Aerosol Hazards from Some Clinical Laboratory Apparatus

Infectious diseases, such as infectious hepatitis, may be transmitted from patients to staff in clinical laboratories and dialysis units during the performance of necessary procedures. This prompted us to examine the propensity of several items of commonly used apparatus to produce aerosols of infec...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:BMJ 1972-03, Vol.1 (5800), p.594-597
Hauptverfasser: Rutter, D. A., Evans, C. G. T.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 597
container_issue 5800
container_start_page 594
container_title BMJ
container_volume 1
creator Rutter, D. A.
Evans, C. G. T.
description Infectious diseases, such as infectious hepatitis, may be transmitted from patients to staff in clinical laboratories and dialysis units during the performance of necessary procedures. This prompted us to examine the propensity of several items of commonly used apparatus to produce aerosols of infected blood. These were a microhaematocrit centrifuge, an AutoAnalyzer stirrer, and a microtonometer. We also compared some blood sample tubes with two types of lid. The results showed that all of these are potential sources of infection for laboratory staff.
doi_str_mv 10.1136/bmj.1.5800.594
format Article
fullrecord <record><control><sourceid>jstor_pubme</sourceid><recordid>TN_cdi_pubmedcentral_primary_oai_pubmedcentral_nih_gov_1787591</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><jstor_id>25417901</jstor_id><sourcerecordid>25417901</sourcerecordid><originalsourceid>FETCH-LOGICAL-b507t-e2b6699eaa31a832bf820ddb9fedb0e48bbd0a5cb4d66c5fb6721427f6e52a013</originalsourceid><addsrcrecordid>eNqFkM1PwyAchonRzGXu6s2kiScPrUD5aC8my-accdEYv44EWqqd7ZjQGudfL3PL1JMnIO_Dj5cHgEMEI4RidqrqWYQimkAY0ZTsgC4iLAlpEse7oAsh5CEihO-DvnMzf8QxT1JGOqBDIYo91QV8oK1xpgom8lPa3AWFNXVwZ2odDKtyXmayCqZSGSsbY5fBYLGQftu6A7BXyMrp_mbtgYfx-f1wEk5vLi6Hg2moKORNqLFiLE21lDGSSYxVkWCY5yotdK6gJolSOZQ0UyRnLKOFYhwjgnnBNMXSl-yBs_XcRatqnWd63lhZiYUta2mXwshS_E3m5Yt4Nu8C8YTTdDXgeDPAmrdWu0bMTGvnvrNHOMPMKyOeitZU5m04q4vtCwiKlWrhVQskVqqFV-0vHP3utcU3Yn_ymfPitjGmBPH0-1_hOi9doz-2ubSvgvGYU3H9OBRj-kSuRrcjQT1_suZXPf7p9gVnLqAC</addsrcrecordid><sourcetype>Open Access Repository</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>1776261464</pqid></control><display><type>article</type><title>Aerosol Hazards from Some Clinical Laboratory Apparatus</title><source>MEDLINE</source><source>JSTOR Archive Collection A-Z Listing</source><source>PubMed Central</source><creator>Rutter, D. A. ; Evans, C. G. T.</creator><creatorcontrib>Rutter, D. A. ; Evans, C. G. T.</creatorcontrib><description>Infectious diseases, such as infectious hepatitis, may be transmitted from patients to staff in clinical laboratories and dialysis units during the performance of necessary procedures. This prompted us to examine the propensity of several items of commonly used apparatus to produce aerosols of infected blood. These were a microhaematocrit centrifuge, an AutoAnalyzer stirrer, and a microtonometer. We also compared some blood sample tubes with two types of lid. The results showed that all of these are potential sources of infection for laboratory staff.</description><identifier>ISSN: 0007-1447</identifier><identifier>ISSN: 0959-8138</identifier><identifier>EISSN: 1468-5833</identifier><identifier>DOI: 10.1136/bmj.1.5800.594</identifier><identifier>PMID: 5013833</identifier><language>eng</language><publisher>England: British Medical Journal Publishing Group</publisher><subject>Aerosols ; Air Microbiology ; Air sampling ; Autoanalysis ; Blood ; Centrifugation ; Health hazards ; Hematocrit ; Hepatitis A ; Humans ; Infections ; Infectious diseases ; Laboratory Infection - etiology ; Papers and Originals ; Radiation counters ; Specimens</subject><ispartof>BMJ, 1972-03, Vol.1 (5800), p.594-597</ispartof><rights>Copyright 1972 British Medical Journal</rights><rights>Copyright BMJ Publishing Group LTD Mar 4, 1972</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-b507t-e2b6699eaa31a832bf820ddb9fedb0e48bbd0a5cb4d66c5fb6721427f6e52a013</citedby></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktopdf>$$Uhttps://www.jstor.org/stable/pdf/25417901$$EPDF$$P50$$Gjstor$$H</linktopdf><linktohtml>$$Uhttps://www.jstor.org/stable/25417901$$EHTML$$P50$$Gjstor$$H</linktohtml><link.rule.ids>230,314,727,780,784,803,885,27924,27925,53791,53793,58017,58250</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/5013833$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Rutter, D. A.