Influence of acquisition parameters on the magnitude of cone beam computed tomography artifacts

To evaluate the influence of kilovoltage (kVp) and metal artifact reduction tool (MAR) on the magnitude of cone beam CT (CBCT) artifacts. A titanium and zirconia implants were inserted alternately in a posterior region of a mandible. CBCT exams were acquired with ProMax 3D (Planmeca Oy, Helsinki, Fi...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Dento-maxillo-facial radiology 2018-12, Vol.47 (8), p.20180151-20180151
Hauptverfasser: Freitas, Deborah Queiroz, Fontenele, Rocharles Cavalcante, Nascimento, Eduarda Helena Leandro, Vasconcelos, Taruska Ventorini, Noujeim, Marcel
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 20180151
container_issue 8
container_start_page 20180151
container_title Dento-maxillo-facial radiology
container_volume 47
creator Freitas, Deborah Queiroz
Fontenele, Rocharles Cavalcante
Nascimento, Eduarda Helena Leandro
Vasconcelos, Taruska Ventorini
Noujeim, Marcel
description To evaluate the influence of kilovoltage (kVp) and metal artifact reduction tool (MAR) on the magnitude of cone beam CT (CBCT) artifacts. A titanium and zirconia implants were inserted alternately in a posterior region of a mandible. CBCT exams were acquired with ProMax 3D (Planmeca Oy, Helsinki, Finland) and Picasso Trio machines (Vatech, Hwaseong, South Korea) using 70 kVp, 80 kVp and 90 kVp with and without MAR activation. The other exposure factors remained fixed at 5mA, field of view 80 × 50 mm and voxel 0.20 mm. The scans were performed before and after the insertion of the implants. Regions of interest were determined in different distances from the artifact production area (15, 25 and 35 mm) in an axial image, in which standard deviation (SD) of grayscale values was measured and contrast-to-noise ratio (CNR) was calculated. Analysis of variance was used to compare the data. Overall, in cases where the artifact was pronounced, MAR was efficient in reducing SD values. MAR also improved the CNR of ProMax images, but did not affect the Picasso images. Additionally, the higher was the kVp, the lower was the SD value and the higher was the CNR in both machines. In both machines, increasing kVp and MAR are effective in decreasing the CBCT artifacts in all their magnitude when they are pronounced. Therefore, the professionals should choose one of those options or even both considering the purpose of the CBCT imaging and radiation dose for the patient.
doi_str_mv 10.1259/dmfr.20180151
format Article
fullrecord <record><control><sourceid>proquest_pubme</sourceid><recordid>TN_cdi_pubmedcentral_primary_oai_pubmedcentral_nih_gov_6326394</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>2057128577</sourcerecordid><originalsourceid>FETCH-LOGICAL-c387t-bfb853095bbae4d76fb7eca3c93bc0be9d85f99b19e8db034c01f68b88ff86053</originalsourceid><addsrcrecordid>eNpVkctLJDEQh8Piso6P416lj15a8-h0JxdBxBcIXnZhbiFJV2aydDptkhb8723XUfRUVdTHrwo-hH4TfEYol-d9cOmMYiIw4eQHWpGma2ve0PUeWmHKcS0YXe-jg5z_YYwbxttfaJ9KSdqO0hVS96MbZhgtVNFV2j7NPvvi41hNOukABVKulqlsoQp6M_oy9_9RG0eoDOiwdGGaC_RViSFukp62L5VOxTttSz5CP50eMhzv6iH6e3P95-qufni8vb-6fKgtE12pjTOCMyy5MRqavmud6cBqZiUzFhuQveBOSkMkiN5g1lhMXCuMEM6JFnN2iC7ec6fZBOgtjCXpQU3JB51eVNRefd-Mfqs28Vm1jLZMNkvA6S4gxacZclHBZwvDoEeIc1YU845QwbtuQet31KaYcwL3eYZg9SZFvUlRH1IW_uTrb5_0hwX2Cg-HjGA</addsrcrecordid><sourcetype>Open Access Repository</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2057128577</pqid></control><display><type>article</type><title>Influence of acquisition parameters on the magnitude of cone beam computed tomography artifacts</title><source>MEDLINE</source><source>Oxford Journals