Retrospective respiratory self‐gating and removal of bulk motion in pulmonary UTE MRI of neonates and adults

Purpose To implement pulmonary three‐dimensional (3D) radial ultrashort echo‐time (UTE) MRI in non‐sedated, free‐breathing neonates and adults with retrospective motion tracking of respiratory and intermittent bulk motion, to obtain diagnostic‐quality, respiratory‐gated images. Methods Pulmonary 3D...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Magnetic resonance in medicine 2017-03, Vol.77 (3), p.1284-1295
Hauptverfasser: Higano, Nara S., Hahn, Andrew D., Tkach, Jean A., Cao, Xuefeng, Walkup, Laura L., Thomen, Robert P., Merhar, Stephanie L., Kingma, Paul S., Fain, Sean B., Woods, Jason C.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Purpose To implement pulmonary three‐dimensional (3D) radial ultrashort echo‐time (UTE) MRI in non‐sedated, free‐breathing neonates and adults with retrospective motion tracking of respiratory and intermittent bulk motion, to obtain diagnostic‐quality, respiratory‐gated images. Methods Pulmonary 3D radial UTE MRI was performed at 1.5 tesla (T) during free breathing in neonates and adult volunteers for validation. Motion‐tracking waveforms were obtained from the time course of each free induction decay's initial point (i.e., k‐space center), allowing for respiratory‐gated image reconstructions that excluded data acquired during bulk motion. Tidal volumes were calculated from end‐expiration and end‐inspiration images. Respiratory rates were calculated from the Fourier transform of the motion‐tracking waveform during quiet breathing, with comparison to physiologic prediction in neonates and validation with spirometry in adults. Results High‐quality respiratory‐gated anatomic images were obtained at inspiration and expiration, with less respiratory blurring at the expense of signal‐to‐noise for narrower gating windows. Inspiration‐expiration volume differences agreed with physiologic predictions (neonates; Bland‐Altman bias = 6.2 mL) and spirometric values (adults; bias = 0.11 L). MRI‐measured respiratory rates compared well with the observed rates (biases = −0.5 and 0.2 breaths/min for neonates and adults, respectively). Conclusions Three‐dimensional radial pulmonary UTE MRI allows for retrospective respiratory self‐gating and removal of intermittent bulk motion in free‐breathing, non‐sedated neonates and adults. Magn Reson Med 77:1284–1295, 2017. © 2016 International Society for Magnetic Resonance in Medicine
ISSN:0740-3194
1522-2594
DOI:10.1002/mrm.26212