cxuh-0G26dGtNZqvW6VI1rnfuvlWtC_V_UhXvalPvlIwhSZK4F7g7CLj6s7O-VdvCZ7YsdWXrzquIxRBTEfGop04HauZq753Nj2coUfsY1BCDEmqIod-4PXV35P89HX4BIlOC7A</recordid><startdate>20190320</startdate><enddate>20190320</enddate><creator>Azimipour, Mehdi</creator><creator>Migacz, Justin V</creator><creator>Zawadzki, Robert J</creator><creator>Werner, John S</creator><creator>Jonnal, Ravi S</creator><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope><scope>5PM</scope><orcidid>https://orcid.org/0000-0002-9574-156X</orcidid><orcidid>https://orcid.org/0000-0002-2033-8207</orcidid></search><sort><creationdate>20190320</creationdate><title>Functional retinal imaging using adaptive optics swept-source OCT at 1.6 MHz</title><author>Azimipour, Mehdi ; Migacz, Justin V ; Zawadzki, Robert J ; Werner, John S ; Jonnal, Ravi S</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c393t-721d3d7fd91fd4078dbaebd5eaf04d4553ee6b14a66010443724577b6df651163</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2019</creationdate><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Azimipour, Mehdi</creatorcontrib><creatorcontrib>Migacz, Justin V</creatorcontrib><creatorcontrib>Zawadzki, Robert J</creatorcontrib><creatorcontrib>Werner, John S</creatorcontrib><creatorcontrib>Jonnal, Ravi S</creatorcontrib><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><collection>PubMed Central (Full Participant titles)</collection><jtitle>Optica</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Azimipour, Mehdi</au><au>Migacz, Justin V</au><au>Zawadzki, Robert J</au><au>Werner, John S</au><au>Jonnal, Ravi S</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Functional retinal imaging using adaptive optics swept-source OCT at 1.6 MHz</atitle><jtitle>Optica</jtitle><addtitle>Optica</addtitle><date>2019-03-20</date><risdate>2019</risdate><volume>6</volume><issue>3</issue><spage>300</spage><epage>303</epage><pages>300-303</pages><issn>2334-2536</issn><eissn>2334-2536</eissn><abstract>Objective optical assessment of photoreceptor function may permit earlier diagnosis of retinal disease than current methods such as perimetry, electrophysiology, and clinical imaging. In this work, we describe an adaptive optics (AO) optical coherence tomography (OCT) system designed to measure functional responses of single cones to visible stimuli. The OCT subsystem consisted of a raster-scanning Fourier-domain mode-locked laser that acquires A scans at 1.64 MHz with a center wavelength of 1063 nm and an AO system operating in closed-loop. Analysis of serial volumetric images revealed phase changes of cone photoreceptors consistent with outer segment elongation and proportional to stimulus intensity, as well as other morphological changes in the outer segment and retinal pigment epithelium.</abstract><cop>United States</cop><pmid>33511257</pmid><doi>10.1364/OPTICA.6.000300</doi><tpages>4</tpages><orcidid>https://orcid.org/0000-0002-9574-156X</orcidid><orcidid>https://orcid.org/0000-0002-2033-8207</orcidid><oa>free_for_read</oa></addata></record>
fulltext fulltext
identifier ISSN: 2334-2536
ispartof Optica, 2019-03, Vol.6 (3), p.300-303
issn 2334-2536
2334-2536
language eng
recordid cdi_pubmedcentral_primary_oai_pubmedcentral_nih_gov_7839998
source DOAJ Directory of Open Access Journals; EZB-FREE-00999 freely available EZB journals
title Functional retinal imaging using adaptive optics swept-source OCT at 1.6 MHz
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-21T07%3A29%3A37IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_pubme&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Functional%20retinal%20imaging%20using%20adaptive%20optics%20swept-source%20OCT%20at%201.6%20MHz&rft.jtitle=Optica&rft.au=Azimipour,%20Mehdi&rft.date=2019-03-20&rft.volume=6&rft.issue=3&rft.spage=300&rft.epage=303&rft.pages=300-303&rft.issn=2334-2536&rft.eissn=2334-2536&rft_id=info:doi/10.1364/OPTICA.6.000300&rft_dat=%3Cproquest_pubme%3E2483816252%3C/proquest_pubme%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=2483816252&rft_id=info:pmid/33511257&rfr_iscdi=true