Online</source><source>EZB Electronic Journals Library</source><creator>Freitas, Deborah Queiroz ; Fontenele, Rocharles Cavalcante ; Nascimento, Eduarda Helena Leandro ; Vasconcelos, Taruska Ventorini ; Noujeim, Marcel</creator><creatorcontrib>Freitas, Deborah Queiroz ; Fontenele, Rocharles Cavalcante ; Nascimento, Eduarda Helena Leandro ; Vasconcelos, Taruska Ventorini ; Noujeim, Marcel</creatorcontrib><description>To evaluate the influence of kilovoltage (kVp) and metal artifact reduction tool (MAR) on the magnitude of cone beam CT (CBCT) artifacts. A titanium and zirconia implants were inserted alternately in a posterior region of a mandible. CBCT exams were acquired with ProMax 3D (Planmeca Oy, Helsinki, Finland) and Picasso Trio machines (Vatech, Hwaseong, South Korea) using 70 kVp, 80 kVp and 90 kVp with and without MAR activation. The other exposure factors remained fixed at 5mA, field of view 80 × 50 mm and voxel 0.20 mm. The scans were performed before and after the insertion of the implants. Regions of interest were determined in different distances from the artifact production area (15, 25 and 35 mm) in an axial image, in which standard deviation (SD) of grayscale values was measured and contrast-to-noise ratio (CNR) was calculated. Analysis of variance was used to compare the data. Overall, in cases where the artifact was pronounced, MAR was efficient in reducing SD values. MAR also improved the CNR of ProMax images, but did not affect the Picasso images. Additionally, the higher was the kVp, the lower was the SD value and the higher was the CNR in both machines. In both machines, increasing kVp and MAR are effective in decreasing the CBCT artifacts in all their magnitude when they are pronounced. Therefore, the professionals should choose one of those options or even both considering the purpose of the CBCT imaging and radiation dose for the patient.</description><identifier>ISSN: 0250-832X</identifier><identifier>EISSN: 1476-542X</identifier><identifier>DOI: 10.1259/dmfr.20180151</identifier><identifier>PMID: 29916722</identifier><language>eng</language><publisher>England: The British Institute of Radiology</publisher><subject>Cone-Beam Computed Tomography ; Dentistry ; Humans ; Mandible - diagnostic imaging ; Phantoms, Imaging</subject><ispartof>Dento-maxillo-facial radiology, 2018-12, Vol.47 (8), p.20180151-20180151</ispartof><rights>2018 The Authors. Published by the British Institute of Radiology 2018 The Authors</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c387t-bfb853095bbae4d76fb7eca3c93bc0be9d85f99b19e8db034c01f68b88ff86053</citedby><cites>FETCH-LOGICAL-c387t-bfb853095bbae4d76fb7eca3c93bc0be9d85f99b19e8db034c01f68b88ff86053</cites><orcidid>0000-0002-1425-5966</orcidid></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>230,315,781,785,886,27929,27930</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/29916722$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Freitas, Deborah Queiroz</creatorcontrib><creatorcontrib>Fontenele, Rocharles Cavalcante</creatorcontrib><creatorcontrib>Nascimento, Eduarda Helena Leandro</creatorcontrib><creatorcontrib>Vasconcelos, Taruska Ventorini</creatorcontrib><creatorcontrib>Noujeim, Marcel</creatorcontrib><title>Influence of acquisition parameters on the magnitude of cone beam computed tomography artifacts</title><title>Dento-maxillo-facial radiology</title><addtitle>Dentomaxillofac Radiol</addtitle><description>To evaluate the influence of kilovoltage (kVp) and metal artifact reduction tool (MAR) on the magnitude of cone beam CT (CBCT) artifacts. A titanium and zirconia implants were inserted alternately in a posterior region