Reducing the airflow waveform distortions from breathing style and body position with improved calibration of respiratory effort belts
Respiratory effort belt measurement is a widely used method to monitor respiration. Signal waveforms of respiratory volume and flow may indicate pathological signs of several diseases and, thus, it would be highly desirable to predict them accurately. Calibrated effort belts are sufficiently accurat...
Gespeichert in:
Veröffentlicht in: | Biomedical engineering online 2013-09, Vol.12 (1), p.97-97 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | 97 |
---|---|
container_issue | 1 |
container_start_page | 97 |
container_title | Biomedical engineering online |
container_volume | 12 |
creator | Seppänen, Tiina M Alho, Olli-Pekka Seppänen, Tapio |
description | Respiratory effort belt measurement is a widely used method to monitor respiration. Signal waveforms of respiratory volume and flow may indicate pathological signs of several diseases and, thus, it would be highly desirable to predict them accurately. Calibrated effort belts are sufficiently accurate for estimating respiratory rate, but the respiratory volume and flow prediction accuracies degrade considerably with changes in the subject's body position and breathing style.
An improved calibration method of respiratory effort belts is presented in this paper. It is based on an optimally trained FIR (Finite Impulse Response) filter bank constructed as a MISO system (Multiple-Input Single-Output) between respiratory effort belt signals and the spirometer in order to reduce waveform errors. Ten healthy adult volunteers were recruited. Breathing was varied between the following styles: metronome-guided controlled breathing rate of 0.1 Hz, 0.15 Hz, 0.25 Hz and 0.33 Hz, and a free rate that was felt normal by each subject. Body position was varied between supine, sitting and standing. The proposed calibration method was tested against these variations and compared with the state-of-the-art methods from the literature.
Relative waveform error decreased 60-70% when predicting airflow under changing breathing styles. The coefficient of determination R2 varied between 0.88-0.95 and 0.65-0.79 with the proposed and the standard method, respectively. Standard deviation of respiratory volume error decreased even 80%. The proposed method outperformed other methods.
Results show that not only the respiratory volume can be computed more precisely from the predicted airflow, but also the flow waveforms are very accurate with the proposed method. The method is robust to breathing style changes and body position changes improving greatly the accuracy of the calibration of respiratory effort belts over the standard method. The enhanced accuracy of the belt calibration offers interesting opportunities, e.g. in pulmonary and critical care medicine when objective measurements are required. |
doi_str_mv | 10.1186/1475-925X-12-97 |
format | Article |
fullrecord | <record><control><sourceid>gale_pubme</sourceid><recordid>TN_cdi_pubmedcentral_primary_oai_pubmedcentral_nih_gov_3852323</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><galeid>A534654663</galeid><sourcerecordid>A534654663</sourcerecordid><originalsourceid>FETCH-LOGICAL-b614t-b9faa1bc75f6458a9f41a3fa33b2d445d570302315f4b2f3b571eb4be68f5f133</originalsourceid><addsrcrecordid>eNqNkktv1DAQxyNERUvhzA1Z4gKHtH7mcUFqKx6VKiEVkLhZdjLedZXEi-3ssl-Az43Dlm2DioR8sD3zm7_HM5NlLwg-IaQqTgkvRV5T8S0nNK_LR9nR3vL43vkwexrCDcYU46J-kh1SjktWEnyU_byGdmzssEBxCUhZbzq3QRu1BuN8j1obovPRuiEg412PtAcVlxMf4rZLEUOLtGu3aOWCnTi0sXGJbL_ybg0talRntVe_Pc4gD2Fl09X5LQKTnohIQxfDs-zAqC7A89v9OPv6_t2Xi4_51acPlxdnV7kuCI-5ro1SRDelMAUXlaoNJ4oZxZimLeeiFSVmmDIiDNfUMC1KApprKCojDGHsOHu7012Nuoe2gSF61cmVt73yW-mUlXPPYJdy4daSVYIyOgmc7wS0df8QmHsa18upD3LqgyRU1mUSeX2bhXffRwhR9jY00HVqADeGxIuaCFZz-h8or2ji-aT66i_0xo1-SOWcKFpiWoj6jlqoDqQdjEtpNpOoPBOMF4IXxfTNkweotFrobeMGMDbZZwFvZgGJifAjLtQYgrz8fD1nT3ds410IHsy-fATLaawfKNjL-23b83_mmP0CxBn0-w</addsrcrecordid><sourcetype>Open Access Repository</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>1442702659</pqid></control><display><type>article</type><title>Reducing the airflow waveform distortions from breathing style and body position with improved calibration of respiratory effort belts</title><source>MEDLINE</source><source>DOAJ Directory of Open Access Journals</source><source>PubMed Central Open Access</source><source>Springer Nature OA Free Journals</source><source>EZB-FREE-00999 freely available EZB journals</source><source>PubMed Central</source><source>SpringerLink Journals - AutoHoldings</source><creator>Seppänen, Tiina M ; Alho, Olli-Pekka ; Seppänen, Tapio</creator><creatorcontrib>Seppänen, Tiina M ; Alho, Olli-Pekka ; Seppänen, Tapio</creatorcontrib><description>Respiratory effort belt measurement is a widely used method to monitor respiration. Signal waveforms of respiratory volume and flow may indicate pathological signs of several diseases and, thus, it would be highly desirable to predict them accurately. Calibrated effort belts are sufficiently accurate for estimating respiratory rate, but the respiratory volume and flow prediction accuracies degrade considerably with changes in the subject's body position and breathing style.