</creatorcontrib><creatorcontrib>Evans, C. G. T.</creatorcontrib><title>Aerosol Hazards from Some Clinical Laboratory Apparatus</title><title>BMJ</title><addtitle>Br Med J</addtitle><description>Infectious diseases, such as infectious hepatitis, may be transmitted from patients to staff in clinical laboratories and dialysis units during the performance of necessary procedures. This prompted us to examine the propensity of several items of commonly used apparatus to produce aerosols of infected blood. These were a microhaematocrit centrifuge, an AutoAnalyzer stirrer, and a microtonometer. We also compared some blood sample tubes with two types of lid. The results showed that all of these are potential sources of infection for laboratory staff.</description><subject>Aerosols</subject><subject>Air Microbiology</subject><subject>Air sampling</subject><subject>Autoanalysis</subject><subject>Blood</subject><subject>Centrifugation</subject><subject>Health hazards</subject><subject>Hematocrit</subject><subject>Hepatitis A</subject><subject>Humans</subject><subject>Infections</subject><subject>Infectious diseases</subject><subject>Laboratory Infection - etiology</subject><subject>Papers and Originals</subject><subject>Radiation counters</subject><subject>Specimens</subject><issn>0007-1447</issn><issn>0959-8138</issn><issn>1468-5833</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>1972</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><sourceid>ABUWG</sourceid><sourceid>AFKRA</sourceid><sourceid>BENPR</sourceid><sourceid>CCPQU</sourceid><recordid>eNqFkM1PwyAchonRzGXu6s2kiScPrUD5aC8my-accdEYv44EWqqd7ZjQGudfL3PL1JMnIO_Dj5cHgEMEI4RidqrqWYQimkAY0ZTsgC4iLAlpEse7oAsh5CEihO-DvnMzf8QxT1JGOqBDIYo91QV8oK1xpgom8lPa3AWFNXVwZ2odDKtyXmayCqZSGSsbY5fBYLGQftu6A7BXyMrp_mbtgYfx-f1wEk5vLi6Hg2moKORNqLFiLE21lDGSSYxVkWCY5yotdK6gJolSOZQ0UyRnLKOFYhwjgnnBNMXSl-yBs_XcRatqnWd63lhZiYUta2mXwshS_E3m5Yt4Nu8C8YTTdDXgeDPAmrdWu0bMTGvnvrNHOMPMKyOeitZU5m04q4vtCwiKlWrhVQskVqqFV-0vHP3utcU3Yn_ymfPitjGmBPH0-1_hOi9doz-2ubSvgvGYU3H9OBRj-kSuRrcjQT1_suZXPf7p9gVnLqAC</recordid><startdate>19720304</startdate><enddate>19720304</enddate><creator>Rutter, D. A.</creator><creator>Evans, C. G. T.</creator><general>British Medical Journal Publishing Group</general><general>British Medical Association</general><general>BMJ Publishing Group LTD</general><scope>BSCLL</scope><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>3V.</scope><scope>7RV</scope><scope>7X7</scope><scope>7XB</scope><scope>88E</scope><scope>8FI</scope><scope>8FJ</scope><scope>8FK</scope><scope>ABUWG</scope><scope>AFKRA</scope><scope>BENPR</scope><scope>BTHHO</scope><scope>CCPQU</scope><scope>FYUFA</scope><scope>GHDGH</scope><scope>K9.</scope><scope>KB0</scope><scope>M0S</scope><scope>M1P</scope><scope>NAPCQ</scope><scope>PQEST</scope><scope>PQQKQ</scope><scope>PQUKI</scope><scope>PRINS</scope><scope>5PM</scope></search><sort><creationdate>19720304</creationdate><title>Aerosol Hazards from Some Clinical Laboratory Apparatus</title><author>Rutter, D. A. ; Evans, C. G. T.</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-b507t-e2b6699eaa31a832bf820ddb9fedb0e48bbd0a5cb4d66c5fb6721427f6e52a013</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>1972</creationdate><topic>Aerosols</topic><topic>Air Microbiology</topic><topic>Air sampling</topic><topic>Autoanalysis</topic><topic>Blood</topic><topic>Centrifugation</topic><topic>Health hazards</topic><topic>Hematocrit</topic><topic>Hepatitis A</topic><topic>Humans</topic><topic>Infections</topic><topic>Infectious diseases</topic><topic>Laboratory Infection - etiology</topic><topic>Papers and Originals</topic><topic>Radiation counters</topic><topic>Specimens</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Rutter, D. A.