Case of an In-Laboratory Vagal Nerve Stimulator Titration for Vagal Nerve Stimulator-Induced Pediatric Obstructive Sleep Apnea

We present the case of a 12-year-old girl with medically refractory epilepsy and a vagal nerve stimulator (VNS), who experienced severe obstructive sleep apnea (OSA) with respiratory events closely matching her VNS settings. We demonstrated a real-time decrease in OSA through an in-laboratory VNS ti...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of clinical sleep medicine 2019-10, Vol.15 (10), p.1539-1542
Hauptverfasser: Chan, Jeremy H M, Owens, James W, Wrede, Joanna E
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 1542
container_issue 10
container_start_page 1539
container_title Journal of clinical sleep medicine
container_volume 15
creator Chan, Jeremy H M
Owens, James W
Wrede, Joanna E
description We present the case of a 12-year-old girl with medically refractory epilepsy and a vagal nerve stimulator (VNS), who experienced severe obstructive sleep apnea (OSA) with respiratory events closely matching her VNS settings. We demonstrated a real-time decrease in OSA through an in-laboratory VNS titration study, decreasing her VNS frequency from 20 Hz to 10 Hz. We were able to demonstrate a baseline level of OSA by turning off the VNS. We then effectively treated her residual OSA with continuous positive airway pressure (CPAP). Novel to our case is that this in-laboratory VNS titration did not result in any subsequent increase in seizure frequency. After 5 months, her seizure frequency had decreased. Our case demonstrates that in-laboratory VNS titration can be an efficient tool for optimizing treatment of VNS-induced OSA and assert that polysomnography before VNS placement is important for guiding future care.
doi_str_mv 10.5664/jcsm.7996
format Article
fullrecord <record><control><sourceid>pubmed</sourceid><recordid>TN_cdi_pubmedcentral_primary_oai_pubmedcentral_nih_gov_6778352</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>31596221</sourcerecordid><originalsourceid>FETCH-LOGICAL-p163t-d48d0db0cd5c4244dcb160d5b6816a1a609171f377694a74bcc2f739d7479edc3</originalsourceid><addsrcrecordid>eNptkNtKw0AQhhdRbK1e-AKyL5C6mz1lb4RSPBSKClZvw55StySbsEkKvfHZTVGLglcz888_H_wDwCVGU8Y5vd6YtpoKKfkRGGPGUCKJFMeHPpMjcNa2G4RoygQ7BSOCmeRpisfgY65aB-sCqgAXIVkqXUfV1XEH39RalfDRxa2DL52v-nKvw5XvBoOvAyyG6X9Tsgi2N87CZ2e96qI38Em3XexN5_fG0rkGzprg1Dk4KVTZuovvOgGvd7er-UOyfLpfzGfLpMGcdImlmUVWI2OZoSml1mjMkWWaZ5grrDiSWOCCCMElVYJqY9JCEGkFFdJZQybg5ovb9LoaBBeGFGXeRF-puMtr5fO_m-Df83W9zbkQGWHpALj6DThc_nySfAJSWng6</addsrcrecordid><sourcetype>Open Access Repository</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype></control><display><type>article</type><title>Case of an In-Laboratory Vagal Nerve Stimulator Titration for Vagal Nerve Stimulator-Induced Pediatric Obstructive Sleep Apnea</title><source>MEDLINE</source><source>EZB-FREE-00999 freely available EZB journals</source><source>PubMed Central</source><source>Alma/SFX Local Collection</source><creator>Chan, Jeremy H M ; Owens, James W ; Wrede, Joanna E</creator><creatorcontrib>Chan, Jeremy H M ; Owens, James W ; Wrede, Joanna E</creatorcontrib><description>We present the case of a 12-year-old girl with medically refractory epilepsy and a vagal nerve stimulator (VNS), who experienced severe obstructive sleep apnea (OSA) with respiratory events closely matching her VNS settings. We demonstrated a real-time decrease in OSA through an in-laboratory VNS titration study, decreasing her VNS frequency from 20 Hz to 10 Hz. We were able to demonstrate a baseline level of OSA by turning off the VNS. We then effectively treated her residual OSA with continuous positive airway pressure (CPAP). Novel to our case is that this in-laboratory VNS titration did not result in any subsequent increase in seizure frequency. After 5 months, her seizure frequency had decreased. Our case demonstrates that in-laboratory VNS titration can be an efficient tool for optimizing treatment of VNS-induced OSA and assert that polysomnography before VNS placement is important for guiding future care.