An improved calibration method of respiratory effort belts is presented in this paper. It is based on an optimally trained FIR (Finite Impulse Response) filter bank constructed as a MISO system (Multiple-Input Single-Output) between respiratory effort belt signals and the spirometer in order to reduce waveform errors. Ten healthy adult volunteers were recruited. Breathing was varied between the following styles: metronome-guided controlled breathing rate of 0.1 Hz, 0.15 Hz, 0.25 Hz and 0.33 Hz, and a free rate that was felt normal by each subject. Body position was varied between supine, sitting and standing. The proposed calibration method was tested against these variations and compared with the state-of-the-art methods from the literature.
Relative waveform error decreased 60-70% when predicting airflow under changing breathing styles. The coefficient of determination R2 varied between 0.88-0.95 and 0.65-0.79 with the proposed and the standard method, respectively. Standard deviation of respiratory volume error decreased even 80%. The proposed method outperformed other methods.
Results show that not only the respiratory volume can be computed more precisely from the predicted airflow, but also the flow waveforms are very accurate with the proposed method. The method is robust to breathing style changes and body position changes improving greatly the accuracy of the calibration of respiratory effort belts over the standard method. The enhanced accuracy of the belt calibration offers interesting opportunities, e.g. in pulmonary and critical care medicine when objective measurements are required.</description><identifier>ISSN: 1475-925X</identifier><identifier>EISSN: 1475-925X</identifier><identifier>DOI: 10.1186/1475-925X-12-97</identifier><identifier>PMID: 24073710</identifier><language>eng</language><publisher>England: BioMed Central Ltd</publisher><subject>Abdomen ; Adult ; Air ; Banks (Finance) ; Calibration ; Female ; Humans ; Linear Models ; Male ; Methods ; Models, Biological ; Otolaryngology ; Physiological aspects ; Posture ; Pulmonary function tests ; Regression analysis ; Respiration ; Respiratory Function Tests - methods ; Studies</subject><ispartof>Biomedical engineering online, 2013-09, Vol.12 (1), p.97-97</ispartof><rights>COPYRIGHT 2013 BioMed Central Ltd.</rights><rights>2013 Seppänen et al.; licensee BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.</rights><rights>Copyright © 2013 Seppänen et al.; licensee BioMed Central Ltd. 2013 Seppänen et al.; licensee BioMed Central Ltd.</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-b614t-b9faa1bc75f6458a9f41a3fa33b2d445d570302315f4b2f3b571eb4be68f5f133</citedby><cites>FETCH-LOGICAL-b614t-b9faa1bc75f6458a9f41a3fa33b2d445d570302315f4b2f3b571eb4be68f5f133</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktopdf>$$Uhttps://www.ncbi.nlm.nih.gov/pmc/articles/PMC3852323/pdf/$$EPDF$$P50$$Gpubmedcentral$$Hfree_for_read</linktopdf><linktohtml>$$Uhttps://www.ncbi.nlm.nih.gov/pmc/articles/PMC3852323/$$EHTML$$P50$$Gpubmedcentral$$Hfree_for_read</linktohtml><link.rule.ids>230,314,727,780,784,864,885,27923,27924,53790,53792</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/24073710$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Seppänen, Tiina M</creatorcontrib><creatorcontrib>Alho, Olli-Pekka</creatorcontrib><creatorcontrib>Seppänen, Tapio</creatorcontrib><title>Reducing the airflow waveform distortions from breathing style and body position with improved calibration of respiratory effort belts</title><title>Biomedical engineering online</title><addtitle>Biomed Eng Online</addtitle><description>Respiratory effort belt measurement is a widely used method to monitor respiration. Signal waveforms of respiratory volume and flow may indicate pathological signs of several diseases and, thus, it would be highly desirable to predict them accurately. Calibrated effort belts are sufficiently accurate for estimating respiratory rate, but the respiratory volume and flow prediction accuracies degrade considerably with changes in the subject's body position and breathing style.
