Large area alpha sources with a lip: Integral counting and spectral distortions
The detection efficiency for large area alpha sources with adjustable heights of a raised lip around the edge were measured by 2π gas-filled proportional counter. The variations in low-energy spectral shape were modeled using a Geant4 radiation and charge transport Monte Carlo simulation, to enable...
Gespeichert in:
Veröffentlicht in: | Applied radiation and isotopes 2018-04, Vol.134, p.376-379 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The detection efficiency for large area alpha sources with adjustable heights of a raised lip around the edge were measured by 2π gas-filled proportional counter. The variations in low-energy spectral shape were modeled using a Geant4 radiation and charge transport Monte Carlo simulation, to enable extrapolation of the spectrum to zero energy. COMSOL Multiphysics finite-element analysis was used to explore changes in the spectrum gain in the presence of a lip. It qualitatively reproduced an increase in peak height due to an increasing height of the source lip. A spectrum analysis procedure was developed to perform integral counting on sources with a lip. The experimental results were used to validate the model, which was then used to predict the changes in 2π counting efficiency for other source-lip geometries.
•Analysis method for proportional counter data for alpha sources with a lip.•Monte Carlo analysis reproduces low-energy scattering shape and efficiency.•Electric field calculations explain peak shift to higher energy. |
---|---|
ISSN: | 0969-8043 1872-9800 |
DOI: | 10.1016/j.apradiso.2017.08.030 |