</creatorcontrib><creatorcontrib>Evans, C. G. T.</creatorcontrib><collection>Istex</collection><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>ProQuest Central (Corporate)</collection><collection>Nursing &amp; Allied Health Database</collection><collection>Health &amp; Medical Collection</collection><collection>ProQuest Central (purchase pre-March 2016)</collection><collection>Medical Database (Alumni Edition)</collection><collection>Hospital Premium Collection</collection><collection>Hospital Premium Collection (Alumni Edition)</collection><collection>ProQuest Central (Alumni) (purchase pre-March 2016)</collection><collection>ProQuest Central (Alumni Edition)</collection><collection>ProQuest Central UK/Ireland</collection><collection>ProQuest Central</collection><collection>BMJ Journals</collection><collection>ProQuest One Community College</collection><collection>Health Research Premium Collection</collection><collection>Health Research Premium Collection (Alumni)</collection><collection>ProQuest Health &amp; Medical Complete (Alumni)</collection><collection>Nursing &amp; Allied Health Database (Alumni Edition)</collection><collection>Health &amp; Medical Collection (Alumni Edition)</collection><collection>Medical Database</collection><collection>Nursing &amp; Allied Health Premium</collection><collection>ProQuest One Academic Eastern Edition (DO NOT USE)</collection><collection>ProQuest One Academic</collection><collection>ProQuest One Academic UKI Edition</collection><collection>ProQuest Central China</collection><collection>PubMed Central (Full Participant titles)</collection><jtitle>BMJ</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Rutter, D. A.</au><au>Evans, C. G. T.</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Aerosol Hazards from Some Clinical Laboratory Apparatus</atitle><jtitle>BMJ</jtitle><addtitle>Br Med J</addtitle><date>1972-03-04</date><risdate>1972</risdate><volume>1</volume><issue>5800</issue><spage>594</spage><epage>597</epage><pages>594-597</pages><issn>0007-1447</issn><issn>0959-8138</issn><eissn>1468-5833</eissn><abstract>Infectious diseases, such as infectious hepatitis, may be transmitted from patients to staff in clinical laboratories and dialysis units during the performance of necessary procedures. This prompted us to examine the propensity of several items of commonly used apparatus to produce aerosols of infected blood. These were a microhaematocrit centrifuge, an AutoAnalyzer stirrer, and a microtonometer. We also compared some blood sample tubes with two types of lid. The results showed that all of these are potential sources of infection for laboratory staff.</abstract><cop>England</cop><pub>British Medical Journal Publishing Group</pub><pmid>5013833</pmid><doi>10.1136/bmj.1.5800.594</doi><tpages>4</tpages><oa>free_for_read</oa></addata></record>
fulltext fulltext
identifier ISSN: 0007-1447
ispartof BMJ, 1972-03, Vol.1 (5800), p.594-597
issn 0007-1447
0959-8138
1468-5833
language eng
recordid cdi_pubmedcentral_primary_oai_pubmedcentral_nih_gov_1787591
source MEDLINE; JSTOR Archive Collection A-Z Listing; PubMed Central
subjects Aerosols
Air Microbiology
Air sampling
Autoanalysis
Blood
Centrifugation
Health hazards
Hematocrit
Hepatitis A
Humans
Infections
Infectious diseases
Laboratory Infection - etiology
Papers and Originals
Radiation counters
Specimens
title Aerosol Hazards from Some Clinical Laboratory Apparatus
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-04T18%3A35%3A33IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-jstor_pubme&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Aerosol%20Hazards%20from%20Some%20Clinical%20Laboratory%20Apparatus&rft.jtitle=BMJ&rft.au=Rutter,%20D.%20A.&rft.date=1972-03-04&rft.volume=1&rft.issue=5800&rft.spage=594&rft.epage=597&rft.pages=594-597&rft.issn=0007-1447&rft.eissn=1468-5833&rft_id=info:doi/10.1136/bmj.1.5800.594&rft_dat=%3Cjstor_pubme%3E25417901%3C/jstor_pubme%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=1776261464&rft_id=info:pmid/5013833&rft_jstor_id=25417901&rfr_iscdi=true