An improved calibration method of respiratory effort belts is presented in this paper. It is based on an optimally trained FIR (Finite Impulse Response) filter bank constructed as a MISO system (Multiple-Input Single-Output) between respiratory effort belt signals and the spirometer in order to reduce waveform errors. Ten healthy adult volunteers were recruited. Breathing was varied between the following styles: metronome-guided controlled breathing rate of 0.1 Hz, 0.15 Hz, 0.25 Hz and 0.33 Hz, and a free rate that was felt normal by each subject. Body position was varied between supine, sitting and standing. The proposed calibration method was tested against these variations and compared with the state-of-the-art methods from the literature.
Relative waveform error decreased 60-70% when predicting airflow under changing breathing styles. The coefficient of determination R2 varied between 0.88-0.95 and 0.65-0.79 with the proposed and the standard method, respectively. Standard deviation of respiratory volume error decreased even 80%. The proposed method outperformed other methods.
Results show that not only the respiratory volume can be computed more precisely from the predicted airflow, but also the flow waveforms are very accurate with the proposed method. The method is robust to breathing style changes and body position changes improving greatly the accuracy of the calibration of respiratory effort belts over the standard method. The enhanced accuracy of the belt calibration offers interesting opportunities, e.g. in pulmonary and critical care medicine when objective measurements are required.</description><subject>Abdomen</subject><subject>Adult</subject><subject>Air</subject><subject>Banks (Finance)</subject><subject>Calibration</subject><subject>Female</subject><subject>Humans</subject><subject>Linear Models</subject><subject>Male</subject><subject>Methods</subject><subject>Models, Biological</subject><subject>Otolaryngology</subject><subject>Physiological aspects</subject><subject>Posture</subject><subject>Pulmonary function tests</subject><subject>Regression analysis</subject><subject>Respiration</subject><subject>Respiratory Function Tests - methods</subject><subject>Studies</subject><issn>1475-925X</issn><issn>1475-925X</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2013</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><sourceid>ABUWG</sourceid><sourceid>AFKRA</sourceid><sourceid>AZQEC</sourceid><sourceid>BENPR</sourceid><sourceid>CCPQU</sourceid><sourceid>DWQXO</sourceid><sourceid>GNUQQ</sourceid><recordid>eNqNkktv1DAQxyNERUvhzA1Z4gKHtH7mcUFqKx6VKiEVkLhZdjLedZXEi-3ssl-Az43Dlm2DioR8sD3zm7_HM5NlLwg-IaQqTgkvRV5T8S0nNK_LR9nR3vL43vkwexrCDcYU46J-kh1SjktWEnyU_byGdmzssEBxCUhZbzq3QRu1BuN8j1obovPRuiEg412PtAcVlxMf4rZLEUOLtGu3aOWCnTi0sXGJbL_ybg0talRntVe_Pc4gD2Fl09X5LQKTnohIQxfDs-zAqC7A89v9OPv6_t2Xi4_51acPlxdnV7kuCI-5ro1SRDelMAUXlaoNJ4oZxZimLeeiFSVmmDIiDNfUMC1KApprKCojDGHsOHu7012Nuoe2gSF61cmVt73yW-mUlXPPYJdy4daSVYIyOgmc7wS0df8QmHsa18upD3LqgyRU1mUSeX2bhXffRwhR9jY00HVqADeGxIuaCFZz-h8or2ji-aT66i_0xo1-SOWcKFpiWoj6jlqoDqQdjEtpNpOoPBOMF4IXxfTNkweotFrobeMGMDbZZwFvZgGJifAjLtQYgrz8fD1nT3ds410IHsy-fATLaawfKNjL-23b83_mmP0CxBn0-w</recordid><startdate>20130928</startdate><enddate>20130928</enddate><creator>Seppänen, Tiina M</creator><creator>Alho, Olli-Pekka</creator><creator>Seppänen, Tapio</creator><general>BioMed Central Ltd</general><general>BioMed Central</general><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>ISR</scope><scope>3V.</scope><scope>7QO</scope><scope>7X7</scope><scope>7XB</scope><scope>88E</scope><scope>8FD</scope><scope>8FE</scope><scope>8FG</scope><scope>8FH</scope><scope>8FI</scope><scope>8FJ</scope><scope>8FK</scope><scope>ABJCF</scope><scope>ABUWG</scope><scope>AFKRA</scope><scope>AZQEC</scope><scope>BBNVY</scope><scope>BENPR</scope><scope>BGLVJ</scope><scope>BHPHI</scope><scope>CCPQU</scope><scope>DWQXO</scope><scope>FR3</scope><scope>FYUFA</scope><scope>GHDGH</scope><scope>GNUQQ</scope><scope>HCIFZ</scope><scope>K9.