Life‐threatening and life‐saving inappropriate implantable cardioverter defibrillator shocks

Key Clinical Message An implantable cardioverter defibrillator (ICD) lead dislodgement into the right atrium is a dangerous situation, particularly in patients in atrial fibrillation because atrial fibrillation can be sensed as ventricular fibrillation and true ventricular fibrillation induced with...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Clinical case reports 2017-04, Vol.5 (4), p.521-525
Hauptverfasser: Pérez‐Rodon, Jordi, Doiny, David, Miranda, Berta, Rivas‐Gandara, Nuria, Roca‐Luque, Ivo, Francisco‐Pascual, Jaume, Lidón, Rosa Maria, García‐Dorado, David, Moya Mitjans, Angel
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 525
container_issue 4
container_start_page 521
container_title Clinical case reports
container_volume 5
creator Pérez‐Rodon, Jordi
Doiny, David
Miranda, Berta
Rivas‐Gandara, Nuria
Roca‐Luque, Ivo
Francisco‐Pascual, Jaume
Lidón, Rosa Maria
García‐Dorado, David
Moya Mitjans, Angel
description Key Clinical Message An implantable cardioverter defibrillator (ICD) lead dislodgement into the right atrium is a dangerous situation, particularly in patients in atrial fibrillation because atrial fibrillation can be sensed as ventricular fibrillation and true ventricular fibrillation induced with an inappropriate shock. In the presence of shocks, ICD interrogation should be performed as soon as possible. An implantable cardioverter defibrillator (ICD) lead dislodgement into the right atrium is a dangerous situation, particularly in patients in atrial fibrillation because atrial fibrillation can be sensed as ventricular fibrillation and true ventricular fibrillation induced with an inappropriate shock. In the presence of shocks, ICD interrogation should be performed as soon as possible.
doi_str_mv 10.1002/ccr3.893
format Article
fullrecord <record><control><sourceid>proquest_pubme</sourceid><recordid>TN_cdi_pubmedcentral_primary_oai_pubmedcentral_nih_gov_5378847</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>2290261033</sourcerecordid><originalsourceid>FETCH-LOGICAL-c3663-43ad85b54dedc6412569364fcf8d2b13d4a431d83ae677afbb5b46b779799853</originalsourceid><addsrcrecordid>eNp1kc1KAzEYRYMotmjBJ5ABN25G8zdJZiNI8Q8KgriPySRjo9PJmEwr7nwEn9EnMaW11IWrhC-Hk8t3AThC8AxBiM-rKpAzUZIdMMSwgDksId3dug_AKMYXCCGCHBcI7oMBFqRkXKAheJq42n5_fvXTYFVvW9c-Z6o1WbMaR7VYTlyrui74LrjEZG7WNartlW5sVqlgnF_Y0NuQGVs7HVzTqN6HLE599RoPwV6tmmhH6_MAPF5fPY5v88n9zd34cpJXhDGSU6KMKHRBjTUVowgXrCSM1lUtDNaIGKooQUYQZRnnqta60JRpzktelqIgB-Bipe3mepYUtu2DamRKPFPhQ3rl5N-X1k3ls1_IgnAhKE-Ck7Ug-Le5jb188fPQpsgS4xJihiAhiTpdUVXwMQZbb35AUC7bkMs2ZGojocfbiTbg7-4TkK-Ad9fYj39Fcjx-IEvhD43Nl90</addsrcrecordid><sourcetype>Open Access Repository</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2290261033</pqid></control><display><type>article</type><title>Life‐threatening and life‐saving inappropriate implantable cardioverter defibrillator shocks</title><source>DOAJ Directory of Open Access Journals</source><source>Elektronische Zeitschriftenbibliothek - Frei zugängliche E-Journals</source><source>Access via Wiley Online Library</source><source>Wiley Online Library (Open Access Collection)</source><source>PubMed Central</source><creator>Pérez‐Rodon, Jordi ; Doiny, David ; Miranda, Berta ; Rivas‐Gandara, Nuria ; Roca‐Luque, Ivo ; Francisco‐Pascual, Jaume ; Lidón, Rosa Maria ; García‐Dorado, David ; Moya Mitjans, Angel</creator><creatorcontrib>Pérez‐Rodon, Jordi ; Doiny, David ; Miranda, Berta ; Rivas‐Gandara, Nuria ; Roca‐Luque, Ivo ; Francisco‐Pascual, Jaume ; Lidón, Rosa Maria ; García‐Dorado, David ; Moya Mitjans, Angel</creatorcontrib><description>Key Clinical Message An implantable cardioverter defibrillator (ICD) lead dislodgement into the right atrium is a dangerous situation, particularly in patients in atrial fibrillation because atrial fibrillation can be sensed as ventricular fibrillation and true ventricular fibrillation induced with an inappropriate shock. In the presence of shocks, ICD interrogation should be performed as soon as possible. An implantable cardioverter defibrillator (ICD) lead dislodgement into the right atrium is a dangerous situation, particularly in patients in atrial fibrillation because atrial fibrillation can be sensed as ventricular fibrillation and true ventricular fibrillation induced with an inappropriate shock. In the presence of shocks, ICD interrogation should be performed as soon as possible.