Three-dimensional Patient-specific Quality Assurance Using Beam Delivery Log Data for Pencil Beam Scanning Carbon-ion Radiotherapy
In the pencil beam scanning carbon-ion radiotherapy, spot positions are arranged in three dimensions throughout the entire target region. Therefore, dose deviations can occur due to spot position errors in the target. However, performing three-dimensional measurements for routine patient-specific qu...
Gespeichert in:
Veröffentlicht in: | In vivo (Athens) 2024-11, Vol.38 (6), p.2935-2944 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | 2944 |
---|---|
container_issue | 6 |
container_start_page | 2935 |
container_title | In vivo (Athens) |
container_volume | 38 |
creator | Takashina, Masaaki Yagi, Masashi Noguchi, Yuuki Nakaji, Taku Hamatani, Noriaki Tsubouchi, Toshiro Kanai, Tatsuaki |
description | In the pencil beam scanning carbon-ion radiotherapy, spot positions are arranged in three dimensions throughout the entire target region. Therefore, dose deviations can occur due to spot position errors in the target. However, performing three-dimensional measurements for routine patient-specific quality assurance (PSQA) is difficult because a simple measurement method has not been established. This study aimed to establish a three-dimensional dose verification method using beam delivery log data.
Pencil beam dose distributions in water were generated through Monte Carlo (MC) calculations. Treatment beam dose distribution was calculated by superposing the pencil beam dose distributions, considering given spot positions and monitor units (referred to as semi-MC, SMC). The aim of this study was to perform gamma analysis (GA) using dose distributions of log data-based SMC instead of measured dose for PSQA. To verify SMC, the SMC depth-dose curves were compared with the measured dose. To assess the equivalence between the proposed and measurement-based methods, pass rates of two-dimensional GA were compared. Furthermore, a three-dimensional GA was performed to investigate clinically suitable criteria.
The SMC dose curves were consistent with measured dose, with a deviation |
doi_str_mv | 10.21873/invivo.13776 |
format | Article |
fullrecord | <record><control><sourceid>proquest_pubme</sourceid><recordid>TN_cdi_pubmedcentral_primary_oai_pubmedcentral_nih_gov_11535962</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>3122637055</sourcerecordid><originalsourceid>FETCH-LOGICAL-c1896-5a4b8a85aacde79816d34046de8ab1de763fe5b8b8f881fbfac00f88360c0e293</originalsourceid><addsrcrecordid>eNpVkc1v1DAQxS1ERbeFI1fkI5cUO44_ckJlSwFpJQq0Ejdr4kx2jRJ7aycr7ZW_vKHbVvQ0o5mf3pvRI-QtZ2clN1p88GHnd_GMC63VC7LguuaFllX9kixYKU1hJP99TE5y_sOY0oyVr8ixqCutRV0vyN_rTUIsWj9gyD4G6OkVjB7DWOQtOt95R39M0PtxT89znhIEh_Qm-7CmnxAGeoG932Ha01Vc0wsYgXYx0SsMzvcH4peDEP7xS0hNDMXsQn9C6-O4wQTb_Wty1EGf8c1DPSU3l5-vl1-L1fcv35bnq8JxU6tCQtUYMBLAtahrw1UrKlapFg00fB4p0aFsTGM6Y3jXdOAYm1uhmGNY1uKUfDzobqdmwNbNPybo7Tb5AdLeRvD2-Sb4jV3HneVcClmrclZ4_6CQ4u2EebSDzw77HgLGKVvBy1IJzaSc0eKAuhRzTtg9-XBm74Ozh-DsfXAz_-7_457ox6TEHR7AmO8</addsrcrecordid><sourcetype>Open Access Repository</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>3122637055</pqid></control><display><type>article</type><title>Three-dimensional Patient-specific Quality Assurance Using Beam Delivery Log Data for Pencil Beam Scanning Carbon-ion Radiotherapy</title><source>MEDLINE</source><source>EZB-FREE-00999 freely available EZB journals</source><source>PubMed Central</source><creator>Takashina, Masaaki ; Yagi, Masashi ; Noguchi, Yuuki ; Nakaji, Taku ; Hamatani, Noriaki ; Tsubouchi, Toshiro ; Kanai, Tatsuaki</creator><creatorcontrib>Takashina, Masaaki ; Yagi, Masashi ; Noguchi, Yuuki ; Nakaji, Taku ; Hamatani, Noriaki ; Tsubouchi, Toshiro ; Kanai, Tatsuaki</creatorcontrib><description>In the pencil beam scanning carbon-ion radiotherapy, spot positions are arranged in three dimensions throughout the entire target region. Therefore, dose deviations can occur due to spot position errors in the target. However, performing three-dimensional measurements for routine patient-specific quality assurance (PSQA) is difficult because a simple measurement method has not been established. This study aimed to establish a three-dimensional dose verification method using beam delivery log data.