</scope><scope>L6V</scope><scope>LK8</scope><scope>M0S</scope><scope>M1P</scope><scope>M7P</scope><scope>M7S</scope><scope>P64</scope><scope>PIMPY</scope><scope>PQEST</scope><scope>PQQKQ</scope><scope>PQUKI</scope><scope>PRINS</scope><scope>PTHSS</scope><scope>7X8</scope><scope>5PM</scope></search><sort><creationdate>20130928</creationdate><title>Reducing the airflow waveform distortions from breathing style and body position with improved calibration of respiratory effort belts</title><author>Seppänen, Tiina M ; Alho, Olli-Pekka ; Seppänen, Tapio</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-b614t-b9faa1bc75f6458a9f41a3fa33b2d445d570302315f4b2f3b571eb4be68f5f133</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2013</creationdate><topic>Abdomen</topic><topic>Adult</topic><topic>Air</topic><topic>Banks (Finance)</topic><topic>Calibration</topic><topic>Female</topic><topic>Humans</topic><topic>Linear Models</topic><topic>Male</topic><topic>Methods</topic><topic>Models, Biological</topic><topic>Otolaryngology</topic><topic>Physiological aspects</topic><topic>Posture</topic><topic>Pulmonary function tests</topic><topic>Regression analysis</topic><topic>Respiration</topic><topic>Respiratory Function Tests - methods</topic><topic>Studies</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Seppänen, Tiina M</creatorcontrib><creatorcontrib>Alho, Olli-Pekka</creatorcontrib><creatorcontrib>Seppänen, Tapio</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>Gale In Context: Science</collection><collection>ProQuest Central (Corporate)</collection><collection>Biotechnology Research Abstracts</collection><collection>ProQuest Health and Medical</collection><collection>ProQuest Central (purchase pre-March 2016)</collection><collection>Medical Database (Alumni Edition)</collection><collection>Technology Research Database</collection><collection>ProQuest SciTech Collection</collection><collection>ProQuest Technology Collection</collection><collection>ProQuest Natural Science Collection</collection><collection>Hospital Premium Collection</collection><collection>Hospital Premium Collection (Alumni Edition)</collection><collection>ProQuest Central (Alumni) (purchase pre-March 2016)</collection><collection>Materials Science & Engineering Collection</collection><collection>ProQuest Central (Alumni Edition)</collection><collection>ProQuest Central UK/Ireland</collection><collection>ProQuest Central Essentials</collection><collection>Biological Science Collection</collection><collection>ProQuest Central</collection><collection>Technology Collection</collection><collection>Natural Science Collection</collection><collection>ProQuest One Community College</collection><collection>ProQuest Central Korea</collection><collection>Engineering Research Database</collection><collection>Health Research Premium Collection</collection><collection>Health Research Premium Collection (Alumni)</collection><collection>ProQuest Central Student</collection><collection>SciTech Premium Collection</collection><collection>ProQuest Health & Medical Complete (Alumni)</collection><collection>ProQuest Engineering Collection</collection><collection>ProQuest Biological Science Collection</collection><collection>Health & Medical Collection (Alumni Edition)</collection><collection>Medical Database</collection><collection>ProQuest Biological Science Journals</collection><collection>Engineering Database</collection><collection>Biotechnology and BioEngineering Abstracts</collection><collection>Publicly Available Content Database</collection><collection>ProQuest One Academic Eastern Edition (DO NOT USE)</collection><collection>ProQuest One Academic</collection><collection>ProQuest One Academic UKI Edition</collection><collection>ProQuest Central China</collection><collection>Engineering Collection</collection><collection>MEDLINE - Academic</collection><collection>PubMed Central (Full Participant titles)</collection><jtitle>Biomedical engineering online</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Seppänen, Tiina M</au><au>Alho, Olli-Pekka</au><au>Seppänen, Tapio</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Reducing the airflow waveform distortions from breathing style and body position with improved calibration of respiratory effort belts</atitle><jtitle>Biomedical engineering online</jtitle><addtitle>Biomed Eng Online</addtitle><date>2013-09-28</date><risdate>2013</risdate><volume>12</volume><issue>1</issue><spage>97</spage><epage>97</epage><pages>97-97</pages><issn>1475-925X</issn><eissn>1475-925X</eissn><abstract>Respiratory effort belt measurement is a widely used method to monitor respiration. Signal waveforms of respiratory volume and flow may indicate pathological signs of several diseases and, thus, it would be highly desirable to predict them accurately. Calibrated effort belts are sufficiently accurate for estimating respiratory rate, but the respiratory volume and flow prediction accuracies degrade considerably with changes in the subject's body position and breathing style.
