Life‐threatening and life‐saving inappropriate implantable cardioverter defibrillator shocks
Key Clinical Message An implantable cardioverter defibrillator (ICD) lead dislodgement into the right atrium is a dangerous situation, particularly in patients in atrial fibrillation because atrial fibrillation can be sensed as ventricular fibrillation and true ventricular fibrillation induced with...
Gespeichert in:
Veröffentlicht in: | Clinical case reports 2017-04, Vol.5 (4), p.521-525 |
---|---|
Hauptverfasser: | , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | 525 |
---|---|
container_issue | 4 |
container_start_page | 521 |
container_title | Clinical case reports |
container_volume | 5 |
creator | Pérez‐Rodon, Jordi Doiny, David Miranda, Berta Rivas‐Gandara, Nuria Roca‐Luque, Ivo Francisco‐Pascual, Jaume Lidón, Rosa Maria García‐Dorado, David Moya Mitjans, Angel |
description | Key Clinical Message
An implantable cardioverter defibrillator (ICD) lead dislodgement into the right atrium is a dangerous situation, particularly in patients in atrial fibrillation because atrial fibrillation can be sensed as ventricular fibrillation and true ventricular fibrillation induced with an inappropriate shock. In the presence of shocks, ICD interrogation should be performed as soon as possible.
An implantable cardioverter defibrillator (ICD) lead dislodgement into the right atrium is a dangerous situation, particularly in patients in atrial fibrillation because atrial fibrillation can be sensed as ventricular fibrillation and true ventricular fibrillation induced with an inappropriate shock. In the presence of shocks, ICD interrogation should be performed as soon as possible. |
doi_str_mv | 10.1002/ccr3.893 |
format | Article |
fullrecord | <record><control><sourceid>proquest_pubme</sourceid><recordid>TN_cdi_pubmedcentral_primary_oai_pubmedcentral_nih_gov_5378847</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>2290261033</sourcerecordid><originalsourceid>FETCH-LOGICAL-c3663-43ad85b54dedc6412569364fcf8d2b13d4a431d83ae677afbb5b46b779799853</originalsourceid><addsrcrecordid>eNp1kc1KAzEYRYMotmjBJ5ABN25G8zdJZiNI8Q8KgriPySRjo9PJmEwr7nwEn9EnMaW11IWrhC-Hk8t3AThC8AxBiM-rKpAzUZIdMMSwgDksId3dug_AKMYXCCGCHBcI7oMBFqRkXKAheJq42n5_fvXTYFVvW9c-Z6o1WbMaR7VYTlyrui74LrjEZG7WNartlW5sVqlgnF_Y0NuQGVs7HVzTqN6HLE599RoPwV6tmmhH6_MAPF5fPY5v88n9zd34cpJXhDGSU6KMKHRBjTUVowgXrCSM1lUtDNaIGKooQUYQZRnnqta60JRpzktelqIgB-Bipe3mepYUtu2DamRKPFPhQ3rl5N-X1k3ls1_IgnAhKE-Ck7Ug-Le5jb188fPQpsgS4xJihiAhiTpdUVXwMQZbb35AUC7bkMs2ZGojocfbiTbg7-4TkK-Ad9fYj39Fcjx-IEvhD43Nl90</addsrcrecordid><sourcetype>Open Access Repository</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2290261033</pqid></control><display><type>article</type><title>Life‐threatening and life‐saving inappropriate implantable cardioverter defibrillator shocks</title><source>DOAJ Directory of Open Access Journals</source><source>Elektronische Zeitschriftenbibliothek - Frei zugängliche E-Journals</source><source>Access via Wiley Online Library</source><source>Wiley Online Library (Open Access Collection)</source><source>PubMed Central</source><creator>Pérez‐Rodon, Jordi ; Doiny, David ; Miranda, Berta ; Rivas‐Gandara, Nuria ; Roca‐Luque, Ivo ; Francisco‐Pascual, Jaume ; Lidón, Rosa Maria ; García‐Dorado, David ; Moya Mitjans, Angel</creator><creatorcontrib>Pérez‐Rodon, Jordi ; Doiny, David ; Miranda, Berta ; Rivas‐Gandara, Nuria ; Roca‐Luque, Ivo ; Francisco‐Pascual, Jaume ; Lidón, Rosa Maria ; García‐Dorado, David ; Moya Mitjans, Angel</creatorcontrib><description>Key Clinical Message
An implantable cardioverter defibrillator (ICD) lead dislodgement into the right atrium is a dangerous situation, particularly in patients in atrial fibrillation because atrial fibrillation can be sensed as ventricular fibrillation and true ventricular fibrillation induced with an inappropriate shock. In the presence of shocks, ICD interrogation should be performed as soon as possible.
