A Simple Method for 2-D In Vivo Dosimetry by Portal Imaging
Purpose: To improve patient safety and treatment quality, verification of dose delivery in radiotherapy is desirable. We present a simple, easy-to-implement, open-source method for in vivo planar dosimetry of conformal radiotherapy by electronic portal imaging device (EPID). Methods: Correlation rat...
Gespeichert in:
Veröffentlicht in: | Technology in cancer research & treatment 2017-12, Vol.16 (6), p.944-955 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | 955 |
---|---|
container_issue | 6 |
container_start_page | 944 |
container_title | Technology in cancer research & treatment |
container_volume | 16 |
creator | Peca, Stefano Brown, Derek Wilson Smith, Wendy Lani |
description | Purpose:
To improve patient safety and treatment quality, verification of dose delivery in radiotherapy is desirable. We present a simple, easy-to-implement, open-source method for in vivo planar dosimetry of conformal radiotherapy by electronic portal imaging device (EPID).
Methods:
Correlation ratios, which relate dose in the mid-depth of slab phantoms to transit EPID signal, were determined for multiple phantom thicknesses and field sizes. Off-axis dose is corrected for by means of model-based convolution. We tested efficacy of dose reconstruction through measurements with off-reference values of attenuator thickness, field size, and monitor units. We quantified the dose calculation error in the presence of thickness changes to simulate anatomical or setup variations. An example of dose calculation on patient data is provided.
Results:
With varying phantom thickness, field size, and monitor units, dose reconstruction was almost always within 3% of planned dose. In the presence of thickness changes from planning CT, the dose discrepancy is exaggerated by up to approximately 1.5% for 1 cm changes upstream of the isocenter plane and 4% for 1 cm changes downstream.
Conclusion:
Our novel electronic portal imaging device in vivo dosimetry allows clinically accurate 2-dimensional reconstruction of dose inside a phantom/patient at isocenter depth. Due to its simplicity, commissioning can be performed in a few hours per energy and may be modified to the user’s needs. It may provide useful dose delivery information to detect harmful errors, guide adaptive radiotherapy, and assure quality of treatment. |
doi_str_mv | 10.1177/1533034617711354 |
format | Article |
fullrecord | <record><control><sourceid>proquest_pubme</sourceid><recordid>TN_cdi_pubmedcentral_primary_oai_pubmedcentral_nih_gov_5762053</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sage_id>10.1177_1533034617711354</sage_id><sourcerecordid>1906467463</sourcerecordid><originalsourceid>FETCH-LOGICAL-c434t-e63130a32b7c3f109f399e9c76412776cefd222c4e44e08339f0f42aa9377f0f3</originalsourceid><addsrcrecordid>eNp1kMtLAzEQxoMotlbvniRHL6t5bdIgCKX1Uago-LiGNM1uU3Y3Ndkt9L93S2tRwdN8zHzzzfAD4ByjK4yFuMYppYgy3mqMacoOQHfTShCl_cO9ZrwDTmJcIEQ4p_gYdEg_7adM4i64GcBXVy4LC59sPfczmPkASTKC4wp-uJWHIx9daeuwhtM1fPGh1gUclzp3VX4KjjJdRHu2qz3wfn_3NnxMJs8P4-FgkhhGWZ3Y9iZFmpKpMDTDSGZUSiuN4AwTIbix2YwQYphlzKI-pTJDGSNaSypEK2kP3G5zl820tDNjqzroQi2DK3VYK6-d-j2p3FzlfqVSwQlqGfTA5S4g-M_GxlqVLhpbFLqyvokKS8QZF4xvrGhrNcHHGGy2P4OR2jBXf5m3Kxc_39svfENuDcnWEHVu1cI3oWpx_R_4BaArhvM</addsrcrecordid><sourcetype>Open Access Repository</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>1906467463</pqid></control><display><type>article</type><title>A Simple Method for 2-D In Vivo Dosimetry by Portal Imaging</title><source>Sage Journals GOLD Open Access 2024</source><source>EZB-FREE-00999 freely available EZB journals</source><source>PubMed Central</source><creator>Peca, Stefano ; Brown, Derek Wilson ; Smith, Wendy Lani</creator><creatorcontrib>Peca, Stefano ; Brown, Derek Wilson ; Smith, Wendy Lani</creatorcontrib><description>Purpose:
To improve patient safety and treatment quality, verification of dose delivery in radiotherapy is desirable. We present a simple, easy-to-implement, open-source method for in vivo planar dosimetry of conformal radiotherapy by electronic portal imaging device (EPID).
Methods:
Correlation ratios, which relate dose in the mid-depth of slab phantoms to transit EPID signal, were determined for multiple phantom thicknesses and field sizes. Off-axis dose is corrected for by means of model-based convolution. We tested efficacy of dose reconstruction through measurements with off-reference values of attenuator thickness, field size, and monitor units. We quantified the dose calculation error in the presence of thickness changes to simulate anatomical or setup variations. An example of dose calculation on patient data is provided.
