Multiplexed and fully automated detection of metabolic biomarkers using microdialysis probe
We report here, the design and development of an automated near real-time continuous detection system for lactate, glutamate, pyruvate and glucose using microdialysis probe. The system developed can automatically push perfusate through microdialysis probe (20, 100 and 1000kDa MWCO cutoff probe) at l...
Gespeichert in:
Veröffentlicht in: | Sensors and actuators. B, Chemical Chemical, 2017, Vol.238, p.633-640 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We report here, the design and development of an automated near real-time continuous detection system for lactate, glutamate, pyruvate and glucose using microdialysis probe. The system developed can automatically push perfusate through microdialysis probe (20, 100 and 1000kDa MWCO cutoff probe) at low to medium flow rate of 0.5–2μL/min with almost 100% fluid recovery. The microdialysate collected from the probe is analyzed automatically for these four metabolite biomarkers. It operates in a continuous mode with measurements of all four biomarkers once every 20min. The dynamic range for these different markers covers the entire clinical range of traumatic brain injury. The prototype shows a low variation of ∼ 7–10% across the entire clinical range for all the biomarkers with fairly good accuracy of ∼95%. The instrument canrun continuously for 24h without user intervention. With a long tubing of 1m to and from the microdialysis probe and associated dead volume, the total lag time for actual event at the probe site versusreported concentration is roughly 1h. |
---|---|
ISSN: | 0925-4005 1873-3077 |
DOI: | 10.1016/j.snb.2016.07.097 |