MR‐linac daily semi‐automated end‐to‐end quality control verification
Purpose Adaptive radiation therapy (ART) on the integrated Elekta Unity magnetic resonance (MR)‐linac requires routine quality assurance to verify delivery accuracy and system data transfer. In this work, our objective was to develop and validate a novel automated end‐to‐end test suite that verifies...
Gespeichert in:
Veröffentlicht in: | Journal of applied clinical medical physics 2023-05, Vol.24 (5), p.e13916-n/a |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | n/a |
---|---|
container_issue | 5 |
container_start_page | e13916 |
container_title | Journal of applied clinical medical physics |
container_volume | 24 |
creator | Malkov, Victor N. Winter, Jeff D. Mateescu, Dan Létourneau, Daniel |
description | Purpose
Adaptive radiation therapy (ART) on the integrated Elekta Unity magnetic resonance (MR)‐linac requires routine quality assurance to verify delivery accuracy and system data transfer. In this work, our objective was to develop and validate a novel automated end‐to‐end test suite that verifies data transfer between multiple software platforms and quantifies the performance of multiple machine subcomponents critical to the ART process.
Methods
We designed and implemented a software tool to quantify the MR and megavoltage (MV) isocenter coincidence, treatment couch positioning consistency, isocenter shift accuracy for the adapted plan as well as the MLC and jaw position accuracy following the beam aperture adaptation. Our tool employs a reference treatment plan with a simulated isocenter shift generated on an MR image of a readily available phantom with MR and MV visible fiducials. Execution of the test occurs within the standard adapt‐to‐position (ATP) clinical workflow with MV images collected of the delivered treatment fields. Using descriptive statistics, we quantified uncertainty in couch positioning, isocentre shift as well as the jaw and MLC positions of the adapted fields. We also executed sensitivity measurements to evaluate the detection algorithm's performance.
Results
We report the results of 301 daily testing instances. We demonstrated consistent tracking of the MR‐to‐MV alignment with respect to the established value and to detect small changes on the order of 0.2 mm following machine service events. We found couch position consistency relative to the test baseline value was within 95% CI [–0.31, 0.26 mm]. For phantom shifts that form the basis for the plan adaptation, we found agreement between MV‐image‐detected phantom shift and online image registration, within ± 1.5 mm in all directions with a 95% CI difference of [–1.29, 0.79 mm]. For beam aperture adaptation accuracy, we found differences between the planned and detected jaw positions had a mean value of 0.27 mm and 95% CI of [–0.29, 0.82 mm] and –0.17 mm and 95% CI of [–0.37, 0.05 mm] for the MLC positions. Automated fiducial detected accuracy was within 0.08 ± 0.20 mm of manual localization. Introduced jaw and MLC position errors (1–10 mm) were detected within 0.55 mm (within 1 mm for 15/256 instances for the jaws). Phantom shifts (1.3 or 5 mm in each cardinal direction) from a reference position were detected within 0.26 mm.