Pencil beam dose distributions in water were generated through Monte Carlo (MC) calculations. Treatment beam dose distribution was calculated by superposing the pencil beam dose distributions, considering given spot positions and monitor units (referred to as semi-MC, SMC). The aim of this study was to perform gamma analysis (GA) using dose distributions of log data-based SMC instead of measured dose for PSQA. To verify SMC, the SMC depth-dose curves were compared with the measured dose. To assess the equivalence between the proposed and measurement-based methods, pass rates of two-dimensional GA were compared. Furthermore, a three-dimensional GA was performed to investigate clinically suitable criteria.
The SMC dose curves were consistent with measured dose, with a deviation <5%. In two-dimensional GA, pass rates for the proposed method were generally lower than those for measurement-based method. The results of the three-dimensional GA indicated that the proposed method, with criteria of 3%-3 mm and 3%-2 mm, had capabilities comparable to the measurement-based method.
The developed three-dimensional log data-based PSQA method with criteria of 3%-3 mm and 3%-2 mm is clinically applicable.</description><identifier>ISSN: 0258-851X</identifier><identifier>ISSN: 1791-7549</identifier><identifier>EISSN: 1791-7549</identifier><identifier>DOI: 10.21873/invivo.13776</identifier><identifier>PMID: 39477399</identifier><language>eng</language><publisher>Greece: International Institute of Anticancer Research</publisher><subject>Heavy Ion Radiotherapy - methods ; Humans ; Monte Carlo Method ; Phantoms, Imaging ; Quality Assurance, Health Care ; Radiotherapy Dosage ; Radiotherapy Planning, Computer-Assisted - methods</subject><ispartof>In vivo (Athens), 2024-11, Vol.38 (6), p.2935-2944</ispartof><rights>Copyright © 2024, International Institute of Anticancer Research (Dr. George J. Delinasios), All rights reserved.</rights><rights>2024 The Author(s). Published by the International Institute of Anticancer Research. 2024</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktopdf>$$Uhttps://www.ncbi.nlm.nih.gov/pmc/articles/PMC11535962/pdf/$$EPDF$$P50$$Gpubmedcentral$$Hfree_for_read</linktopdf><linktohtml>$$Uhttps://www.ncbi.nlm.nih.gov/pmc/articles/PMC11535962/$$EHTML$$P50$$Gpubmedcentral$$Hfree_for_read</linktohtml><link.rule.ids>230,314,727,780,784,885,27924,27925,53791,53793</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/39477399$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Takashina, Masaaki</creatorcontrib><creatorcontrib>Yagi, Masashi</creatorcontrib><creatorcontrib>Noguchi, Yuuki</creatorcontrib><creatorcontrib>Nakaji, Taku</creatorcontrib><creatorcontrib>Hamatani, Noriaki</creatorcontrib><creatorcontrib>Tsubouchi, Toshiro</creatorcontrib><creatorcontrib>Kanai, Tatsuaki</creatorcontrib><title>Three-dimensional Patient-specific Quality Assurance Using Beam Delivery Log Data for Pencil Beam Scanning Carbon-ion Radiotherapy</title><title>In vivo (Athens)</title><addtitle>In Vivo</addtitle><description>In the pencil beam scanning carbon-ion radiotherapy, spot positions are arranged in three dimensions throughout the entire target region. Therefore, dose deviations can occur due to spot position errors in the target. However, performing three-dimensional measurements for routine patient-specific quality assurance (PSQA) is difficult because a simple measurement method has not been established. This study aimed to establish a three-dimensional dose verification method using beam delivery log data.