</description><identifier>ISSN: 1550-9389</identifier><identifier>EISSN: 1550-9397</identifier><identifier>DOI: 10.5664/jcsm.7996</identifier><identifier>PMID: 31596221</identifier><language>eng</language><publisher>United States: American Academy of Sleep Medicine</publisher><subject>Case Reports ; Child ; Continuous Positive Airway Pressure - methods ; Female ; Humans ; Polysomnography - methods ; Sleep Apnea, Obstructive - diagnosis ; Sleep Apnea, Obstructive - etiology ; Sleep Apnea, Obstructive - therapy ; Treatment Outcome ; Vagus Nerve Stimulation - adverse effects</subject><ispartof>Journal of clinical sleep medicine, 2019-10, Vol.15 (10), p.1539-1542</ispartof><rights>2019 American Academy of Sleep Medicine.</rights><rights>2019 American Academy of Sleep Medicine 2019</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktopdf>$$Uhttps://www.ncbi.nlm.nih.gov/pmc/articles/PMC6778352/pdf/$$EPDF$$P50$$Gpubmedcentral$$H</linktopdf><linktohtml>$$Uhttps://www.ncbi.nlm.nih.gov/pmc/articles/PMC6778352/$$EHTML$$P50$$Gpubmedcentral$$H</linktohtml><link.rule.ids>230,314,727,780,784,885,27924,27925,53791,53793</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/31596221$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Chan, Jeremy H M</creatorcontrib><creatorcontrib>Owens, James W</creatorcontrib><creatorcontrib>Wrede, Joanna E</creatorcontrib><title>Case of an In-Laboratory Vagal Nerve Stimulator Titration for Vagal Nerve Stimulator-Induced Pediatric Obstructive Sleep Apnea</title><title>Journal of clinical sleep medicine</title><addtitle>J Clin Sleep Med</addtitle><description>We present the case of a 12-year-old girl with medically refractory epilepsy and a vagal nerve stimulator (VNS), who experienced severe obstructive sleep apnea (OSA) with respiratory events closely matching her VNS settings. We demonstrated a real-time decrease in OSA through an in-laboratory VNS titration study, decreasing her VNS frequency from 20 Hz to 10 Hz. We were able to demonstrate a baseline level of OSA by turning off the VNS. We then effectively treated her residual OSA with continuous positive airway pressure (CPAP). Novel to our case is that this in-laboratory VNS titration did not result in any subsequent increase in seizure frequency. After 5 months, her seizure frequency had decreased. Our case demonstrates that in-laboratory VNS titration can be an efficient tool for optimizing treatment of VNS-induced OSA and assert that polysomnography before VNS placement is important for guiding future care.</description><subject>Case Reports</subject><subject>Child</subject><subject>Continuous Positive Airway Pressure - methods</subject><subject>Female</subject><subject>Humans</subject><subject>Polysomnography - methods</subject><subject>Sleep Apnea, Obstructive - diagnosis</subject><subject>Sleep Apnea, Obstructive - etiology</subject><subject>Sleep Apnea, Obstructive - therapy</subject><subject>Treatment Outcome</subject><subject>Vagus Nerve Stimulation - adverse effects</subject><issn>1550-9389</issn><issn>1550-9397</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2019</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNptkNtKw0AQhhdRbK1e-AKyL5C6mz1lb4RSPBSKClZvw55StySbsEkKvfHZTVGLglcz888_H_wDwCVGU8Y5vd6YtpoKKfkRGGPGUCKJFMeHPpMjcNa2G4RoygQ7BSOCmeRpisfgY65aB-sCqgAXIVkqXUfV1XEH39RalfDRxa2DL52v-nKvw5XvBoOvAyyG6X9Tsgi2N87CZ2e96qI38Em3XexN5_fG0rkGzprg1Dk4KVTZuovvOgGvd7er-UOyfLpfzGfLpMGcdImlmUVWI2OZoSml1mjMkWWaZ5grrDiSWOCCCMElVYJqY9JCEGkFFdJZQybg5ovb9LoaBBeGFGXeRF-puMtr5fO_m-Df83W9zbkQGWHpALj6DThc_nySfAJSWng6</recordid><startdate>20191015</startdate><enddate>20191015</enddate><creator>Chan, Jeremy H M</creator><creator>Owens, James W</creator><creator>Wrede, Joanna E</creator><general>American Academy of Sleep Medicine</general><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>5PM</scope></search><sort><creationdate>20191015</creationdate><title>Case of an In-Laboratory Vagal Nerve Stimulator Titration for Vagal Nerve Stimulator-Induced Pediatric Obstructive Sleep Apnea</title><author>Chan, Jeremy H M ; Owens, James W ; Wrede, Joanna