Influence of the ultrasound transducer tilt on muscle thickness and echo intensity of the rectus femoris muscle of healthy subjects
[Purpose] This study aimed to assess the influence of ultrasound (US) transducer tilt on muscle thickness and echo intensity of the rectus femoris muscle (RF) in healthy subjects. [Subjects and Methods] Fourteen healthy male subjects (20.8 ± 0.8 years) participated in this study. Transducer tilt was...
Gespeichert in:
Veröffentlicht in: | Journal of Physical Therapy Science 2017, Vol.29(12), pp.2190-2193 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | 2193 |
---|---|
container_issue | 12 |
container_start_page | 2190 |
container_title | Journal of Physical Therapy Science |
container_volume | 29 |
creator | Ishida, Hiroshi Suehiro, Tadanobu Suzuki, Keita Yoneda, Tsuyoshi Watanabe, Susumu |
description | [Purpose] This study aimed to assess the influence of ultrasound (US) transducer tilt on muscle thickness and echo intensity of the rectus femoris muscle (RF) in healthy subjects. [Subjects and Methods] Fourteen healthy male subjects (20.8 ± 0.8 years) participated in this study. Transducer tilt was measured during US, with a digital angle gauge. Muscle thickness and echo intensity were measured in 4 transducer tilt conditions: reference angle; +3°; +6°; and +9° cranial from the reference angle. [Results] All differences in transducer tilt relative to the reference condition were larger than the minimal detectable change (MDC) of the reference condition. All differences in muscle thickness relative to the reference condition were not larger than the MDC of the reference condition. All differences in the echo intensity relative to the reference condition, except between the reference and the +3° condition, were larger than the MDC of the reference condition. [Conclusion] Our results indicated that an examiner should maintain a precise transducer tilt during repeated US measurements to quantify the minimal change in the echo intensity of the RF. |
doi_str_mv | 10.1589/jpts.29.190 |
format | Article |
fullrecord | <record><control><sourceid>proquest_pubme</sourceid><recordid>TN_cdi_pubmedcentral_primary_oai_pubmedcentral_nih_gov_5890228</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>2014584428</sourcerecordid><originalsourceid>FETCH-LOGICAL-c6450-df49dd40be842707e7c63a05794f6700fce2ef17c124857c1f57345954bf92d23</originalsourceid><addsrcrecordid>eNpdkk2P0zAQhiMEYsvCiTuyxAUJtdiOHccXJLSCpdJKXOBsuc5k4-DaxR9IPfPHcem2BS7jSPPk8UzeNM1LgleE9_LdvMtpReWKSPyoWVDSiyXvaPe4WWBJ-JLTXlw1z1KaMaYCs_5pc0Vlx9oO00Xza-1HV8AbQGFEeQJUXI46heIHVB98GoqBiLJ1GQWPtiUZBxW05ruHlJCuHJgpIOsz-GTz_iSKYHJJaIRtiDad3qzNCbTL0x6lspkrk543T0btErx4OK-bb58-fr35vLz7cru--XC3NB3jeDmMTA4DwxvoWV1EgDBdqzEXko2dwHg0QGEkwhDKel6PkYuWccnZZpR0oO118_7o3ZXNFgYDvi7o1C7arY57FbRV_3a8ndR9-KnqV8aU9lXw5kEQw48CKautTQac0x5CSYpiypgQhOOKvv4PnUOJvq5XKcJ4z9gf4dsjZWJIKcJ4HoZgdQhXHcJVVKoabqVf_T3_mT2lWYHbI1C71mgXvLMeLjcb3c9HISZC1d9BEqpqVbTqD6VtMeZYHkzro2lOWd_D-Sods60pXsaih3pxMizOjJl0VODb3w4L09M</addsrcrecordid><sourcetype>Open Access Repository</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2014584428</pqid></control><display><type>article</type><title>Influence of the ultrasound transducer tilt on muscle thickness and echo intensity of the rectus femoris muscle of healthy subjects</title><source>J-STAGE Free</source><source>EZB-FREE-00999 freely available EZB journals</source><source>PubMed Central</source><source>PubMed Central Open Access</source><creator>Ishida, Hiroshi ; Suehiro, Tadanobu ; Suzuki, Keita ; Yoneda, Tsuyoshi ; Watanabe, Susumu</creator><creatorcontrib>Ishida, Hiroshi ; Suehiro, Tadanobu ; Suzuki, Keita ; Yoneda, Tsuyoshi ; Watanabe, Susumu ; Faculty of Health Science