Malfunction of a central venous multilumen access catheter caused by kinking: A case report

A multilumen access catheter (MAC) is a large-bore catheter that prevents kinking or collapsing. Misplacement is a frequently reported complication. A 59-year-old man who was diagnosed with a hemoperitoneum due to a liver laceration after a fall. After catheterization, we were able to aspirate blood...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Medicine (Baltimore) 2018-07, Vol.97 (30), p.e11622-e11622
Hauptverfasser: Cho, Ho Bum, Kim, Sang Hyun, Yoo, Jae Hwa, Gong, Hyung Youn, Seo, Yong Han, Park, Sun Young, Chung, Ji Won, Kim, Mun Gyu, Chung, Jin Hun, Kim, Sang Ho
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A multilumen access catheter (MAC) is a large-bore catheter that prevents kinking or collapsing. Misplacement is a frequently reported complication. A 59-year-old man who was diagnosed with a hemoperitoneum due to a liver laceration after a fall. After catheterization, we were able to aspirate blood through the 9 Fr, but not the 12-gauge line. Thus, we assumed that the catheter was misplaced. Nevertheless, ultrasonographic findings did not reveal the reason for the malfunction. We found kinking in the distal part of the catheter after removing it. The MAC was removed soon after stopping the active bleeding. A hemihepatectomy was successfully performed, and the patient was transferred to the intensive care unit. Anesthesiologists should consider kinking of large-bore catheters, including MACs.
ISSN:0025-7974
1536-5964
DOI:10.1097/MD.0000000000011622