Pulse oximetry signal loss during hypoxic episodes in preterm infants receiving automated oxygen control

The aim of this study was to analyze signal loss (SL) resulting from low signal quality of pulse oximetry-derived hemoglobin oxygen saturation (SpO 2 ) measurements during prolonged hypoxemic episodes (pHE) in very preterm infants receiving automatic oxygen control (AOC). We did a post hoc analysis...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:European journal of pediatrics 2024-07, Vol.183 (7), p.2865-2869
Hauptverfasser: Langanky, Lukas O., Kreutzer, Karen B., Poets, Christian F., Franz, Axel R., Schwarz, Christoph E.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The aim of this study was to analyze signal loss (SL) resulting from low signal quality of pulse oximetry-derived hemoglobin oxygen saturation (SpO 2 ) measurements during prolonged hypoxemic episodes (pHE) in very preterm infants receiving automatic oxygen control (AOC). We did a post hoc analysis of a randomized crossover study of AOC, programmed to set FiO 2 to “back-up FiO 2 ” during SL. In 24 preterm infants (median (interquartile range)) gestational age 25.3 (24.6 to 25.6) weeks, recording time 12.7 h (12.2 to 13.6 h) per infant, we identified 76 pHEs (median duration 119 s (86 to 180 s)). In 50 (66%) pHEs, SL occurred for a median duration of 51 s (33 to 85 s) and at a median frequency of 2 (1 to 2) SL-periods per pHE. SpO 2 before and after SL was similar (82% (76 to 88%) vs 82% (76 to 87%), p  = 0.3)).   Conclusion : SL is common during pHE and must hence be considered in AOC-algorithm designs. Administering a “backup FiO 2 ” (which reflects FiO 2 -requirements during normoxemia) during SL may prolong pHE with SL.   Trial registration : The study was registered at www.clinicaltrials.gov under the registration no. NCT03785899. What is Known: • Previous studies examined SpO2 signal loss (SL) during routine manual oxygen control being rare, but pronounced in lower SpO2 states. • Oxygen titration during SL is unlikely to be beneficial as SpO2 may recover to a normoxic range. What is New: • Periods of low signal quality of SpO2 are common during pHEs and while supported with automated oxygen control (SPOC), FiO2 is set to a back-up value reflecting FiO2 requirements during normoxemia in response to SL, although SpO2 remained below target until signal recovery. • FiO2 overshoots following pHEs were rare during AOC and occurred with a delayed onset; therefore, increased FiO2 during SL does not necessarily lead to overshoots.
ISSN:1432-1076
0340-6199
1432-1076
DOI:10.1007/s00431-024-05549-9