of a mandible. CBCT exams were acquired with ProMax 3D (Planmeca Oy, Helsinki, Finland) and Picasso Trio machines (Vatech, Hwaseong, South Korea) using 70 kVp, 80 kVp and 90 kVp with and without MAR activation. The other exposure factors remained fixed at 5mA, field of view 80 × 50 mm and voxel 0.20 mm. The scans were performed before and after the insertion of the implants. Regions of interest were determined in different distances from the artifact production area (15, 25 and 35 mm) in an axial image, in which standard deviation (SD) of grayscale values was measured and contrast-to-noise ratio (CNR) was calculated. Analysis of variance was used to compare the data. Overall, in cases where the artifact was pronounced, MAR was efficient in reducing SD values. MAR also improved the CNR of ProMax images, but did not affect the Picasso images. Additionally, the higher was the kVp, the lower was the SD value and the higher was the CNR in both machines. In both machines, increasing kVp and MAR are effective in decreasing the CBCT artifacts in all their magnitude when they are pronounced. Therefore, the professionals should choose one of those options or even both considering the purpose of the CBCT imaging and radiation dose for the patient.</description><subject>Cone-Beam Computed Tomography</subject><subject>Dentistry</subject><subject>Humans</subject><subject>Mandible - diagnostic imaging</subject><subject>Phantoms, Imaging</subject><issn>0250-832X</issn><issn>1476-542X</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2018</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNpVkctLJDEQh8Piso6P416lj15a8-h0JxdBxBcIXnZhbiFJV2aydDptkhb8723XUfRUVdTHrwo-hH4TfEYol-d9cOmMYiIw4eQHWpGma2ve0PUeWmHKcS0YXe-jg5z_YYwbxttfaJ9KSdqO0hVS96MbZhgtVNFV2j7NPvvi41hNOukABVKulqlsoQp6M_oy9_9RG0eoDOiwdGGaC_RViSFukp62L5VOxTttSz5CP50eMhzv6iH6e3P95-qufni8vb-6fKgtE12pjTOCMyy5MRqavmud6cBqZiUzFhuQveBOSkMkiN5g1lhMXCuMEM6JFnN2iC7ec6fZBOgtjCXpQU3JB51eVNRefd-Mfqs28Vm1jLZMNkvA6S4gxacZclHBZwvDoEeIc1YU845QwbtuQet31KaYcwL3eYZg9SZFvUlRH1IW_uTrb5_0hwX2Cg-HjGA</recordid><startdate>201812</startdate><enddate>201812</enddate><creator>Freitas, Deborah Queiroz</creator><creator>Fontenele, Rocharles Cavalcante</creator><creator>Nascimento, Eduarda Helena Leandro</creator><creator>Vasconcelos, Taruska Ventorini</creator><creator>Noujeim, Marcel</creator><general>The British Institute of Radiology</general><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope><scope>5PM</scope><orcidid>https://orcid.org/0000-0002-1425-5966</orcidid></search><sort><creationdate>201812</creationdate><title>Influence of acquisition parameters on the magnitude of cone beam computed tomography artifacts</title><author>Freitas, Deborah Queiroz ; Fontenele, Rocharles Cavalcante ; Nascimento, Eduarda Helena Leandro ; Vasconcelos, Taruska Ventorini ; Noujeim, Marcel</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c387t-bfb853095bbae4d76fb7eca3c93bc0be9d85f99b19e8db034c01f68b88ff86053</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2018</creationdate><topic>Cone-Beam Computed Tomography</topic><topic>Dentistry</topic><topic>Humans</topic><topic>Mandible - diagnostic imaging</topic><topic>Phantoms, Imaging</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Freitas, Deborah Queiroz</creatorcontrib><creatorcontrib>Fontenele, Rocharles Cavalcante</creatorcontrib><creatorcontrib>Nascimento, Eduarda Helena Leandro</creatorcontrib><creatorcontrib>Vasconcelos, Taruska Ventorini</creatorcontrib><creatorcontrib>Noujeim, Marcel</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><collection>PubMed Central (Full Participant titles)</collection><jtitle>Dento-maxillo-facial