Role of aortic stent graft oversizing and barb characteristics on folding

Objective To evaluate folding in infrarenal stent grafts in relation to oversizing, barb angle, and barb length using computed tomography images of stent grafts deployed in explanted bovine aortas. Methods Computed tomography data from an in vitro investigation on the effect of oversizing of 4% to 4...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of vascular surgery 2012-05, Vol.55 (5), p.1401-1409
Hauptverfasser: Lin, Kathleen K., BS, Kratzberg, Jarin A., PhD, Raghavan, Madhavan L., PhD
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Objective To evaluate folding in infrarenal stent grafts in relation to oversizing, barb angle, and barb length using computed tomography images of stent grafts deployed in explanted bovine aortas. Methods Computed tomography data from an in vitro investigation on the effect of oversizing of 4% to 45% (n = 19), barb length of 2 to 7 mm (n = 11), and barb angle of 10° to 90° (n = 7) on device fixation were examined for instances of folding. Folding was classified as circumferential or longitudinal and quantified on an ordinal scale based on codified criteria. Cumulative fold ranking from 0 (no fold) to 6 (two severe folds) for each deployment was used as the measure of folding observed. Results Of the 37 cases, cumulative mean ± standard deviation fold ranking for stent grafts oversized >30% (n = 5) was significantly greater than the rest (3.4 ± 1.7 vs 0.5 ± 1.2, respectively; Mann-Whitney U test; P < .005). When barb length was varied from 2 to 7 mm (oversizing held at 10%-20%), folding was noted in one of 11 cases. Similarly, when barb angle was varied from 0° (vertical) to 90° (horizontal), folding was not noted in any of the seven cases. The pullout force was not significantly different between stent grafts with and without folding (5.4 ± 1.95 vs 5.12 ± 1.89 N, respectively; P > .5). At least one instance of folding was noted in the seven of seven (100%) stent grafts with oversizing >23.5% and in only five of 30 (14%) stent grafts with oversizing 30%. Large variations in barb length and angle did not aggravate folding risk when oversized within the recommended range of 10% to 20%.
ISSN:0741-5214
1097-6809
DOI:10.1016/j.jvs.2011.10.080