and Technology ; Kawasaki University of Medical Welfare ; Department of Sensory Science ; Department of Rehabilitation</creatorcontrib><description>[Purpose] This study aimed to assess the influence of ultrasound (US) transducer tilt on muscle thickness and echo intensity of the rectus femoris muscle (RF) in healthy subjects. [Subjects and Methods] Fourteen healthy male subjects (20.8 ± 0.8 years) participated in this study. Transducer tilt was measured during US, with a digital angle gauge. Muscle thickness and echo intensity were measured in 4 transducer tilt conditions: reference angle; +3°; +6°; and +9° cranial from the reference angle. [Results] All differences in transducer tilt relative to the reference condition were larger than the minimal detectable change (MDC) of the reference condition. All differences in muscle thickness relative to the reference condition were not larger than the MDC of the reference condition. All differences in the echo intensity relative to the reference condition, except between the reference and the +3° condition, were larger than the MDC of the reference condition. [Conclusion] Our results indicated that an examiner should maintain a precise transducer tilt during repeated US measurements to quantify the minimal change in the echo intensity of the RF.</description><identifier>ISSN: 0915-5287</identifier><identifier>EISSN: 2187-5626</identifier><identifier>DOI: 10.1589/jpts.29.190</identifier><identifier>PMID: 29643602</identifier><language>eng</language><publisher>Japan: The Society of Physical Therapy Science</publisher><subject>Echo intensity ; Muscle thickness ; Original ; Ultrasonic imaging ; Ultrasound</subject><ispartof>Journal of Physical Therapy Science, 2017, Vol.29(12), pp.2190-2193</ispartof><rights>2017 by the Society of Physical Therapy Science. Published by IPEC Inc.</rights><rights>Copyright Japan Science and Technology Agency 2017</rights><rights>2017©by the Society of Physical Therapy Science. Published by IPEC Inc. 2017</rights><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c6450-df49dd40be842707e7c63a05794f6700fce2ef17c124857c1f57345954bf92d23</citedby><cites>FETCH-LOGICAL-c6450-df49dd40be842707e7c63a05794f6700fce2ef17c124857c1f57345954bf92d23</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktopdf>$$Uhttps://www.ncbi.nlm.nih.gov/pmc/articles/PMC5890228/pdf/$$EPDF$$P50$$Gpubmedcentral$$Hfree_for_read</linktopdf><linktohtml>$$Uhttps://www.ncbi.nlm.nih.gov/pmc/articles/PMC5890228/$$EHTML$$P50$$Gpubmedcentral$$Hfree_for_read</linktohtml><link.rule.ids>230,314,723,776,780,881,1877,4010,27900,27901,27902,53766,53768</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/29643602$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Ishida, Hiroshi</creatorcontrib><creatorcontrib>Suehiro, Tadanobu</creatorcontrib><creatorcontrib>Suzuki, Keita</creatorcontrib><creatorcontrib>Yoneda, Tsuyoshi</creatorcontrib><creatorcontrib>Watanabe, Susumu</creatorcontrib><creatorcontrib>Faculty of Health Science and Technology</creatorcontrib><creatorcontrib>Kawasaki University of Medical Welfare</creatorcontrib><creatorcontrib>Department of Sensory Science</creatorcontrib><creatorcontrib>Department of Rehabilitation</creatorcontrib><title>Influence of the ultrasound transducer tilt on muscle thickness and echo intensity of the rectus femoris muscle of healthy subjects</title><title>Journal of Physical Therapy Science</title><addtitle>Journal of Physical Therapy Science</addtitle><description>[Purpose] This study aimed to assess the influence of ultrasound (US) transducer tilt on muscle thickness and echo intensity of the rectus femoris muscle (RF) in healthy subjects. [Subjects and Methods] Fourteen healthy male