</description><identifier>ISSN: 2050-0904</identifier><identifier>EISSN: 2050-0904</identifier><identifier>DOI: 10.1002/ccr3.893</identifier><identifier>PMID: 28396781</identifier><language>eng</language><publisher>England: John Wiley &amp; Sons, Inc</publisher><subject>Atrial fibrillation ; Cardiac arrhythmia ; Cardiovascular disease ; Case Report ; Case Reports ; Consciousness ; Defibrillators ; Disease prevention ; Electrocardiography ; implantable cardioverter defibrillator ; inappropriate shock ; lead dislodgement ; Patients ; Prostheses ; Quality of life ; Sinuses ; ventricular fibrillation</subject><ispartof>Clinical case reports, 2017-04, Vol.5 (4), p.521-525</ispartof><rights>2017 The Authors. published by John Wiley &amp; Sons Ltd.</rights><rights>2017. This work is published under http://creativecommons.org/licenses/by/4.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><cites>FETCH-LOGICAL-c3663-43ad85b54dedc6412569364fcf8d2b13d4a431d83ae677afbb5b46b779799853</cites><orcidid>0000-0003-2234-5248</orcidid></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktopdf>$$Uhttps://www.ncbi.nlm.nih.gov/pmc/articles/PMC5378847/pdf/$$EPDF$$P50$$Gpubmedcentral$$Hfree_for_read</linktopdf><linktohtml>$$Uhttps://www.ncbi.nlm.nih.gov/pmc/articles/PMC5378847/$$EHTML$$P50$$Gpubmedcentral$$Hfree_for_read</linktohtml><link.rule.ids>230,314,727,780,784,864,885,1417,11562,27924,27925,45574,45575,46052,46476,53791,53793</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/28396781$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Pérez‐Rodon, Jordi</creatorcontrib><creatorcontrib>Doiny, David</creatorcontrib><creatorcontrib>Miranda, Berta</creatorcontrib><creatorcontrib>Rivas‐Gandara, Nuria</creatorcontrib><creatorcontrib>Roca‐Luque, Ivo</creatorcontrib><creatorcontrib>Francisco‐Pascual, Jaume</creatorcontrib><creatorcontrib>Lidón, Rosa Maria</creatorcontrib><creatorcontrib>García‐Dorado, David</creatorcontrib><creatorcontrib>Moya Mitjans, Angel</creatorcontrib><title>Life‐threatening and life‐saving inappropriate implantable cardioverter defibrillator shocks</title><title>Clinical case reports</title><addtitle>Clin Case Rep</addtitle><description>Key Clinical Message An implantable cardioverter defibrillator (ICD) lead dislodgement into the right atrium is a dangerous situation, particularly in patients in atrial fibrillation because atrial fibrillation can be sensed as ventricular fibrillation and true ventricular fibrillation induced with an inappropriate shock. In the presence of shocks, ICD interrogation should be performed as soon as possible. An implantable cardioverter defibrillator (ICD) lead dislodgement into the right atrium is a dangerous situation, particularly in patients in atrial fibrillation because atrial fibrillation can be sensed as ventricular fibrillation and true ventricular fibrillation induced with an inappropriate shock. In the presence of shocks, ICD interrogation should be performed as soon as possible.</description><subject>Atrial fibrillation</subject><subject>Cardiac arrhythmia</subject><subject>Cardiovascular disease</subject><subject>Case Report</subject><subject>Case Reports</subject><subject>Consciousness</subject><subject>Defibrillators</subject><subject>Disease prevention</subject><subject>Electrocardiography</subject><subject>implantable cardioverter defibrillator</subject><subject>inappropriate shock</subject><subject>lead dislodgement</subject><subject>Patients</subject><subject>Prostheses</subject><subject>Quality of life</subject><subject>Sinuses</subject><subject>ventricular