Conclusions
We have demonstrated the accuracy and sensi |
doi_str_mv | 10.1002/acm2.13916 |
format | Article |
fullrecord | <record><control><sourceid>proquest_pubme</sourceid><recordid>TN_cdi_pubmedcentral_primary_oai_pubmedcentral_nih_gov_10161066</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>2775623859</sourcerecordid><originalsourceid>FETCH-LOGICAL-c4776-16e1807cd1c47e6e876267596ce578fcfc3fc91adb2d944042b2d76597b5d663</originalsourceid><addsrcrecordid>eNp9kctKxDAYhYMoXkY3PoAU3Igwmj9tbiuRwRvMIIj7kElTjaSNNu3I7HwEn9EnMeOoqAs3yfmTj8P5OQjtAj4CjMmxNjU5glwCW0GbQAkbSgnF6g-9gbZifMAYQORiHW3kjLMcC7qJJpObt5dX7xptslI7P8-irV160n0Xat3ZMrNNmeYupCPJ7KnX3nXzzISma4PPZrZ1lTO6c6HZRmuV9tHufN4DdHt-dju6HI6vL65Gp-OhKThnQ2AWBOamhDRbZgVnhHEqmbGUi8pUJq-MBF1OSSmLAhckCc6o5FNaMpYP0MnS9rGf1rY0NiXRXj22rtbtXAXt1O-fxt2ruzBTgIEB_nA4-HRow1NvY6dqF431Xjc29FERzikjuaAyoft_0IfQt01aTxGBZS4kowvDwyVl2hBja6vvNIDVoiW1aEl9tJTgvZ_5v9GvWhIAS-DZeTv_x0qdjiZkafoO_86gcw</addsrcrecordid><sourcetype>Open Access Repository</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2809389656</pqid></control><display><type>article</type><title>MR‐linac daily semi‐automated end‐to‐end quality control verification</title><source>MEDLINE</source><source>Wiley Online Library</source><source>Wiley Online Library Open Access</source><source>PubMed Central</source><source>Directory of Open Access Journals</source><source>EZB Electronic Journals Library</source><creator>Malkov, Victor N. ; Winter, Jeff D. ; Mateescu, Dan ; Létourneau, Daniel</creator><creatorcontrib>Malkov, Victor N. ; Winter, Jeff D. ; Mateescu, Dan ; Létourneau, Daniel</creatorcontrib><description>Purpose
Adaptive radiation therapy (ART) on the integrated Elekta Unity magnetic resonance (MR)‐linac requires routine quality assurance to verify delivery accuracy and system data transfer. In this work, our objective was to develop and validate a novel automated end‐to‐end test suite that verifies data transfer between multiple software platforms and quantifies the performance of multiple machine subcomponents critical to the ART process.
Methods
We designed and implemented a software tool to quantify the MR and megavoltage (MV) isocenter coincidence, treatment couch positioning consistency, isocenter shift accuracy for the adapted plan as well as the MLC and jaw position accuracy following the beam aperture adaptation. Our tool employs a reference treatment plan with a simulated isocenter shift generated on an MR image of a readily available phantom with MR and MV visible fiducials. Execution of the test occurs within the standard adapt‐to‐position (ATP) clinical workflow with MV images collected of the delivered treatment fields. Using descriptive statistics, we quantified uncertainty in couch positioning, isocentre shift as well as the jaw and MLC positions of the adapted fields. We also executed sensitivity measurements to evaluate the detection algorithm's performance.
Results
We report the results of 301 daily testing instances. We demonstrated consistent tracking of the MR‐to‐MV alignment with respect to the established value and to detect small changes on the order of 0.2 mm following machine service events. We found couch position consistency relative to the test baseline value was within 95% CI [–0.31, 0.26 mm]. For phantom shifts that form the basis for the plan adaptation, we found agreement between MV‐image‐detected phantom shift and online image registration, within ± 1.5 mm in all directions with a 95% CI difference of [–1.29, 0.79 mm]. For beam aperture adaptation accuracy, we found differences between the planned and detected jaw positions had a mean value of 0.27 mm and 95% CI of [–0.29, 0.82 mm] and –0.17 mm and 95% CI of [–0.37, 0.05 mm] for the MLC positions. Automated fiducial detected accuracy was within 0.08 ± 0.20 mm of manual localization. Introduced jaw and MLC position errors (1–10 mm) were detected within 0.55 mm (within 1 mm for 15/256 instances for the jaws). Phantom shifts (1.3 or 5 mm in each cardinal direction) from a reference position were detected within 0.26 mm.