Pencil beam dose distributions in water were generated through Monte Carlo (MC) calculations. Treatment beam dose distribution was calculated by superposing the pencil beam dose distributions, considering given spot positions and monitor units (referred to as semi-MC, SMC). The aim of this study was to perform gamma analysis (GA) using dose distributions of log data-based SMC instead of measured dose for PSQA. To verify SMC, the SMC depth-dose curves were compared with the measured dose. To assess the equivalence between the proposed and measurement-based methods, pass rates of two-dimensional GA were compared. Furthermore, a three-dimensional GA was performed to investigate clinically suitable criteria.
The SMC dose curves were consistent with measured dose, with a deviation <5%. In two-dimensional GA, pass rates for the proposed method were generally lower than those for measurement-based method. The results of the three-dimensional GA indicated that the proposed method, with criteria of 3%-3 mm and 3%-2 mm, had capabilities comparable to the measurement-based method.
The developed three-dimensional log data-based PSQA method with criteria of 3%-3 mm and 3%-2 mm is clinically applicable.</description><subject>Heavy Ion Radiotherapy - methods</subject><subject>Humans</subject><subject>Monte Carlo Method</subject><subject>Phantoms, Imaging</subject><subject>Quality Assurance, Health Care</subject><subject>Radiotherapy Dosage</subject><subject>Radiotherapy Planning, Computer-Assisted - methods</subject><issn>0258-851X</issn><issn>1791-7549</issn><issn>1791-7549</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2024</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNpVkc1v1DAQxS1ERbeFI1fkI5cUO44_ckJlSwFpJQq0Ejdr4kx2jRJ7aycr7ZW_vKHbVvQ0o5mf3pvRI-QtZ2clN1p88GHnd_GMC63VC7LguuaFllX9kixYKU1hJP99TE5y_sOY0oyVr8ixqCutRV0vyN_rTUIsWj9gyD4G6OkVjB7DWOQtOt95R39M0PtxT89znhIEh_Qm-7CmnxAGeoG932Ha01Vc0wsYgXYx0SsMzvcH4peDEP7xS0hNDMXsQn9C6-O4wQTb_Wty1EGf8c1DPSU3l5-vl1-L1fcv35bnq8JxU6tCQtUYMBLAtahrw1UrKlapFg00fB4p0aFsTGM6Y3jXdOAYm1uhmGNY1uKUfDzobqdmwNbNPybo7Tb5AdLeRvD2-Sb4jV3HneVcClmrclZ4_6CQ4u2EebSDzw77HgLGKVvBy1IJzaSc0eKAuhRzTtg9-XBm74Ozh-DsfXAz_-7_457ox6TEHR7AmO8</recordid><startdate>202411</startdate><enddate>202411</enddate><creator>Takashina, Masaaki</creator><creator>Yagi, Masashi</creator><creator>Noguchi, Yuuki</creator><creator>Nakaji, Taku</creator><creator>Hamatani, Noriaki</creator><creator>Tsubouchi, Toshiro</creator><creator>Kanai, Tatsuaki</creator><general>International Institute of Anticancer Research</general><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope><scope>5PM</scope></search><sort><creationdate>202411</creationdate><title>Three-dimensional Patient-specific Quality Assurance Using Beam Delivery Log Data for Pencil Beam Scanning Carbon-ion Radiotherapy</title><author>Takashina, Masaaki ; Yagi, Masashi ; Noguchi, Yuuki ; Nakaji, Taku ; Hamatani, Noriaki ; Tsubouchi, Toshiro ; Kanai, Tatsuaki</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c1896-5a4b8a85aacde79816d34046de8ab1de763fe5b8b8f881fbfac00f88360c0e293</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2024</creationdate><topic>Heavy Ion Radiotherapy - methods</topic><topic>Humans</topic><topic>Monte Carlo Method</topic><topic>Phantoms, Imaging</topic><topic>Quality Assurance, Health Care</topic><topic>Radiotherapy Dosage</topic><topic>Radiotherapy Planning, Computer-Assisted - methods</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Takashina, Masaaki</creatorcontrib><creatorcontrib>Yagi, Masashi</creatorcontrib><creatorcontrib>Noguchi, Yuuki</creatorcontrib><creatorcontrib>Nakaji, Taku</creatorcontrib><creatorcontrib>Hamatani, Noriaki</creatorcontrib><creatorcontrib>Tsubouchi, Toshiro</creatorcontrib><creatorcontrib>Kanai, Tatsuaki</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><collection>PubMed Central (Full Participant titles)</collection><jtitle>In vivo (Athens)</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Takashina, Masaaki</au><au>Yagi, Masashi</au><au>Noguchi, Yuuki</au><au>Nakaji, Taku</au><au>Hamatani, Noriaki</au><au>Tsubouchi, Toshiro</au><au>Kanai, Tatsuaki</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Three-dimensional Patient-specific Quality Assurance Using Beam Delivery Log Data for Pencil Beam Scanning Carbon-ion Radiotherapy</atitle><jtitle>In vivo (Athens)</jtitle><addtitle>In Vivo</addtitle><date>2024-11</date><risdate>2024</risdate><volume>38</volume><issue>6</issue><spage>2935</spage><epage>2944</epage><pages>2935-2944</pages><issn>0258-851X</issn><issn>1791-7549</issn><eissn>1791-7549</eissn><abstract>In the pencil beam scanning carbon-ion radiotherapy, spot positions are arranged in three dimensions throughout the entire target region. Therefore, dose deviations can occur due to spot position errors in the target. However, performing three-dimensional measurements for routine patient-specific quality assurance (PSQA) is difficult because a simple measurement method has not been established. This study aimed to establish a three-dimensional dose verification method using beam delivery log data.
Pencil beam dose distributions in water were generated through Monte Carlo (MC) calculations. Treatment beam dose distribution was calculated by superposing the pencil beam dose distributions, considering given spot positions and monitor units (referred to as semi-MC, SMC). The aim of this study was to perform gamma analysis (GA) using dose distributions of log data-based SMC instead of measured dose for PSQA. To verify SMC, the SMC depth-dose curves were compared with the measured dose. To assess the equivalence between the proposed and measurement-based methods, pass rates of two-dimensional GA were compared. Furthermore, a three-dimensional GA was performed to investigate clinically suitable criteria.
The SMC dose curves were consistent with measured dose, with a deviation <5%. In two-dimensional GA, pass rates for the proposed method were generally lower than those for measurement-based method. The results of the three-dimensional GA indicated that the proposed method, with criteria of 3%-3 mm and 3%-2 mm, had capabilities comparable to the measurement-based method.
The developed three-dimensional log data-based PSQA method with criteria of 3%-3 mm and 3%-2 mm is clinically applicable.</abstract><cop>Greece</cop><pub>International Institute of Anticancer Research</pub><pmid>39477399</pmid><doi>10.21873/invivo.13776</doi><tpages>10</tpages><oa>free_for_read</oa></addata></record> |
fulltext | fulltext |
identifier | ISSN: 0258-851X |
ispartof | In vivo (Athens), 2024-11, Vol.38 (6), p.2935-2944 |
issn | 0258-851X 1791-7549 1791-7549 |
language | eng |
recordid | cdi_pubmedcentral_primary_oai_pubmedcentral_nih_gov_11535962 |
source | MEDLINE; EZB-FREE-00999 freely available EZB journals; PubMed Central |
subjects | Heavy Ion Radiotherapy - methods Humans Monte Carlo Method Phantoms, Imaging Quality Assurance, Health Care Radiotherapy Dosage Radiotherapy Planning, Computer-Assisted - methods |
title | Three-dimensional Patient-specific Quality Assurance Using Beam Delivery Log Data for Pencil Beam Scanning Carbon-ion Radiotherapy |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2024-12-20T10%3A10%3A41IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_pubme&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Three-dimensional%20Patient-specific%20Quality%20Assurance%20Using%20Beam%20Delivery%20Log%20Data%20for%20Pencil%20Beam%20Scanning%20Carbon-ion%20Radiotherapy&rft.jtitle=In%20vivo%20(Athens)&rft.au=Takashina,%20Masaaki&rft.date=2024-11&rft.volume=38&rft.issue=6&rft.spage=2935&rft.epage=2944&rft.pages=2935-2944&rft.issn=0258-851X&rft.eissn=1791-7549&rft_id=info:doi/10.21873/invivo.13776&rft_dat=%3Cproquest_pubme%3E3122637055%3C/proquest_pubme%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=3122637055&rft_id=info:pmid/39477399&rfr_iscdi=true |