Results:
With varying phantom thickness, field size, and monitor units, dose reconstruction was almost always within 3% of planned dose. In the presence of thickness changes from planning CT, the dose discrepancy is exaggerated by up to approximately 1.5% for 1 cm changes upstream of the isocenter plane and 4% for 1 cm changes downstream.
Conclusion:
Our novel electronic portal imaging device in vivo dosimetry allows clinically accurate 2-dimensional reconstruction of dose inside a phantom/patient at isocenter depth. Due to its simplicity, commissioning can be performed in a few hours per energy and may be modified to the user’s needs. It may provide useful dose delivery information to detect harmful errors, guide adaptive radiotherapy, and assure quality of treatment.</description><identifier>ISSN: 1533-0346</identifier><identifier>EISSN: 1533-0338</identifier><identifier>DOI: 10.1177/1533034617711354</identifier><identifier>PMID: 28585491</identifier><language>eng</language><publisher>Los Angeles, CA: SAGE Publications</publisher><subject>Original</subject><ispartof>Technology in cancer research & treatment, 2017-12, Vol.16 (6), p.944-955</ispartof><rights>The Author(s) 2017</rights><rights>The Author(s) 2017 2017 SAGE Publications</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c434t-e63130a32b7c3f109f399e9c76412776cefd222c4e44e08339f0f42aa9377f0f3</citedby><cites>FETCH-LOGICAL-c434t-e63130a32b7c3f109f399e9c76412776cefd222c4e44e08339f0f42aa9377f0f3</cites><orcidid>0000-0001-5940-6046</orcidid></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktopdf>$$Uhttps://www.ncbi.nlm.nih.gov/pmc/articles/PMC5762053/pdf/$$EPDF$$P50$$Gpubmedcentral$$Hfree_for_read</linktopdf><linktohtml>$$Uhttps://www.ncbi.nlm.nih.gov/pmc/articles/PMC5762053/$$EHTML$$P50$$Gpubmedcentral$$Hfree_for_read</linktohtml><link.rule.ids>230,315,728,781,785,886,21971,27858,27929,27930,44950,45338,53796,53798</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/28585491$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Peca, Stefano</creatorcontrib><creatorcontrib>Brown, Derek Wilson</creatorcontrib><creatorcontrib>Smith, Wendy Lani</creatorcontrib><title>A Simple Method for 2-D In Vivo Dosimetry by Portal Imaging</title><title>Technology in cancer research & treatment</title><addtitle>Technol Cancer Res Treat</addtitle><description>Purpose:
To improve patient safety and treatment quality, verification of dose delivery in radiotherapy is desirable. We present a simple, easy-to-implement, open-source method for in vivo planar dosimetry of conformal radiotherapy by electronic portal imaging device (EPID).
Methods:
Correlation ratios, which relate dose in the mid-depth of slab phantoms to transit EPID signal, were determined for multiple phantom thicknesses and field sizes. Off-axis dose is corrected for by means of model-based convolution. We tested efficacy of dose reconstruction through measurements with off-reference values of attenuator thickness, field size, and monitor units. We quantified the dose calculation error in the presence of thickness changes to simulate anatomical or setup variations. An example of dose calculation on patient data is provided.
Results:
With varying phantom thickness, field size, and monitor units, dose reconstruction was almost always within 3% of planned dose. In the presence of thickness changes from planning CT, the dose discrepancy is exaggerated by up to approximately 1.5% for 1 cm changes upstream of the isocenter plane and 4% for 1 cm changes downstream.