An implantable cardioverter defibrillator (ICD) lead dislodgement into the right atrium is a dangerous situation, particularly in patients in atrial fibrillation because atrial fibrillation can be sensed as ventricular fibrillation and true ventricular fibrillation induced with an inappropriate shock. In the presence of shocks, ICD interrogation should be performed as soon as possible.</description><identifier>ISSN: 2050-0904</identifier><identifier>EISSN: 2050-0904</identifier><identifier>DOI: 10.1002/ccr3.893</identifier><identifier>PMID: 28396781</identifier><language>eng</language><publisher>England: John Wiley & Sons, Inc</publisher><subject>Atrial fibrillation ; Cardiac arrhythmia ; Cardiovascular disease ; Case Report ; Case Reports ; Consciousness ; Defibrillators ; Disease prevention ; Electrocardiography ; implantable cardioverter defibrillator ; inappropriate shock ; lead dislodgement ; Patients ; Prostheses ; Quality of life ; Sinuses ; ventricular fibrillation</subject><ispartof>Clinical case reports, 2017-04, Vol.5 (4), p.521-525</ispartof><rights>2017 The Authors. published by John Wiley & Sons Ltd.</rights><rights>2017. This work is published under http://creativecommons.org/licenses/by/4.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><cites>FETCH-LOGICAL-c3663-43ad85b54dedc6412569364fcf8d2b13d4a431d83ae677afbb5b46b779799853</cites><orcidid>0000-0003-2234-5248</orcidid></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktopdf>$$Uhttps://www.ncbi.nlm.nih.gov/pmc/articles/PMC5378847/pdf/$$EPDF$$P50$$Gpubmedcentral$$Hfree_for_read</linktopdf><linktohtml>$$Uhttps://www.ncbi.nlm.nih.gov/pmc/articles/PMC5378847/$$EHTML$$P50$$Gpubmedcentral$$Hfree_for_read</linktohtml><link.rule.ids>230,314,727,780,784,864,885,1417,11562,27924,27925,45574,45575,46052,46476,53791,53793</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/28396781$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Pérez‐Rodon, Jordi</creatorcontrib><creatorcontrib>Doiny, David</creatorcontrib><creatorcontrib>Miranda, Berta</creatorcontrib><creatorcontrib>Rivas‐Gandara, Nuria</creatorcontrib><creatorcontrib>Roca‐Luque, Ivo</creatorcontrib><creatorcontrib>Francisco‐Pascual, Jaume</creatorcontrib><creatorcontrib>Lidón, Rosa Maria</creatorcontrib><creatorcontrib>García‐Dorado, David</creatorcontrib><creatorcontrib>Moya Mitjans, Angel</creatorcontrib><title>Life‐threatening and life‐saving inappropriate implantable cardioverter defibrillator shocks</title><title>Clinical case reports</title><addtitle>Clin Case Rep</addtitle><description>Key Clinical Message
An implantable cardioverter defibrillator (ICD) lead dislodgement into the right atrium is a dangerous situation, particularly in patients in atrial fibrillation because atrial fibrillation can be sensed as ventricular fibrillation and true ventricular fibrillation induced with an inappropriate shock. In the presence of shocks, ICD interrogation should be performed as soon as possible.