An improved calibration method of respiratory effort belts is presented in this paper. It is based on an optimally trained FIR (Finite Impulse Response) filter bank constructed as a MISO system (Multiple-Input Single-Output) between respiratory effort belt signals and the spirometer in order to reduce waveform errors. Ten healthy adult volunteers were recruited. Breathing was varied between the following styles: metronome-guided controlled breathing rate of 0.1 Hz, 0.15 Hz, 0.25 Hz and 0.33 Hz, and a free rate that was felt normal by each subject. Body position was varied between supine, sitting and standing. The proposed calibration method was tested against these variations and compared with the state-of-the-art methods from the literature.
Relative waveform error decreased 60-70% when predicting airflow under changing breathing styles. The coefficient of determination R2 varied between 0.88-0.95 and 0.65-0.79 with the proposed and the standard method, respectively. Standard deviation of respiratory volume error decreased even 80%. The proposed method outperformed other methods.
Results show that not only the respiratory volume can be computed more precisely from the predicted airflow, but also the flow waveforms are very accurate with the proposed method. The method is robust to breathing style changes and body position changes improving greatly the accuracy of the calibration of respiratory effort belts over the standard method. The enhanced accuracy of the belt calibration offers interesting opportunities, e.g. in pulmonary and critical care medicine when objective measurements are required.</abstract><cop>England</cop><pub>BioMed Central Ltd</pub><pmid>24073710</pmid><doi>10.1186/1475-925X-12-97</doi><tpages>1</tpages><oa>free_for_read</oa></addata></record> |
fulltext | fulltext |
identifier | ISSN: 1475-925X |
ispartof | Biomedical engineering online, 2013-09, Vol.12 (1), p.97-97 |
issn | 1475-925X 1475-925X |
language | eng |
recordid | cdi_pubmedcentral_primary_oai_pubmedcentral_nih_gov_3852323 |
source | MEDLINE; DOAJ Directory of Open Access Journals; PubMed Central Open Access; Springer Nature OA Free Journals; EZB-FREE-00999 freely available EZB journals; PubMed Central; SpringerLink Journals - AutoHoldings |
subjects | Abdomen Adult Air Banks (Finance) Calibration Female Humans Linear Models Male Methods Models, Biological Otolaryngology Physiological aspects Posture Pulmonary function tests Regression analysis Respiration Respiratory Function Tests - methods Studies |
title | Reducing the airflow waveform distortions from breathing style and body position with improved calibration of respiratory effort belts |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-10T22%3A53%3A39IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-gale_pubme&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Reducing%20the%20airflow%20waveform%20distortions%20from%20breathing%20style%20and%20body%20position%20with%20improved%20calibration%20of%20respiratory%20effort%20belts&rft.jtitle=Biomedical%20engineering%20online&rft.au=Sepp%C3%A4nen,%20Tiina%20M&rft.date=2013-09-28&rft.volume=12&rft.issue=1&rft.spage=97&rft.epage=97&rft.pages=97-97&rft.issn=1475-925X&rft.eissn=1475-925X&rft_id=info:doi/10.1186/1475-925X-12-97&rft_dat=%3Cgale_pubme%3EA534654663%3C/gale_pubme%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=1442702659&rft_id=info:pmid/24073710&rft_galeid=A534654663&rfr_iscdi=true |