Conclusion:
Our novel electronic portal imaging device in vivo dosimetry allows clinically accurate 2-dimensional reconstruction of dose inside a phantom/patient at isocenter depth. Due to its simplicity, commissioning can be performed in a few hours per energy and may be modified to the user’s needs. It may provide useful dose delivery information to detect harmful errors, guide adaptive radiotherapy, and assure quality of treatment.</description><subject>Original</subject><issn>1533-0346</issn><issn>1533-0338</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2017</creationdate><recordtype>article</recordtype><sourceid>AFRWT</sourceid><recordid>eNp1kMtLAzEQxoMotlbvniRHL6t5bdIgCKX1Uago-LiGNM1uU3Y3Ndkt9L93S2tRwdN8zHzzzfAD4ByjK4yFuMYppYgy3mqMacoOQHfTShCl_cO9ZrwDTmJcIEQ4p_gYdEg_7adM4i64GcBXVy4LC59sPfczmPkASTKC4wp-uJWHIx9daeuwhtM1fPGh1gUclzp3VX4KjjJdRHu2qz3wfn_3NnxMJs8P4-FgkhhGWZ3Y9iZFmpKpMDTDSGZUSiuN4AwTIbix2YwQYphlzKI-pTJDGSNaSypEK2kP3G5zl820tDNjqzroQi2DK3VYK6-d-j2p3FzlfqVSwQlqGfTA5S4g-M_GxlqVLhpbFLqyvokKS8QZF4xvrGhrNcHHGGy2P4OR2jBXf5m3Kxc_39svfENuDcnWEHVu1cI3oWpx_R_4BaArhvM</recordid><startdate>20171201</startdate><enddate>20171201</enddate><creator>Peca, Stefano</creator><creator>Brown, Derek Wilson</creator><creator>Smith, Wendy Lani</creator><general>SAGE Publications</general><scope>AFRWT</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope><scope>5PM</scope><orcidid>https://orcid.org/0000-0001-5940-6046</orcidid></search><sort><creationdate>20171201</creationdate><title>A Simple Method for 2-D In Vivo Dosimetry by Portal Imaging</title><author>Peca, Stefano ; Brown, Derek Wilson ; Smith, Wendy Lani</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c434t-e63130a32b7c3f109f399e9c76412776cefd222c4e44e08339f0f42aa9377f0f3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2017</creationdate><topic>Original</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Peca, Stefano</creatorcontrib><creatorcontrib>Brown, Derek Wilson</creatorcontrib><creatorcontrib>Smith, Wendy Lani</creatorcontrib><collection>Sage Journals GOLD Open Access 2024</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><collection>PubMed Central (Full Participant titles)</collection><jtitle>Technology in cancer research & treatment</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Peca, Stefano</au><au>Brown, Derek Wilson</au><au>Smith, Wendy Lani</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>A Simple Method for 2-D In Vivo Dosimetry by Portal Imaging</atitle><jtitle>Technology in cancer research & treatment</jtitle><addtitle>Technol Cancer Res Treat</addtitle><date>2017-12-01</date><risdate>2017</risdate><volume>16</volume><issue>6</issue><spage>944</spage><epage>955</epage><pages>944-955</pages><issn>1533-0346</issn><eissn>1533-0338</eissn><abstract>Purpose:
To improve patient safety and treatment quality, verification of dose delivery in radiotherapy is desirable. We present a simple, easy-to-implement, open-source method for in vivo planar dosimetry of conformal radiotherapy by electronic portal imaging device (EPID).
Methods:
Correlation ratios, which relate dose in the mid-depth of slab phantoms to transit EPID signal, were determined for multiple phantom thicknesses and field sizes. Off-axis dose is corrected for by means of model-based convolution. We tested efficacy of dose reconstruction through measurements with off-reference values of attenuator thickness, field size, and monitor units. We quantified the dose calculation error in the presence of thickness changes to simulate anatomical or setup variations. An example of dose calculation on patient data is provided.
Results:
With varying phantom thickness, field size, and monitor units, dose reconstruction was almost always within 3% of planned dose. In the presence of thickness changes from planning CT, the dose discrepancy is exaggerated by up to approximately 1.5% for 1 cm changes upstream of the isocenter plane and 4% for 1 cm changes downstream.
Conclusion:
Our novel electronic portal imaging device in vivo dosimetry allows clinically accurate 2-dimensional reconstruction of dose inside a phantom/patient at isocenter depth. Due to its simplicity, commissioning can be performed in a few hours per energy and may be modified to the user’s needs. It may provide useful dose delivery information to detect harmful errors, guide adaptive radiotherapy, and assure quality of treatment.</abstract><cop>Los Angeles, CA</cop><pub>SAGE Publications</pub><pmid>28585491</pmid><doi>10.1177/1533034617711354</doi><tpages>12</tpages><orcidid>https://orcid.org/0000-0001-5940-6046</orcidid><oa>free_for_read</oa></addata></record> |
fulltext | fulltext |
identifier | ISSN: 1533-0346 |
ispartof | Technology in cancer research & treatment, 2017-12, Vol.16 (6), p.944-955 |
issn | 1533-0346 1533-0338 |
language | eng |
recordid | cdi_pubmedcentral_primary_oai_pubmedcentral_nih_gov_5762053 |
source | Sage Journals GOLD Open Access 2024; EZB-FREE-00999 freely available EZB journals; PubMed Central |
subjects | Original |
title | A Simple Method for 2-D In Vivo Dosimetry by Portal Imaging |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2024-12-13T10%3A46%3A13IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_pubme&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=A%20Simple%20Method%20for%202-D%20In%20Vivo%20Dosimetry%20by%20Portal%20Imaging&rft.jtitle=Technology%20in%20cancer%20research%20&%20treatment&rft.au=Peca,%20Stefano&rft.date=2017-12-01&rft.volume=16&rft.issue=6&rft.spage=944&rft.epage=955&rft.pages=944-955&rft.issn=1533-0346&rft.eissn=1533-0338&rft_id=info:doi/10.1177/1533034617711354&rft_dat=%3Cproquest_pubme%3E1906467463%3C/proquest_pubme%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=1906467463&rft_id=info:pmid/28585491&rft_sage_id=10.1177_1533034617711354&rfr_iscdi=true |