An implantable cardioverter defibrillator (ICD) lead dislodgement into the right atrium is a dangerous situation, particularly in patients in atrial fibrillation because atrial fibrillation can be sensed as ventricular fibrillation and true ventricular fibrillation induced with an inappropriate shock. In the presence of shocks, ICD interrogation should be performed as soon as possible.</description><subject>Atrial fibrillation</subject><subject>Cardiac arrhythmia</subject><subject>Cardiovascular disease</subject><subject>Case Report</subject><subject>Case Reports</subject><subject>Consciousness</subject><subject>Defibrillators</subject><subject>Disease prevention</subject><subject>Electrocardiography</subject><subject>implantable cardioverter defibrillator</subject><subject>inappropriate shock</subject><subject>lead dislodgement</subject><subject>Patients</subject><subject>Prostheses</subject><subject>Quality of life</subject><subject>Sinuses</subject><subject>ventricular fibrillation</subject><issn>2050-0904</issn><issn>2050-0904</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2017</creationdate><recordtype>article</recordtype><sourceid>24P</sourceid><sourceid>WIN</sourceid><sourceid>ABUWG</sourceid><sourceid>AFKRA</sourceid><sourceid>AZQEC</sourceid><sourceid>BENPR</sourceid><sourceid>CCPQU</sourceid><sourceid>DWQXO</sourceid><recordid>eNp1kc1KAzEYRYMotmjBJ5ABN25G8zdJZiNI8Q8KgriPySRjo9PJmEwr7nwEn9EnMaW11IWrhC-Hk8t3AThC8AxBiM-rKpAzUZIdMMSwgDksId3dug_AKMYXCCGCHBcI7oMBFqRkXKAheJq42n5_fvXTYFVvW9c-Z6o1WbMaR7VYTlyrui74LrjEZG7WNartlW5sVqlgnF_Y0NuQGVs7HVzTqN6HLE599RoPwV6tmmhH6_MAPF5fPY5v88n9zd34cpJXhDGSU6KMKHRBjTUVowgXrCSM1lUtDNaIGKooQUYQZRnnqta60JRpzktelqIgB-Bipe3mepYUtu2DamRKPFPhQ3rl5N-X1k3ls1_IgnAhKE-Ck7Ug-Le5jb188fPQpsgS4xJihiAhiTpdUVXwMQZbb35AUC7bkMs2ZGojocfbiTbg7-4TkK-Ad9fYj39Fcjx-IEvhD43Nl90</recordid><startdate>201704</startdate><enddate>201704</enddate><creator>Pérez‐Rodon, Jordi</creator><creator>Doiny, David</creator><creator>Miranda, Berta</creator><creator>Rivas‐Gandara, Nuria</creator><creator>Roca‐Luque, Ivo</creator><creator>Francisco‐Pascual, Jaume</creator><creator>Lidón, Rosa Maria</creator><creator>García‐Dorado, David</creator><creator>Moya Mitjans, Angel</creator><general>John Wiley & Sons, Inc</general><general>John Wiley and Sons Inc</general><scope>24P</scope><scope>WIN</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>3V.</scope><scope>7X7</scope><scope>7XB</scope><scope>8FI</scope><scope>8FJ</scope><scope>8FK</scope><scope>ABUWG</scope><scope>AFKRA</scope><scope>AZQEC</scope><scope>BENPR</scope><scope>CCPQU</scope><scope>DWQXO</scope><scope>FYUFA</scope><scope>GHDGH</scope><scope>K9.</scope><scope>M0S</scope><scope>PIMPY</scope><scope>PQEST</scope><scope>PQQKQ</scope><scope>PQUKI</scope><scope>PRINS</scope><scope>5PM</scope><orcidid>https://orcid.org/0000-0003-2234-5248</orcidid></search><sort><creationdate>201704</creationdate><title>Life‐threatening and life‐saving inappropriate implantable cardioverter defibrillator shocks</title><author>Pérez‐Rodon, Jordi ; Doiny, David ; Miranda, Berta ; Rivas‐Gandara, Nuria ; Roca‐Luque, Ivo ; Francisco‐Pascual, Jaume ; Lidón, Rosa Maria ; García‐Dorado, David ; Moya Mitjans, Angel</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c3663-43ad85b54dedc6412569364fcf8d2b13d4a431d83ae677afbb5b46b779799853</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2017</creationdate><topic>Atrial fibrillation</topic><topic>Cardiac arrhythmia</topic><topic>Cardiovascular disease</topic><topic>Case Report</topic><topic>Case Reports</topic><topic>Consciousness</topic><topic>Defibrillators</topic><topic>Disease prevention</topic><topic>Electrocardiography</topic><topic>implantable cardioverter defibrillator</topic><topic>inappropriate shock</topic><topic>lead dislodgement</topic><topic>Patients</topic><topic>Prostheses</topic><topic>Quality of life</topic><topic>Sinuses</topic><topic>ventricular fibrillation</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Pérez‐Rodon, Jordi</creatorcontrib><creatorcontrib>Doiny, David</creatorcontrib><creatorcontrib>Miranda, Berta</creatorcontrib><creatorcontrib>Rivas‐Gandara, Nuria</creatorcontrib><creatorcontrib>Roca‐Luque, Ivo</creatorcontrib><creatorcontrib>Francisco‐Pascual, Jaume</creatorcontrib><creatorcontrib>Lidón, Rosa Maria</creatorcontrib><creatorcontrib>García‐Dorado, David</creatorcontrib><creatorcontrib>Moya Mitjans, Angel</creatorcontrib><collection>Wiley Online Library (Open Access Collection)</collection><collection>Wiley Online Library (Open Access Collection)</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>ProQuest