Conclusions
We have demonstrated the accuracy and sensitivity of a daily end‐to‐end test suite capable of detecting errors in multiple machine subcomponents including system data transfer. Our test suite evaluates the entire treatment workflow and has captured system communication issues prior to patient treatment. With automated processing and the use of a standard vendor‐provided phantom, it is possible to expand to other Unity sites.</description><identifier>ISSN: 1526-9914</identifier><identifier>EISSN: 1526-9914</identifier><identifier>DOI: 10.1002/acm2.13916</identifier><identifier>PMID: 36763085</identifier><language>eng</language><publisher>United States: John Wiley & Sons, Inc</publisher><subject>Accuracy ; adaptive ; Automation ; daily ; Elekta Unity ; end‐to‐end ; Humans ; Magnetic Resonance Imaging - methods ; Magnetic Resonance Spectroscopy ; MRL ; Particle Accelerators ; Phantoms, Imaging ; Planning ; Quality Control ; Radiation Oncology Physics ; Radiation therapy ; Radiotherapy Planning, Computer-Assisted - methods ; Registration ; Software</subject><ispartof>Journal of applied clinical medical physics, 2023-05, Vol.24 (5), p.e13916-n/a</ispartof><rights>2023 The Authors. published by Wiley Periodicals, LLC on behalf of The American Association of Physicists in Medicine.</rights><rights>2023 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, LLC on behalf of The American Association of Physicists in Medicine.</rights><rights>2023. This work is published under http://creativecommons.org/licenses/by/4.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c4776-16e1807cd1c47e6e876267596ce578fcfc3fc91adb2d944042b2d76597b5d663</citedby><cites>FETCH-LOGICAL-c4776-16e1807cd1c47e6e876267596ce578fcfc3fc91adb2d944042b2d76597b5d663</cites><orcidid>0000-0002-6765-7248</orcidid></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktopdf>$$Uhttps://www.ncbi.nlm.nih.gov/pmc/articles/PMC10161066/pdf/$$EPDF$$P50$$Gpubmedcentral$$Hfree_for_read</linktopdf><linktohtml>$$Uhttps://www.ncbi.nlm.nih.gov/pmc/articles/PMC10161066/$$EHTML$$P50$$Gpubmedcentral$$Hfree_for_read</linktohtml><link.rule.ids>230,314,723,776,780,860,881,1411,11541,27901,27902,45550,45551,46027,46451,53766,53768</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/36763085$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Malkov, Victor N.</creatorcontrib><creatorcontrib>Winter, Jeff D.</creatorcontrib><creatorcontrib>Mateescu, Dan</creatorcontrib><creatorcontrib>Létourneau, Daniel</creatorcontrib><title>MR‐linac daily semi‐automated end‐to‐end quality control verification</title><title>Journal of applied clinical medical physics</title><addtitle>J Appl Clin Med Phys</addtitle><description>Purpose
Adaptive radiation therapy (ART) on the integrated Elekta Unity magnetic resonance (MR)‐linac requires routine quality assurance to verify delivery accuracy and system data transfer. In this work, our objective was to develop and validate a novel automated end‐to‐end test suite that verifies data transfer between multiple software platforms and quantifies the performance of multiple machine subcomponents critical to the ART process.
Methods
We designed and implemented a software tool to quantify the MR and megavoltage (MV) isocenter coincidence, treatment couch positioning consistency, isocenter shift accuracy for the adapted plan as well as the MLC and jaw position accuracy following the beam aperture adaptation. Our tool employs a reference treatment plan with a simulated isocenter shift generated on an MR image of a readily available phantom with MR and MV visible fiducials. Execution of the test occurs within the standard adapt‐to‐position (ATP) clinical workflow with MV images collected of the delivered treatment fields. Using descriptive statistics, we quantified uncertainty in couch positioning, isocentre shift as well as the jaw and MLC positions of the adapted fields. We also executed sensitivity measurements to evaluate the detection algorithm's performance.
Results
We report the results of 301 daily testing instances. We demonstrated consistent tracking of the MR‐to‐MV alignment with respect to the established value and to detect small changes on the order of 0.2 mm following machine service events. We found couch position consistency relative to the test baseline value was within 95% CI [–0.31, 0.26 mm]. For phantom shifts that form the basis for the plan adaptation, we found agreement between MV‐image‐detected phantom shift and online image registration, within ± 1.5 mm in all directions with a 95% CI difference of [–1.29, 0.79 mm]. For beam aperture adaptation accuracy, we found differences between the planned and detected jaw positions had a mean value of 0.27 mm and 95% CI of [–0.29, 0.82 mm] and –0.17 mm and 95% CI of [–0.37, 0.05 mm] for the MLC positions. Automated fiducial detected accuracy was within 0.08 ± 0.20 mm of manual localization. Introduced jaw and MLC position errors (1–10 mm) were detected within 0.55 mm (within 1 mm for 15/256 instances for the jaws). Phantom shifts (1.3 or 5 mm in each cardinal direction) from a reference position were detected within 0.26 mm.
