Background Electron beam melting (E-beam) is a new technology to produce 3-dimensional surface topographies for cementless orthopedic implants. Methods The friction coefficients of two newly developed E-beam produced surface topographies were in vitro compared with sandblasted E-beam and titanium pl...
Gespeichert in:
Veröffentlicht in: | Archives of orthopaedic and trauma surgery 2011-05, Vol.131 (5), p.711-718 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Background
Electron beam melting (E-beam) is a new technology to produce 3-dimensional surface topographies for cementless orthopedic implants.
Methods
The friction coefficients of two newly developed E-beam produced surface topographies were in vitro compared with sandblasted E-beam and titanium plasma sprayed controls. Bone ingrowth (direct bone–implant contact) was determined by implanting the samples in the femoral condyles of 6 goats for a period of 6 weeks.
Results
Friction coefficients of the new structures were comparable to the titanium plasma sprayed control. The direct bone–implant contact was 23.9 and 24.5% for the new surface structures. Bone–implant contact of the sandblasted and titanium plasma sprayed control was 18.2 and 25.5%, respectively.
Conclusions
The frictional and bone ingrowth properties of the E-beam produced surface structures are similar to the plasma-sprayed control. However, since the maximal bone ingrowth had not been reached for the E-beam structures during the relatively short-term period, longer-term follow-up studies are needed to assess whether the E-beam structures lead to a better long-term performance than surfaces currently in use, such as titanium plasma spray coating. |
---|---|