radiology</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Freitas, Deborah Queiroz</au><au>Fontenele, Rocharles Cavalcante</au><au>Nascimento, Eduarda Helena Leandro</au><au>Vasconcelos, Taruska Ventorini</au><au>Noujeim, Marcel</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Influence of acquisition parameters on the magnitude of cone beam computed tomography artifacts</atitle><jtitle>Dento-maxillo-facial radiology</jtitle><addtitle>Dentomaxillofac Radiol</addtitle><date>2018-12</date><risdate>2018</risdate><volume>47</volume><issue>8</issue><spage>20180151</spage><epage>20180151</epage><pages>20180151-20180151</pages><issn>0250-832X</issn><eissn>1476-542X</eissn><abstract>To evaluate the influence of kilovoltage (kVp) and metal artifact reduction tool (MAR) on the magnitude of cone beam CT (CBCT) artifacts. A titanium and zirconia implants were inserted alternately in a posterior region of a mandible. CBCT exams were acquired with ProMax 3D (Planmeca Oy, Helsinki, Finland) and Picasso Trio machines (Vatech, Hwaseong, South Korea) using 70 kVp, 80 kVp and 90 kVp with and without MAR activation. The other exposure factors remained fixed at 5mA, field of view 80 × 50 mm and voxel 0.20 mm. The scans were performed before and after the insertion of the implants. Regions of interest were determined in different distances from the artifact production area (15, 25 and 35 mm) in an axial image, in which standard deviation (SD) of grayscale values was measured and contrast-to-noise ratio (CNR) was calculated. Analysis of variance was used to compare the data. Overall, in cases where the artifact was pronounced, MAR was efficient in reducing SD values. MAR also improved the CNR of ProMax images, but did not affect the Picasso images. Additionally, the higher was the kVp, the lower was the SD value and the higher was the CNR in both machines. In both machines, increasing kVp and MAR are effective in decreasing the CBCT artifacts in all their magnitude when they are pronounced. Therefore, the professionals should choose one of those options or even both considering the purpose of the CBCT imaging and radiation dose for the patient.</abstract><cop>England</cop><pub>The British Institute of Radiology</pub><pmid>29916722</pmid><doi>10.1259/dmfr.20180151</doi><tpages>1</tpages><orcidid>https://orcid.org/0000-0002-1425-5966</orcidid><oa>free_for_read</oa></addata></record>
fulltext fulltext
identifier ISSN: 0250-832X
ispartof Dento-maxillo-facial radiology, 2018-12, Vol.47 (8), p.20180151-20180151
issn 0250-832X
1476-542X
language eng
recordid cdi_pubmedcentral_primary_oai_pubmedcentral_nih_gov_6326394
source MEDLINE; Oxford Journals Online; EZB Electronic Journals Library
subjects Cone-Beam Computed Tomography
Dentistry
Humans
Mandible - diagnostic imaging
Phantoms, Imaging
title Influence of acquisition parameters on the magnitude of cone beam computed tomography artifacts
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2024-12-13T00%3A26%3A30IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_pubme&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Influence%20of%20acquisition%20parameters%20on%20the%20magnitude%20of%20cone%20beam%20computed%20tomography%20artifacts&rft.jtitle=Dento-maxillo-facial%20radiology&rft.au=Freitas,%20Deborah%20Queiroz&rft.date=2018-12&rft.volume=47&rft.issue=8&rft.spage=20180151&rft.epage=20180151&rft.pages=20180151-20180151&rft.issn=0250-832X&rft.eissn=1476-542X&rft_id=info:doi/10.1259/dmfr.20180151&rft_dat=%3Cproquest_pubme%3E2057128577%3C/proquest_pubme%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=2057128577&rft_id=info:pmid/29916722&rfr_iscdi=true