subjects (20.8 ± 0.8 years) participated in this study. Transducer tilt was measured during US, with a digital angle gauge. Muscle thickness and echo intensity were measured in 4 transducer tilt conditions: reference angle; +3°; +6°; and +9° cranial from the reference angle. [Results] All differences in transducer tilt relative to the reference condition were larger than the minimal detectable change (MDC) of the reference condition. All differences in muscle thickness relative to the reference condition were not larger than the MDC of the reference condition. All differences in the echo intensity relative to the reference condition, except between the reference and the +3° condition, were larger than the MDC of the reference condition. [Conclusion] Our results indicated that an examiner should maintain a precise transducer tilt during repeated US measurements to quantify the minimal change in the echo intensity of the RF.</description><subject>Echo intensity</subject><subject>Muscle thickness</subject><subject>Original</subject><subject>Ultrasonic imaging</subject><subject>Ultrasound</subject><issn>0915-5287</issn><issn>2187-5626</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2017</creationdate><recordtype>article</recordtype><recordid>eNpdkk2P0zAQhiMEYsvCiTuyxAUJtdiOHccXJLSCpdJKXOBsuc5k4-DaxR9IPfPHcem2BS7jSPPk8UzeNM1LgleE9_LdvMtpReWKSPyoWVDSiyXvaPe4WWBJ-JLTXlw1z1KaMaYCs_5pc0Vlx9oO00Xza-1HV8AbQGFEeQJUXI46heIHVB98GoqBiLJ1GQWPtiUZBxW05ruHlJCuHJgpIOsz-GTz_iSKYHJJaIRtiDad3qzNCbTL0x6lspkrk543T0btErx4OK-bb58-fr35vLz7cru--XC3NB3jeDmMTA4DwxvoWV1EgDBdqzEXko2dwHg0QGEkwhDKel6PkYuWccnZZpR0oO118_7o3ZXNFgYDvi7o1C7arY57FbRV_3a8ndR9-KnqV8aU9lXw5kEQw48CKautTQac0x5CSYpiypgQhOOKvv4PnUOJvq5XKcJ4z9gf4dsjZWJIKcJ4HoZgdQhXHcJVVKoabqVf_T3_mT2lWYHbI1C71mgXvLMeLjcb3c9HISZC1d9BEqpqVbTqD6VtMeZYHkzro2lOWd_D-Sods60pXsaih3pxMizOjJl0VODb3w4L09M</recordid><startdate>2017</startdate><enddate>2017</enddate><creator>Ishida, Hiroshi</creator><creator>Suehiro, Tadanobu</creator><creator>Suzuki, Keita</creator><creator>Yoneda, Tsuyoshi</creator><creator>Watanabe, Susumu</creator><general>The Society of Physical Therapy Science</general><general>Japan Science and Technology Agency</general><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7TS</scope><scope>NAPCQ</scope><scope>7X8</scope><scope>5PM</scope></search><sort><creationdate>2017</creationdate><title>Influence of the ultrasound transducer tilt on muscle thickness and echo intensity of the rectus femoris muscle of healthy subjects</title><author>Ishida, Hiroshi ; Suehiro, Tadanobu ; Suzuki, Keita ; Yoneda, Tsuyoshi ; Watanabe, Susumu</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c6450-df49dd40be842707e7c63a05794f6700fce2ef17c124857c1f57345954bf92d23</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2017</creationdate><topic>Echo intensity</topic><topic>Muscle thickness</topic><topic>Original</topic><topic>Ultrasonic imaging</topic><topic>Ultrasound</topic><toplevel>online_resources</toplevel><creatorcontrib>Ishida, Hiroshi</creatorcontrib><creatorcontrib>Suehiro, Tadanobu</creatorcontrib><creatorcontrib>Suzuki, Keita</creatorcontrib><creatorcontrib>Yoneda, Tsuyoshi</creatorcontrib><creatorcontrib>Watanabe, Susumu</creatorcontrib><creatorcontrib>Faculty of Health Science and Technology</creatorcontrib><creatorcontrib>Kawasaki University of Medical Welfare</creatorcontrib><creatorcontrib>Department of Sensory Science</creatorcontrib><creatorcontrib>Department of Rehabilitation</creatorcontrib><collection>PubMed</collection><collection>CrossRef</collection><collection>Physical Education Index</collection><collection>Nursing & Allied Health Premium</collection><collection>MEDLINE - Academic</collection><collection>PubMed