Conclusions
We have demonstrated the accuracy and sensitivity of a daily end‐to‐end test suite capable of detecting errors in multiple machine subcomponents including system data transfer. Our test suite evaluates the entire treatment workflow and has captured system communication issues prior to patient treatment. With automated processing and the use of a standard vendor‐provided phantom, it is possible to expand to other Unity sites.</description><subject>Accuracy</subject><subject>adaptive</subject><subject>Automation</subject><subject>daily</subject><subject>Elekta Unity</subject><subject>end‐to‐end</subject><subject>Humans</subject><subject>Magnetic Resonance Imaging - methods</subject><subject>Magnetic Resonance Spectroscopy</subject><subject>MRL</subject><subject>Particle Accelerators</subject><subject>Phantoms, Imaging</subject><subject>Planning</subject><subject>Quality Control</subject><subject>Radiation Oncology Physics</subject><subject>Radiation therapy</subject><subject>Radiotherapy Planning, Computer-Assisted - methods</subject><subject>Registration</subject><subject>Software</subject><issn>1526-9914</issn><issn>1526-9914</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2023</creationdate><recordtype>article</recordtype><sourceid>24P</sourceid><sourceid>EIF</sourceid><sourceid>BENPR</sourceid><recordid>eNp9kctKxDAYhYMoXkY3PoAU3Igwmj9tbiuRwRvMIIj7kElTjaSNNu3I7HwEn9EnMeOoqAs3yfmTj8P5OQjtAj4CjMmxNjU5glwCW0GbQAkbSgnF6g-9gbZifMAYQORiHW3kjLMcC7qJJpObt5dX7xptslI7P8-irV160n0Xat3ZMrNNmeYupCPJ7KnX3nXzzISma4PPZrZ1lTO6c6HZRmuV9tHufN4DdHt-dju6HI6vL65Gp-OhKThnQ2AWBOamhDRbZgVnhHEqmbGUi8pUJq-MBF1OSSmLAhckCc6o5FNaMpYP0MnS9rGf1rY0NiXRXj22rtbtXAXt1O-fxt2ruzBTgIEB_nA4-HRow1NvY6dqF431Xjc29FERzikjuaAyoft_0IfQt01aTxGBZS4kowvDwyVl2hBja6vvNIDVoiW1aEl9tJTgvZ_5v9GvWhIAS-DZeTv_x0qdjiZkafoO_86gcw</recordid><startdate>202305</startdate><enddate>202305</enddate><creator>Malkov, Victor N.</creator><creator>Winter, Jeff D.</creator><creator>Mateescu, Dan</creator><creator>Létourneau, Daniel</creator><general>John Wiley & Sons, Inc</general><general>John Wiley and Sons Inc</general><scope>24P</scope><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>3V.</scope><scope>7X7</scope><scope>7XB</scope><scope>88I</scope><scope>8FI</scope><scope>8FJ</scope><scope>8FK</scope><scope>ABUWG</scope><scope>AFKRA</scope><scope>AZQEC</scope><scope>BENPR</scope><scope>CCPQU</scope><scope>DWQXO</scope><scope>FYUFA</scope><scope>GHDGH</scope><scope>GNUQQ</scope><scope>HCIFZ</scope><scope>K9.</scope><scope>M0S</scope><scope>M2P</scope><scope>PIMPY</scope><scope>PQEST</scope><scope>PQQKQ</scope><scope>PQUKI</scope><scope>PRINS</scope><scope>Q9U</scope><scope>7X8</scope><scope>5PM</scope><orcidid>https://orcid.org/0000-0002-6765-7248</orcidid></search><sort><creationdate>202305</creationdate><title>MR‐linac daily semi‐automated end‐to‐end quality control verification</title><author>Malkov, Victor N. ; Winter, Jeff D. ; Mateescu, Dan ; Létourneau, Daniel</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c4776-16e1807cd1c47e6e876267596ce578fcfc3fc91adb2d944042b2d76597b5d663</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2023</creationdate><topic>Accuracy</topic><topic>adaptive</topic><topic>Automation</topic><topic>daily</topic><topic>Elekta Unity</topic><topic>end‐to‐end</topic><topic>Humans</topic><topic>Magnetic Resonance Imaging - methods</topic><topic>Magnetic Resonance Spectroscopy</topic><topic>MRL</topic><topic>Particle Accelerators</topic><topic>Phantoms, Imaging</topic><topic>Planning</topic><topic>Quality Control</topic><topic>Radiation Oncology Physics</topic><topic>Radiation therapy</topic><topic>Radiotherapy Planning, Computer-Assisted - methods</topic><topic>Registration</topic><topic>Software</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Malkov, Victor N.