Novel ALK immunohistochemistry assay (clone OTI1A4, Dako) is a sensitive, reliable marker for identifying ALK rearrangements in lung adenocarcinomas: A validation study

Abstract Objectives We present the first study validating the recent Dako ALK assay (clone OTI1A4, in vitro diagnostic) for detecting ALK rearrangements in lung adenocarcinoma. Methods Lung adenocarcinoma cases between 2011 and 2023 were retrospectively collected to create a cohort of 203 samples. C...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:American journal of clinical pathology 2024-01, Vol.161 (1), p.71-82
Hauptverfasser: Eren, Ozgur Can, Mericoz, Cisel Aydin, Bozkurtlar, Emine, Bulutay, Pinar, Baygul, Arzu, Kulac, Ibrahim
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Abstract Objectives We present the first study validating the recent Dako ALK assay (clone OTI1A4, in vitro diagnostic) for detecting ALK rearrangements in lung adenocarcinoma. Methods Lung adenocarcinoma cases between 2011 and 2023 were retrospectively collected to create a cohort of 203 samples. Cases were stained with Dako ALK OTI1A4 and Ventana ALK D5F3 and reviewed by 3 pathologists independently. Correlation between assays, including their sensitivity and specificity, was evaluated. Results The cohort (n = 203) consisted of resections, core needle biopsies, and cell blocks. Agreement between Dako ALK OTI1A4 and Ventana ALK D5F3 assays was “almost perfect,” with κ = 0.89. The sensitivity and specificity of the Dako ALK OTI1A4 assay were 93.3% and 96%, respectively, in a subgroup of 55 molecularly confirmed cases (n = 30 with and n = 25 without ALK rearrangement). Conclusions Immunohistochemistry-based assays provide a valid and reasonably priced alternative, especially in settings where molecular confirmatory tests are neither offered nor accessible. Given high interassay and molecular concordance, we propose that the novel Dako OTI1A4 assay can be reliably used to identify cases with ALK rearrangement.
ISSN:0002-9173
1943-7722
DOI:10.1093/ajcp/aqad111