Central (Corporate)</collection><collection>Health & Medical Collection</collection><collection>ProQuest Central (purchase pre-March 2016)</collection><collection>Hospital Premium Collection</collection><collection>Hospital Premium Collection (Alumni Edition)</collection><collection>ProQuest Central (Alumni) (purchase pre-March 2016)</collection><collection>ProQuest Central (Alumni Edition)</collection><collection>ProQuest Central UK/Ireland</collection><collection>ProQuest Central Essentials</collection><collection>ProQuest Central</collection><collection>ProQuest One Community College</collection><collection>ProQuest Central Korea</collection><collection>Health Research Premium Collection</collection><collection>Health Research Premium Collection (Alumni)</collection><collection>ProQuest Health & Medical Complete (Alumni)</collection><collection>Health & Medical Collection (Alumni Edition)</collection><collection>Publicly Available Content Database</collection><collection>ProQuest One Academic Eastern Edition (DO NOT USE)</collection><collection>ProQuest One Academic</collection><collection>ProQuest One Academic UKI Edition</collection><collection>ProQuest Central China</collection><collection>PubMed Central (Full Participant titles)</collection><jtitle>Clinical case reports</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Pérez‐Rodon, Jordi</au><au>Doiny, David</au><au>Miranda, Berta</au><au>Rivas‐Gandara, Nuria</au><au>Roca‐Luque, Ivo</au><au>Francisco‐Pascual, Jaume</au><au>Lidón, Rosa Maria</au><au>García‐Dorado, David</au><au>Moya Mitjans, Angel</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Life‐threatening and life‐saving inappropriate implantable cardioverter defibrillator shocks</atitle><jtitle>Clinical case reports</jtitle><addtitle>Clin Case Rep</addtitle><date>2017-04</date><risdate>2017</risdate><volume>5</volume><issue>4</issue><spage>521</spage><epage>525</epage><pages>521-525</pages><issn>2050-0904</issn><eissn>2050-0904</eissn><abstract>Key Clinical Message
An implantable cardioverter defibrillator (ICD) lead dislodgement into the right atrium is a dangerous situation, particularly in patients in atrial fibrillation because atrial fibrillation can be sensed as ventricular fibrillation and true ventricular fibrillation induced with an inappropriate shock. In the presence of shocks, ICD interrogation should be performed as soon as possible.
An implantable cardioverter defibrillator (ICD) lead dislodgement into the right atrium is a dangerous situation, particularly in patients in atrial fibrillation because atrial fibrillation can be sensed as ventricular fibrillation and true ventricular fibrillation induced with an inappropriate shock. In the presence of shocks, ICD interrogation should be performed as soon as possible.</abstract><cop>England</cop><pub>John Wiley & Sons, Inc</pub><pmid>28396781</pmid><doi>10.1002/ccr3.893</doi><tpages>5</tpages><orcidid>https://orcid.org/0000-0003-2234-5248</orcidid><oa>free_for_read</oa></addata></record> |
fulltext | fulltext |
identifier | ISSN: 2050-0904 |
ispartof | Clinical case reports, 2017-04, Vol.5 (4), p.521-525 |
issn | 2050-0904 2050-0904 |
language | eng |
recordid | cdi_pubmedcentral_primary_oai_pubmedcentral_nih_gov_5378847 |
source | DOAJ Directory of Open Access Journals; Elektronische Zeitschriftenbibliothek - Frei zugängliche E-Journals; Access via Wiley Online Library; Wiley Online Library (Open Access Collection); PubMed Central |
subjects | Atrial fibrillation Cardiac arrhythmia Cardiovascular disease Case Report Case Reports Consciousness Defibrillators Disease prevention Electrocardiography implantable cardioverter defibrillator inappropriate shock lead dislodgement Patients Prostheses Quality of life Sinuses ventricular fibrillation |
title | Life‐threatening and life‐saving inappropriate implantable cardioverter defibrillator shocks |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2024-12-25T18%3A03%3A36IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_pubme&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Life%E2%80%90threatening%20and%20life%E2%80%90saving%20inappropriate%20implantable%20cardioverter%20defibrillator%20shocks&rft.jtitle=Clinical%20case%20reports&rft.au=P%C3%A9rez%E2%80%90Rodon,%20Jordi&rft.date=2017-04&rft.volume=5&rft.issue=4&rft.spage=521&rft.epage=525&rft.pages=521-525&rft.issn=2050-0904&rft.eissn=2050-0904&rft_id=info:doi/10.1002/ccr3.893&rft_dat=%3Cproquest_pubme%3E2290261033%3C/proquest_pubme%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=2290261033&rft_id=info:pmid/28396781&rfr_iscdi=true |