</creatorcontrib><creatorcontrib>Winter, Jeff D.</creatorcontrib><creatorcontrib>Mateescu, Dan</creatorcontrib><creatorcontrib>Létourneau, Daniel</creatorcontrib><collection>Wiley Online Library Open Access</collection><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>ProQuest Central (Corporate)</collection><collection>Health & Medical Collection</collection><collection>ProQuest Central (purchase pre-March 2016)</collection><collection>Science Database (Alumni Edition)</collection><collection>Hospital Premium Collection</collection><collection>Hospital Premium Collection (Alumni Edition)</collection><collection>ProQuest Central (Alumni) (purchase pre-March 2016)</collection><collection>ProQuest Central (Alumni)</collection><collection>ProQuest Central</collection><collection>ProQuest Central Essentials</collection><collection>ProQuest Central</collection><collection>ProQuest One Community College</collection><collection>ProQuest Central</collection><collection>Health Research Premium Collection</collection><collection>Health Research Premium Collection (Alumni)</collection><collection>ProQuest Central Student</collection><collection>SciTech Premium Collection</collection><collection>ProQuest Health & Medical Complete (Alumni)</collection><collection>Health & Medical Collection (Alumni Edition)</collection><collection>Science Database</collection><collection>ProQuest - Publicly Available Content Database</collection><collection>ProQuest One Academic Eastern Edition (DO NOT USE)</collection><collection>ProQuest One Academic</collection><collection>ProQuest One Academic UKI Edition</collection><collection>ProQuest Central China</collection><collection>ProQuest Central Basic</collection><collection>MEDLINE - Academic</collection><collection>PubMed Central (Full Participant titles)</collection><jtitle>Journal of applied clinical medical physics</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Malkov, Victor N.</au><au>Winter, Jeff D.</au><au>Mateescu, Dan</au><au>Létourneau, Daniel</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>MR‐linac daily semi‐automated end‐to‐end quality control verification</atitle><jtitle>Journal of applied clinical medical physics</jtitle><addtitle>J Appl Clin Med Phys</addtitle><date>2023-05</date><risdate>2023</risdate><volume>24</volume><issue>5</issue><spage>e13916</spage><epage>n/a</epage><pages>e13916-n/a</pages><issn>1526-9914</issn><eissn>1526-9914</eissn><abstract>Purpose
Adaptive radiation therapy (ART) on the integrated Elekta Unity magnetic resonance (MR)‐linac requires routine quality assurance to verify delivery accuracy and system data transfer. In this work, our objective was to develop and validate a novel automated end‐to‐end test suite that verifies data transfer between multiple software platforms and quantifies the performance of multiple machine subcomponents critical to the ART process.