fibrillation</subject><issn>2050-0904</issn><issn>2050-0904</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2017</creationdate><recordtype>article</recordtype><sourceid>24P</sourceid><sourceid>WIN</sourceid><sourceid>ABUWG</sourceid><sourceid>AFKRA</sourceid><sourceid>AZQEC</sourceid><sourceid>BENPR</sourceid><sourceid>CCPQU</sourceid><sourceid>DWQXO</sourceid><recordid>eNp1kc1KAzEYRYMotmjBJ5ABN25G8zdJZiNI8Q8KgriPySRjo9PJmEwr7nwEn9EnMaW11IWrhC-Hk8t3AThC8AxBiM-rKpAzUZIdMMSwgDksId3dug_AKMYXCCGCHBcI7oMBFqRkXKAheJq42n5_fvXTYFVvW9c-Z6o1WbMaR7VYTlyrui74LrjEZG7WNartlW5sVqlgnF_Y0NuQGVs7HVzTqN6HLE599RoPwV6tmmhH6_MAPF5fPY5v88n9zd34cpJXhDGSU6KMKHRBjTUVowgXrCSM1lUtDNaIGKooQUYQZRnnqta60JRpzktelqIgB-Bipe3mepYUtu2DamRKPFPhQ3rl5N-X1k3ls1_IgnAhKE-Ck7Ug-Le5jb188fPQpsgS4xJihiAhiTpdUVXwMQZbb35AUC7bkMs2ZGojocfbiTbg7-4TkK-Ad9fYj39Fcjx-IEvhD43Nl90</recordid><startdate>201704</startdate><enddate>201704</enddate><creator>Pérez‐Rodon, Jordi</creator><creator>Doiny, David</creator><creator>Miranda, Berta</creator><creator>Rivas‐Gandara, Nuria</creator><creator>Roca‐Luque, Ivo</creator><creator>Francisco‐Pascual, Jaume</creator><creator>Lidón, Rosa Maria</creator><creator>García‐Dorado, David</creator><creator>Moya Mitjans, Angel</creator><general>John Wiley &amp; Sons, Inc</general><general>John Wiley and Sons Inc</general><scope>24P</scope><scope>WIN</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>3V.</scope><scope>7X7</scope><scope>7XB</scope><scope>8FI</scope><scope>8FJ</scope><scope>8FK</scope><scope>ABUWG</scope><scope>AFKRA</scope><scope>AZQEC</scope><scope>BENPR</scope><scope>CCPQU</scope><scope>DWQXO</scope><scope>FYUFA</scope><scope>GHDGH</scope><scope>K9.</scope><scope>M0S</scope><scope>PIMPY</scope><scope>PQEST</scope><scope>PQQKQ</scope><scope>PQUKI</scope><scope>PRINS</scope><scope>5PM</scope><orcidid>https://orcid.org/0000-0003-2234-5248</orcidid></search><sort><creationdate>201704</creationdate><title>Life‐threatening and life‐saving inappropriate implantable cardioverter defibrillator shocks</title><author>Pérez‐Rodon, Jordi ; Doiny, David ; Miranda, Berta ; Rivas‐Gandara, Nuria ; Roca‐Luque, Ivo ; Francisco‐Pascual, Jaume ; Lidón, Rosa Maria ; García‐Dorado, David ; Moya Mitjans, Angel</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c3663-43ad85b54dedc6412569364fcf8d2b13d4a431d83ae677afbb5b46b779799853</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2017</creationdate><topic>Atrial fibrillation</topic><topic>Cardiac arrhythmia</topic><topic>Cardiovascular disease</topic><topic>Case Report</topic><topic>Case Reports</topic><topic>Consciousness</topic><topic>Defibrillators</topic><topic>Disease prevention</topic><topic>Electrocardiography</topic><topic>implantable cardioverter defibrillator</topic><topic>inappropriate shock</topic><topic>lead dislodgement</topic><topic>Patients</topic><topic>Prostheses</topic><topic>Quality of life</topic><topic>Sinuses</topic><topic>ventricular fibrillation</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Pérez‐Rodon, Jordi</creatorcontrib><creatorcontrib>Doiny, David</creatorcontrib><creatorcontrib>Miranda, Berta</creatorcontrib><creatorcontrib>Rivas‐Gandara, Nuria</creatorcontrib><creatorcontrib>Roca‐Luque, Ivo</creatorcontrib><creatorcontrib>Francisco‐Pascual, Jaume</creatorcontrib><creatorcontrib>Lidón, Rosa Maria</creatorcontrib><creatorcontrib>García‐Dorado, David</creatorcontrib><creatorcontrib>Moya Mitjans, Angel</creatorcontrib><collection>Wiley Online Library (Open Access Collection)</collection><collection>Wiley Online Library (Open Access Collection)</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>ProQuest Central (Corporate)</collection><collection>Health &amp; Medical Collection</collection><collection>ProQuest Central (purchase pre-March 2016)</collection><collection>Hospital Premium Collection</collection><collection>Hospital Premium Collection (Alumni Edition)</collection><collection>ProQuest Central (Alumni) (purchase pre-March 2016)</collection><collection>ProQuest Central (Alumni Edition)</collection><collection>ProQuest Central UK/Ireland</collection><collection>ProQuest Central Essentials</collection><collection>ProQuest