E</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-p163t-d48d0db0cd5c4244dcb160d5b6816a1a609171f377694a74bcc2f739d7479edc3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2019</creationdate><topic>Case Reports</topic><topic>Child</topic><topic>Continuous Positive Airway Pressure - methods</topic><topic>Female</topic><topic>Humans</topic><topic>Polysomnography - methods</topic><topic>Sleep Apnea, Obstructive - diagnosis</topic><topic>Sleep Apnea, Obstructive - etiology</topic><topic>Sleep Apnea, Obstructive - therapy</topic><topic>Treatment Outcome</topic><topic>Vagus Nerve Stimulation - adverse effects</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Chan, Jeremy H M</creatorcontrib><creatorcontrib>Owens, James W</creatorcontrib><creatorcontrib>Wrede, Joanna E</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>PubMed Central (Full Participant titles)</collection><jtitle>Journal of clinical sleep medicine</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Chan, Jeremy H M</au><au>Owens, James W</au><au>Wrede, Joanna E</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Case of an In-Laboratory Vagal Nerve Stimulator Titration for Vagal Nerve Stimulator-Induced Pediatric Obstructive Sleep Apnea</atitle><jtitle>Journal of clinical sleep medicine</jtitle><addtitle>J Clin Sleep Med</addtitle><date>2019-10-15</date><risdate>2019</risdate><volume>15</volume><issue>10</issue><spage>1539</spage><epage>1542</epage><pages>1539-1542</pages><issn>1550-9389</issn><eissn>1550-9397</eissn><abstract>We present the case of a 12-year-old girl with medically refractory epilepsy and a vagal nerve stimulator (VNS), who experienced severe obstructive sleep apnea (OSA) with respiratory events closely matching her VNS settings. We demonstrated a real-time decrease in OSA through an in-laboratory VNS titration study, decreasing her VNS frequency from 20 Hz to 10 Hz. We were able to demonstrate a baseline level of OSA by turning off the VNS. We then effectively treated her residual OSA with continuous positive airway pressure (CPAP). Novel to our case is that this in-laboratory VNS titration did not result in any subsequent increase in seizure frequency. After 5 months, her seizure frequency had decreased. Our case demonstrates that in-laboratory VNS titration can be an efficient tool for optimizing treatment of VNS-induced OSA and assert that polysomnography before VNS placement is important for guiding future care.</abstract><cop>United States</cop><pub>American Academy of Sleep Medicine</pub><pmid>31596221</pmid><doi>10.5664/jcsm.7996</doi><tpages>4</tpages><oa>free_for_read</oa></addata></record>
fulltext fulltext
identifier ISSN: 1550-9389
ispartof Journal of clinical sleep medicine, 2019-10, Vol.15 (10), p.1539-1542
issn 1550-9389
1550-9397
language eng
recordid cdi_pubmedcentral_primary_oai_pubmedcentral_nih_gov_6778352
source MEDLINE; EZB-FREE-00999 freely available EZB journals; PubMed Central; Alma/SFX Local Collection
subjects Case Reports
Child
Continuous Positive Airway Pressure - methods
Female
Humans
Polysomnography - methods
Sleep Apnea, Obstructive - diagnosis
Sleep Apnea, Obstructive - etiology
Sleep Apnea, Obstructive - therapy
Treatment Outcome
Vagus Nerve Stimulation - adverse effects
title Case of an In-Laboratory Vagal Nerve Stimulator Titration for Vagal Nerve Stimulator-Induced Pediatric Obstructive Sleep Apnea
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-03T03%3A49%3A29IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-pubmed&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Case%20of%20an%20In-Laboratory%20Vagal%20Nerve%20Stimulator%20Titration%20for%20Vagal%20Nerve%20Stimulator-Induced%20Pediatric%20Obstructive%20Sleep%20Apnea&rft.jtitle=Journal%20of%20clinical%20sleep%20medicine&rft.au=Chan,%20Jeremy%20H%20M&rft.date=2019-10-15&rft.volume=15&rft.issue=10&rft.spage=1539&rft.epage=1542&rft.pages=1539-1542&rft.issn=1550-9389&rft.eissn=1550-9397&rft_id=info:doi/10.5664/jcsm.7996&rft_dat=%3Cpubmed%3E31596221%3C/pubmed%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_id=info:pmid/31596221&rfr_iscdi=true