Methods
We designed and implemented a software tool to quantify the MR and megavoltage (MV) isocenter coincidence, treatment couch positioning consistency, isocenter shift accuracy for the adapted plan as well as the MLC and jaw position accuracy following the beam aperture adaptation. Our tool employs a reference treatment plan with a simulated isocenter shift generated on an MR image of a readily available phantom with MR and MV visible fiducials. Execution of the test occurs within the standard adapt‐to‐position (ATP) clinical workflow with MV images collected of the delivered treatment fields. Using descriptive statistics, we quantified uncertainty in couch positioning, isocentre shift as well as the jaw and MLC positions of the adapted fields. We also executed sensitivity measurements to evaluate the detection algorithm's performance.
Results
We report the results of 301 daily testing instances. We demonstrated consistent tracking of the MR‐to‐MV alignment with respect to the established value and to detect small changes on the order of 0.2 mm following machine service events. We found couch position consistency relative to the test baseline value was within 95% CI [–0.31, 0.26 mm]. For phantom shifts that form the basis for the plan adaptation, we found agreement between MV‐image‐detected phantom shift and online image registration, within ± 1.5 mm in all directions with a 95% CI difference of [–1.29, 0.79 mm]. For beam aperture adaptation accuracy, we found differences between the planned and detected jaw positions had a mean value of 0.27 mm and 95% CI of [–0.29, 0.82 mm] and –0.17 mm and 95% CI of [–0.37, 0.05 mm] for the MLC positions. Automated fiducial detected accuracy was within 0.08 ± 0.20 mm of manual localization. Introduced jaw and MLC position errors (1–10 mm) were detected within 0.55 mm (within 1 mm for 15/256 instances for the jaws). Phantom shifts (1.3 or 5 mm in each cardinal direction) from a reference position were detected within 0.26 mm.
Conclusions
We have demonstrated the accuracy and sensitivity of a daily end‐to‐end test suite capable of detecting errors in multiple machine subcomponents including system data transfer. Our test suite evaluates the entire treatment workflow and has captured system communication issues prior to patient treatment. With automated processing and the use of a standard vendor‐provided phantom, it is possible to expand to other Unity sites.</abstract><cop>United States</cop><pub>John Wiley & Sons, Inc</pub><pmid>36763085</pmid><doi>10.1002/acm2.13916</doi><tpages>11</tpages><orcidid>https://orcid.org/0000-0002-6765-7248</orcidid><oa>free_for_read</oa></addata></record> |
fulltext | fulltext |
identifier | ISSN: 1526-9914 |
ispartof | Journal of applied clinical medical physics, 2023-05, Vol.24 (5), p.e13916-n/a |
issn | 1526-9914 1526-9914 |
language | eng |
recordid | cdi_pubmedcentral_primary_oai_pubmedcentral_nih_gov_10161066 |
source | MEDLINE; Wiley Online Library; Wiley Online Library Open Access; PubMed Central; Directory of Open Access Journals; EZB Electronic Journals Library |
subjects | Accuracy adaptive Automation daily Elekta Unity end‐to‐end Humans Magnetic Resonance Imaging - methods Magnetic Resonance Spectroscopy MRL Particle Accelerators Phantoms, Imaging Planning Quality Control Radiation Oncology Physics Radiation therapy Radiotherapy Planning, Computer-Assisted - methods Registration Software |
title | MR‐linac daily semi‐automated end‐to‐end quality control verification |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-02-04T13%3A51%3A42IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_pubme&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=MR%E2%80%90linac%20daily%20semi%E2%80%90automated%20end%E2%80%90to%E2%80%90end%20quality%20control%20verification&rft.jtitle=Journal%20of%20applied%20clinical%20medical%20physics&rft.au=Malkov,%20Victor%20N.&rft.date=2023-05&rft.volume=24&rft.issue=5&rft.spage=e13916&rft.epage=n/a&rft.pages=e13916-n/a&rft.issn=1526-9914&rft.eissn=1526-9914&rft_id=info:doi/10.1002/acm2.13916&rft_dat=%3Cproquest_pubme%3E2775623859%3C/proquest_pubme%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=2809389656&rft_id=info:pmid/36763085&rfr_iscdi=true |