Central</collection><collection>ProQuest One Community College</collection><collection>ProQuest Central Korea</collection><collection>Health Research Premium Collection</collection><collection>Health Research Premium Collection (Alumni)</collection><collection>ProQuest Health &amp; Medical Complete (Alumni)</collection><collection>Health &amp; Medical Collection (Alumni Edition)</collection><collection>Publicly Available Content Database</collection><collection>ProQuest One Academic Eastern Edition (DO NOT USE)</collection><collection>ProQuest One Academic</collection><collection>ProQuest One Academic UKI Edition</collection><collection>ProQuest Central China</collection><collection>PubMed Central (Full Participant titles)</collection><jtitle>Clinical case reports</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Pérez‐Rodon, Jordi</au><au>Doiny, David</au><au>Miranda, Berta</au><au>Rivas‐Gandara, Nuria</au><au>Roca‐Luque, Ivo</au><au>Francisco‐Pascual, Jaume</au><au>Lidón, Rosa Maria</au><au>García‐Dorado, David</au><au>Moya Mitjans, Angel</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Life‐threatening and life‐saving inappropriate implantable cardioverter defibrillator shocks</atitle><jtitle>Clinical case reports</jtitle><addtitle>Clin Case Rep</addtitle><date>2017-04</date><risdate>2017</risdate><volume>5</volume><issue>4</issue><spage>521</spage><epage>525</epage><pages>521-525</pages><issn>2050-0904</issn><eissn>2050-0904</eissn><abstract>Key Clinical Message An implantable cardioverter defibrillator (ICD) lead dislodgement into the right atrium is a dangerous situation, particularly in patients in atrial fibrillation because atrial fibrillation can be sensed as ventricular fibrillation and true ventricular fibrillation induced with an inappropriate shock. In the presence of shocks, ICD interrogation should be performed as soon as possible. An implantable cardioverter defibrillator (ICD) lead dislodgement into the right atrium is a dangerous situation, particularly in patients in atrial fibrillation because atrial fibrillation can be sensed as ventricular fibrillation and true ventricular fibrillation induced with an inappropriate shock. In the presence of shocks, ICD interrogation should be performed as soon as possible.</abstract><cop>England</cop><pub>John Wiley &amp; Sons, Inc</pub><pmid>28396781</pmid><doi>10.1002/ccr3.893</doi><tpages>5</tpages><orcidid>https://orcid.org/0000-0003-2234-5248</orcidid><oa>free_for_read</oa></addata></record>
fulltext fulltext
identifier ISSN: 2050-0904
ispartof Clinical case reports, 2017-04, Vol.5 (4), p.521-525
issn 2050-0904
2050-0904
language eng
recordid cdi_pubmedcentral_primary_oai_pubmedcentral_nih_gov_5378847
source DOAJ Directory of Open Access Journals; Elektronische Zeitschriftenbibliothek - Frei zugängliche E-Journals; Access via Wiley Online Library; Wiley Online Library (Open Access Collection); PubMed Central
subjects Atrial fibrillation
Cardiac arrhythmia
Cardiovascular disease
Case Report
Case Reports
Consciousness
Defibrillators
Disease prevention
Electrocardiography
implantable cardioverter defibrillator
inappropriate shock
lead dislodgement
Patients
Prostheses
Quality of life
Sinuses
ventricular fibrillation
title Life‐threatening and life‐saving inappropriate implantable cardioverter defibrillator shocks
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2024-12-25T18%3A03%3A36IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_pubme&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Life%E2%80%90threatening%20and%20life%E2%80%90saving%20inappropriate%20implantable%20cardioverter%20defibrillator%20shocks&rft.jtitle=Clinical%20case%20reports&rft.au=P%C3%A9rez%E2%80%90Rodon,%20Jordi&rft.date=2017-04&rft.volume=5&rft.issue=4&rft.spage=521&rft.epage=525&rft.pages=521-525&rft.issn=2050-0904&rft.eissn=2050-0904&rft_id=info:doi/10.1002/ccr3.893&rft_dat=%3Cproquest_pubme%3E2290261033%3C/proquest_pubme%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=2290261033&rft_id=info:pmid/28396781&rfr_iscdi=true