Central (Full Participant titles)</collection><jtitle>Journal of Physical Therapy Science</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Ishida, Hiroshi</au><au>Suehiro, Tadanobu</au><au>Suzuki, Keita</au><au>Yoneda, Tsuyoshi</au><au>Watanabe, Susumu</au><aucorp>Faculty of Health Science and Technology</aucorp><aucorp>Kawasaki University of Medical Welfare</aucorp><aucorp>Department of Sensory Science</aucorp><aucorp>Department of Rehabilitation</aucorp><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Influence of the ultrasound transducer tilt on muscle thickness and echo intensity of the rectus femoris muscle of healthy subjects</atitle><jtitle>Journal of Physical Therapy Science</jtitle><addtitle>Journal of Physical Therapy Science</addtitle><date>2017</date><risdate>2017</risdate><volume>29</volume><issue>12</issue><spage>2190</spage><epage>2193</epage><pages>2190-2193</pages><issn>0915-5287</issn><eissn>2187-5626</eissn><abstract>[Purpose] This study aimed to assess the influence of ultrasound (US) transducer tilt on muscle thickness and echo intensity of the rectus femoris muscle (RF) in healthy subjects. [Subjects and Methods] Fourteen healthy male subjects (20.8 ± 0.8 years) participated in this study. Transducer tilt was measured during US, with a digital angle gauge. Muscle thickness and echo intensity were measured in 4 transducer tilt conditions: reference angle; +3°; +6°; and +9° cranial from the reference angle. [Results] All differences in transducer tilt relative to the reference condition were larger than the minimal detectable change (MDC) of the reference condition. All differences in muscle thickness relative to the reference condition were not larger than the MDC of the reference condition. All differences in the echo intensity relative to the reference condition, except between the reference and the +3° condition, were larger than the MDC of the reference condition. [Conclusion] Our results indicated that an examiner should maintain a precise transducer tilt during repeated US measurements to quantify the minimal change in the echo intensity of the RF.</abstract><cop>Japan</cop><pub>The Society of Physical Therapy Science</pub><pmid>29643602</pmid><doi>10.1589/jpts.29.190</doi><tpages>4</tpages><oa>free_for_read</oa></addata></record> |
fulltext | fulltext |
identifier | ISSN: 0915-5287 |
ispartof | Journal of Physical Therapy Science, 2017, Vol.29(12), pp.2190-2193 |
issn | 0915-5287 2187-5626 |
language | eng |
recordid | cdi_pubmedcentral_primary_oai_pubmedcentral_nih_gov_5890228 |
source | J-STAGE Free; EZB-FREE-00999 freely available EZB journals; PubMed Central; PubMed Central Open Access |
subjects | Echo intensity Muscle thickness Original Ultrasonic imaging Ultrasound |
title | Influence of the ultrasound transducer tilt on muscle thickness and echo intensity of the rectus femoris muscle of healthy subjects |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-02-12T11%3A41%3A36IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_pubme&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Influence%20of%20the%20ultrasound%20transducer%20tilt%20on%20muscle%20thickness%20and%20echo%20intensity%20of%20the%20rectus%20femoris%20muscle%20of%20healthy%20subjects&rft.jtitle=Journal%20of%20Physical%20Therapy%20Science&rft.au=Ishida,%20Hiroshi&rft.aucorp=Faculty%20of%20Health%20Science%20and%20Technology&rft.date=2017&rft.volume=29&rft.issue=12&rft.spage=2190&rft.epage=2193&rft.pages=2190-2193&rft.issn=0915-5287&rft.eissn=2187-5626&rft_id=info:doi/10.1589/jpts.29.190&rft_dat=%3Cproquest_pubme%3E2014584428%3C/proquest_pubme%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=2014584428&rft_id=info:pmid/29643602&rfr_iscdi=true |